{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70127",
      "recalling_firm": "Teleflex, Inc.",
      "address_1": "550 E Swedesford Rd Ste 400",
      "address_2": "N/A",
      "postal_code": "19087-1601",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in the states of FL, MI, WV, and KS.",
      "recall_number": "Z-1008-2015",
      "product_description": "MAQUET Servo Humidifier 163; Model No.: 01-06-8125-8, catalog no. XKC01-06-8125-8.    A heat and moisture condenser (artificial nose) is a device intended to be positioned over a tracheotomy (a surgically created opening in the throat) or tracheal tube (a tube inserted into the trachea) to warm and humidify gases breathed in by a patient.",
      "product_quantity": "18",
      "reason_for_recall": "Connector mount cracks which may lead to a leak failure during use and necessitate immediate replacement in the breathing circuit.  If leak exists and is left untreated, patient may be deprived of adequate anesthetic gases over a period of time and serious adverse health consequences may occur.",
      "recall_initiation_date": "20141204",
      "center_classification_date": "20150206",
      "termination_date": "20170321",
      "report_date": "20150218",
      "code_info": "Batch Numbers: 201321, 201323, and 201324."
    }
  ]
}