      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1008-2014",
      "product_description": "RECON SCREW LG 5.5MM DIA  and M/DN FEM IM NAIL 8MM DIA  to 16 MM DIA. Item codes starting with 0022520, 0022521, 002522, 0022523, 0022524 and 0022525.",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "60177183  60185167  60358113  60497829  60497830  60505959  60537292  60537294  60575711  60620308  60714139  60725700  60817575  60863968  60876909  61031020  61075699  61140183  61178962  61286463  61475180  61530431  60099578  60152959  60185168  60265704  60279629  60358114  60461498  60497831  60505960  60537296  60550315  60575712  60602426  60615919  60690735  60714140  60735634  60857991  60863973  60950791  60966531  61104682  61175267  61194460  61453383  61489974  60099579  60185169  60195438  60265705  60279630  60358115  60380439  60450626  60497832  60505961  60537298  60569917  60575713  60602428  60615938  60711131  60735635  60819473  60857992  60863980  61024758  61066603  61075700  61133155  61158610  61175268  61178939  61242313  61258929  61280067  61303205  61446260  61484472  61604368  61664930  60152960  60177184  60185170  60208612  60279631  60323289  60450628  60461502  60461503  60505962  60537299  60537301  60569919  60602430  60615936  60714142  60735636  60819454  60857994  60863981  60876912  61041742  61071417  61120273  61143242  61158614  61175269  61178942  61256011  61303206  61414385  61439695  61446261  61512407  61530432  61551620  62296412  60095053  60117030  60123298  60137565  60152961  60167141  60185171  60208614  60279632  60301507  60367530  60450630  60451182  60461504  60461505  60497835  60497836  60550316  60561121  60569927  60575715  60620326  60620327  60711137  60714143  60735637  60757038  60819450  60853602  60863982  61004026  61064849  61071418  61104683  61109127  61120958  61158616  61164106  61175270  61190330  61303207  61425887  61453384  61516664  61538686  61607673  61657614  60129672  60140389  60152323  60152962  60179286  60201124  60208616  60279633  60288489  60338428  60358117  60367533  60387524  60422606  60438312  60461506  60461508  60497837  60505963  60505964  60537303  60550317  60569928  60575716  60615937  60653330  60690737  60714144  60735638  60819460  60857995  60863983  60863986  60979593  61041738  61071419  61104684  61120960  61138172  61143243  61147777  61175271  61178947  61211621  61221429  61263171  61286467  61303220  61425889  61453385  61512408  61516671  61560199  61607674  61621137  61746795  62262050  60095055  60117033  60135208  60145195  60152963  60155332  60186866  60201126  60208618  60265706  60279635  60309129  60323292  60338429  60358118  60380444  60399608  60450636  60461510  60461512  60497838  60513225  60538407  60569930  60602433  60602434  60602436  60620329  60653331  60690738  60714145  60735639  60757039  60819463  60863990  60863991  60966631  60979602  61066607  61066609  61079805  61138177  61147780  61158619  61175272  61178937  61190325  61215268  61242315  61258932  61280070  61425890  61453386  61535655  61560201  61564737  61657627  61746796  61746797  60095056  60117035  60135209  60145196  60152325  60155333  60179287  60208620  60253764  60279636  60309132  60358119  60367535  60452041  60461515  60461517  60497839  60513226  60531965  60531967  60569931  60569932  60602440  60615940  60620332  60653332  60674158  60714146  60757040  60819459  60863994  60876915  61024662  61030979  61071420  61133156  61140187  61175273  61178941  61211622  61263173  61280076  61504077  61535665  61560202  61564738  61657628  62262051  60135210  60145197  60152964  60165625  60201127  60208622  60253765  60323294  60358120  60461521  60497840  60513227  60531969  60531971  60550318  60569934  60575717  60602443  60615929  60630479  60653333  60690739  60714147  60735640  60757041  60807998  60807999  60863997  61031021  61079806  61084120  61104685  61133157  61153533  61211623  61221430  61421223  61504078  61535666  61564739  61729527  61746798  60123299  60165626  60208624  60265707  60451180  60461527  60513228  60550319  60569941  60575718  60602445  60653336  60690740  60714148  60757042  60819247  60864001  60864002  61133158  61153534  61175274  61178935  61190324  61530434  61542635  61657630  60142916  60152965  60196073  60196074  60208627  60323299  60358122  60380445  60451181  60468382  60630480  60675786  60711138  60747674  60797058  60819339  60931182  61096515  61104686  61120274  61133159  61175276  61178955  61464938  61530435  61578714  60653339  60690741  60714149  60735641  60757043  60808000  60808001  61037808  61067785  61075701  61120275  61133160  61175279  61512409  60721529  60789988  60789989  60819461  60876917  61037805  61071421  61091730  61120276  61133161  61138180  61158622  61178943  60750995  60822394  60876920  61004033  61091731  61109130  61133162  61138182  61158625  61175281  61625240  60789986  60817577  60819289  60876922  61037810  61091732  61133163  61138185  61178940  61740355  60161893  60465841  60465842  60482812  61748922  60465841R  60465842R  368065  60139987  60665005  60918979  61026905  61069802  61310736  61338965  61584812  61748931  60665005R  61026905R  61069802R  61310736R  61338965R  368064  60651189  61143354  61243929  61267989  61310737  61382368  61310737R  60165697  60187088  60787289  60918982  61100901  61248165  61284067  61303623  60787289R  61100901R  367997  60161894  60885938  60985890  61172870  61184355  61344226  61490707  60985890R  61172870R  61184355R  368061  60161895  60885939  60984609  61303625  61413324  00367996R  60885939R  60984609R  61303625R  368060  60806260  61100902  61263717  61275930  61338966  61476307  60806260R  61100902R  368059  60161896  60576875  61266207  61294493  61344227  61394798  00368059R  60576875R  61266207R  367127  60161897  61069803  61117728  61151158  61220909  61259372  61310738  61509887  00368058R  61069803R  61151158R  61220909R  61259372R  61310738R  367995  369369  61049656  61075883  61248162  61272873  61310739  61490709  61049656R  60665102  60813857  60885940  61277868  61394799  61490710  60813857R  367994  61191146  61294494  61452749  00367994R  61294494R  367993  60615884  61188824  61266209  61307897  61341135  61428518  61597081  61307897R  61428518R  367992  60159589  60426973  60588695  60718440  61172902  61191147  61262579  61284068  61317588  61375266  61468132  00367992R  60426973R  60718440R  61172902R  61262579R  61284068R  61375266R  61069804  61113771  61191148  61248164  61284069  61332761  61446564  61332761R  61446564R  367832  368053  60484444  61113772  61191149  61255542  61338967  61351674  61377324  61502914  60598249  60885942  61243936  61272874  61313667  61446576  60885942R  367989  60159590  61026904  61040160  61172880  61248163  61369915  61415291  61026904R  61172880R  61248163R  369077  61228313  61244686  61310740  61420485  00369077R  00369314R  61310740R  368049  368829  60483127  60513078  60527048  60619792  61303626  61357284  61375267  61446585  61584820  61748921  61263719R  61303626R  60385436  60482822  60500475  60536363  60583474  61092435  61191151  61251810  61275931  61338968  61413325  61484891  61275931R  60119163  60241659  60416111  60483129  60505111  60530689  60558960  60789398  60973760  61049150  61243934  61251808  61277869  61328153  61475198  61049150R  368046  60260476  60416119  60416121  60416124  60530920  60530922  60530924  60577283  60577284  60577286  61502915  61578042  61607676  60260476R  60416124R  60577284R  368045  60127662  60130297  60130298  60513085  60513086  60536366  60574361  60127966R  60130298R  60342209  60830521  60885943  61069805  61159182  61255544  61294504  61344228  60175545R  60342209R  60885943R  367103  368043  60159591  60787290  61175870  61240383  61293408  61375268  61523827  60159591R  368830  369179  60159592  61175871  61251809  61294507  61428520  00369179R  60159592R  61175871R  61294507R  368041  60753356  61178285  61276454  61394801  61458940  60753356R  61276454R  61394801R  60214453  60513089  60530662  60817715  61228305  61284070  61332764  61415297  60817715R  367748  368039  60127970  60173881  60202904  60241660  60249931  60446762  60500194  60525510  60583476  60817718  61069806  61092436  61159183  61172879  61185468  61191153  61255545  61267991  61303627  61434764  61468133  61516758  61092436R  61159183R  367346  369058  60093393  60127972  60187089  60249932  60249938  60416137  60505123  60530696  60530698  60691689  60789399  60813876  60816201  60925345  61049134  61092437  61092438  61117729  61120885  61198487  61234872  61244687  61277870  61313658  61452751  61490711  61556933  61565508  61578043  61581166  61584822  60416137R  60691689R  60816201R  61092438R  60093394  60123993  60127975  60161898  60173883  60177386  60247155  60252722  60416143  60416145  60483274  60530936  60563282  60563283  60563284  60758085  60789401  60813875  60816204  60915923  60915924  60915925  60915926  60915928  61446606  61452752  61490712  60252722R  60915924R  60127663  60127665  60127976  60130287  60130288  60130289  60252727  60252728  60541670  60541671  60541676  60574363  60574364  60576879  61428521  61452753  61523828  61565510  368035  368209  60074241  60202905  60462370  60508696  60541684  60565828  60885944  61240298  61290150  61351682  61446614  61533702  60074241R  60541684R  61290150R  367101  60165698  60472297  60859067  61184353  61240299  61307898  61338969  61413327  60859067R  61240299R  368033  368206  60684923  60854055  61175872  61276456  61328154  60854055R  61175872R  368032  60753355  61184352  61276459  61317589  61479998  00368032R  60753355R  61276459R  368031  60483288  60513093  60530664  61276462  61294508  61332774  61341136  61584824  60093395  60119165  60154570  60165699  60173884  60175546  60176997  60249939  60387658  60405434  60405435  60405436  60446764  60446765  60615887  60654156  60735563  60822110  60992646  61100906  61120886  61143355  61159185  61159186  61191158  61191165  61191166  61191167  61199930  61746117  61191167R  368029  60093396  60098429  60105022  60123996  60127978  60131252  60139998  60173885  60173886  60176998  60214454  60241663  60247156  60264247  60278659  60361389  60367743  60380243  60444853  60478991  60483307  60513094  60513099  60530705  60530712  60541702  60601711  60615888  60654157  60680601  60715082  60735564  60758086  60789402  60789403  60813886  60817720  61049137  61049138  61049139  61113774  61117731  61120887  61159188  61172892  61191168  61198488  61217302  61234873  61244689  61313657  61452754  61468135  61590995  61602067  367874  368220  60093397  60098430  60127979  60140015  60140018  60173887  60192546  60219130  60231128  60264251  60303657  60303658  60405440  60437865  60444855  60483310  60530940  60530949  60530951  60530952  60530954  60645738  60654158  60671230  60680608  60691701  60758087  60789405  60789406  60789407  60817721  60931430  60931431  60931432  60931433  61100907  61113775  61120888  61170152  61172867  61191169  61234878  61237548  61262581  61267995  61270621  61293410  61458942  61476309  61516760  61542183  61614575  61746121  60654158R  60931430R  61120888R  367107  368067  368215  60082935  60098431  60123997  60123998  60127981  60127985  60127987  60127988  60173888  60173889  60208365  60247157  60249941  60462385  60541642  60541643  60541647  60574367  60615889  60680609  60789408  60816211  60816212  60822111  60828293  60828299  61018699  61117732  61120889  61172886  61234870  61303629  61475199  61502917  61542185  61565511  00367107R  368831  60124001  60156389  60187091  60208366  60214455  60278662  60387659  60478995  60483312  60508708  60541724  60541725  60565819  60565820  60645737  60654159  60654160  60715083  60789409  60828359  60828360  60124001R  60208366R  60483312R  60565820R  60645737R  60715083R  367348  368832  369263  60140033  60192547  60281334  60422975  60558961  60573182  60573532  60640169  61198489  61276463  61382369  60140033R  60422975R  60640169R  61198489R  367081  367749  369061  60127989  60601712  60821407  60854056  61276464  61359840  61490715  61626797  60127989R  60601712R  60821407R  60854056R  60426974  60601713  60751917  61533703  60426974R  60601713R  60751917R  367071  367777  60241664  60462393  60483314  60513109  60536424  61117733  61255547  61259375  61272879  61275927  61344229  60536424R  368027  60093398  60119167  60127993  60140036  60173890  60176999  60231129  60247158  60252730  60278665  60444858  60462397  60500198  60500206  60601714  60615891  60725702  60725703  60735568  60833588  61069811  61114017  61114019  61117734  61159189  61159190  61172887  61191170  61191171  61191172  61191174  61217305  61700309  60176999R  60462397R  61114017R  367876  60089209  60093399  60093401  60105023  60124003  60124004  60131253  60138547  60138548  60140039  60161899  60173891  60175547  60177001  60202907  60208369  60214457  60214458  60241665  60241668  60258752  60264256  60278668  60294014  60310256  60317275  60324247  60462403  60479002  60479003  60505144  60530716  60530718  60530721  60530722  60530724  60530730  60530731  60601715  60601716  60645730  60645731  60671234  60680610  60715085  60715086  60789411  60807330  60807334  60925350  60931446  60931447  60973764  61008297  61049127  61069813  61069816  61114020  61114022  61147941  61147942  61147944  61147945  61147946  61172893  61198490  61217306  61244692  61262582  61307901  61351678  61452756  61476310  61490717  61509889  61542187  61562058  61565512  61581168  61641815  61746125  61746126  60241665R  60671234R  60807334R  368214  60124005  60127674  60127677  60130280  60131254  60131255  60138550  60154572  60161901  60173893  60173894  60202909  60202910  60202911  60208370  60208371  60214459  60214460  60214461  60247159  60264260  60264262  60324248  60333255  60348177  60348178  60354622  60354623  60358059  60358060  60414833  60444862  60462406  60479015  60483091  60483323  60513121  60530966  60530978  60563286  60563287  60630437  60630438  60671235  60680612  60680613  60789413  60789414  60789415  60807338  60874408  60949603  60949604  61049145  61065826  61097049  61113776  61114023  61114024  61114025  61147947  61172878  61217308  61228311  61234866  61234867  61244698  61259376  61275928  61277874  61303633  61307902  61313656  61476311  61490719  61508357  61509890  61578046  61581169  61618122  61641816  61746128  60089210  60119170  60124006  60127678  60131256  60131257  60140046  60154573  60154574  60156390  60156391  60173895  60205398  60214462  60214463  60241670  60241673  60241674  60264263  60264264  60264265  60348179  60385446  60399557  60401272  60541853  60558962  60567075  60567077  60567082  60567083  60567085  60574368  60576888  60654163  60680614  60735571  60735572  60779101  60779102  60779103  60807344  60807345  60828361  60833599  60843568  60844198  61018700  61100911  61100912  61114026  61114027  61132995  61147948  61164634  61170151  61211719  61220908  61228308  61228309  61244754  61270624  61307903  61313664  61468137  61508358  61565513  61581170  61748976  60127679R  60214462R  60241673R  60654163R  368810  60098437  60119171  60131258  60154575  60173896  60192548  60208372  60258756  60264266  60264267  60444839  60462413  60483327  60508713  60508715  60508716  60536425  60541728  60541729  60541730  60565813  60565814  60565817  60565818  60565821  60565822  60565823  60565824  60565825  60848495  61172891  61220904  61228306  61307904  61602068  60131258R  60462413R  60541728R  367087  367100  367877  368811  60119173  60140047  60154576  60187092  60192549  60260477  60462417  60479021  60483329  60525555  60536426  60558965  60680615  60827839  60836664  60854057  61285983  00367100R  00368811R  60483329R  60836664R  60465843  60513128  60536431  60569168  60569169  60574360  60848497  60854058  60536431R  60574360R  60854058R  365825  368212  60423628  60789416  60858672  60859069  61428522  00368212R  60423628R  60789416R  60859069R  367136  368024  60482813  60513130  60515295  60530665  61255553  61294509  00368024R  60515294R  60515295R  60530665R  60119174  60140048  60173897  60175548  60231131  60231132  60241675  60241676  60338270  60444863  60482826  60530674  60530675  60597743  60601720  60615893  60621834  60671237  60691703  60715089  60925351  60973762  61008097  61069817  61079843  61117735  61147949  61147950  61191176  61255554  61259378  61275929  61277875  61502919  61746131  60231131R  367349  60093403  60124009  60124010  60127682  60128002  60131260  60138552  60140085  60140086  60154577  60156392  60156393  60173898  60173899  60173900  60187093  60214464  60214465  60214466  60219131  60231133  60231134  60241677  60241678  60241679  60269661  60278672  60294016  60294017  60294019  60303663  60324251  60333257  60479022  60479024  60505066  60505068  60513132  60530733  60530734  60530735  60530736  60530743  60597744  60601721  60645726  60645727  60671240  60715090  60735577  60735578  60789417  60807348  60807350  60874410  60973757  60973758  60973759  61049131  61049132  61065832  61065837  61097067  61114029  61114030  61117736  61133004  61147951  61147952  61147953  61147954  61147955  61198492  61217310  61217311  61234877  61293412  61452758  61452759  61476312  61523830  61523833  61523836  61562061  61565515  61746141  62250623  60231133R  60671240R  366982  367109  368021  60089215  60089223  60093404  60093405  60127685  60127686  60130285  60130286  60131261  60138553  60140088  60154578  60173903  60208373  60214468  60219132  60219133  60219134  60241680  60241681  60241682  60264269  60278675  60303666  60303668  60310264  60317277  60319975  60324252  60479027  60483335  60513135  60513136  60530986  60530991  60530997  60531005  60601722  60601724  60615895  60630440  60645723  60671243  60691705  60691706  60758091  60758092  60789420  60789421  60817724  60874480  60925347  60931454  60931456  61049141  61049143  61065850  61065852  61114031  61114032  61114034  61147958  61172872  61199932  61217312  61234871  61244759  61255555  61267992  61270620  61293413  61313661  61338979  61468138  61475201  61509891  61542191  61562062  61641818  61674950  61677996  60089223R  368836  60089224  60098441  60124014  60127689  60128009  60130262  60130263  60131263  60131264  60140089  60140090  60154579  60165701  60173904  60173905  60177002  60192550  60192554  60208374  60214469  60219135  60219136  60219137  60247161  60269663  60278676  60294020  60294021  60303673  60310265  60310266  60317278  60319976  60319977  60324253  60333260  60333261  60472537  60541626  60541745  60541747  60541751  60541752  60567088  60567091  60576893  60576898  60576902  60576904  60645740  60645742  60654167  60654168  60715094  60715095  60758094  60789422  60807355  60807358  60827431  60830533  60836666  60838308  60885946  61040162  61114040  61114045  61120892  61133006  61147959  61147960  61172866  61191178  61199933  61217313  61220903  61228307  61237549  61244760  61267994  61270626  61285981  61307907  61313659  61458944  61468139  61476313  61509893  61542192  61618126  60567091R  60576896R  60654167R  365821  367072  368068  60082945  60108241  60124015  60154580  60175550  60187095  60197215  60208375  60214470  60249942  60278678  60324254  60338271  60354624  60358061  60483342  60508733  60536442  60565829  60565834  60565835  60597746  60601725  60645746  60671247  60680617  60715096  60715097  60789423  60829846  60829847  60830534  60836667  60843571  60843573  60844199  60844201  60848499  61211720  61220911  61313660  61542193  61565517  61602069  61780093  60214470R  60338271R  60354624R  60462478R  60536442R  368838  368839  60108243  60140091  60154581  60187096  60231135  60247162  60342210  60358063  60361392  60444842  60513143  60550453  60558969  60558970  60680619  60807359  60828362  60836670  60848502  60854049  60854060  61199934  61203580  61220905  00368839R  60247162R  60680619R  367009  367750  60386472  60465845  60465846  60482810  60586962  60747393  60848504  60854061  60747393R  365824  367350  368323  60426975  60836671  60854062  60859074  60854062R  368020  60483349  60513150  60513151  60513154  60536719  60513154R  60093413  60119175  60128015  60197216  60231136  60483352  60500214  60525536  60583475  60615896  60630442  60680620  60691707  60735580  60735581  60822108  61013362  61114046  61114047  61114051  61117738  61120893  61151160  61172901  61191179  61191180  61191181  60175551R  60680620R  60822109R  367351  368000  60119176  60124016  60127690  60127691  60128017  60130292  60131265  60138556  60140094  60154582  60202914  60208376  60214471  60214472  60231137  60278680  60278682  60294023  60319978  60324258  60333263  60483354  60530745  60530750  60597747  60601727  60601728  60645719  60654169  60715099  60758097  60789424  60931448  60931449  60931450  60931451  61049147  61049148  61049149  61092439  61114057  61114061  61114062  61147963  61147964  61170150  61211718  61220906  61244761  61270622  61277878  61382370  61413332  61502922  61533706  61556935  61562063  61574971  61614578  61746150  60140093R  60310268R  60601727R  60715099R  60874478R  368221  368341  60093417  60124017  60128023  60130268  60130269  60131266  60138557  60140102  60140103  60154583  60187097  60187098  60192555  60192556  60202916  60202917  60205401  60219138  60247163  60264271  60278684  60294024  60294025  60303681  60310271  60317280  60479040  60479042  60483361  60513161  60513164  60531010  60531014  60531015  60531020  60601729  60615897  60630443  60680623  60680624  60758098  60789426  60807362  60817725  60822113  60931434  60931435  60931437  60931438  61049120  61049121  61092440  61114063  61114064  61114065  61117739  61120894  61147965  61172881  61191182  61199929  61234879  61244762  61259381  61389798  61413333  61468140  61490730  61508359  61509894  61578049  61675989  61676004  60691708R  60089225  60093419  60119179  60128027  60130278  60131268  60138558  60156395  60156396  60173906  60173907  60177003  60192557  60202918  60202919  60219139  60219141  60247164  60252736  60252737  60252738  60252739  60278691  60278693  60294026  60303685  60319982  60324263  60333265  60444868  60462488  60483363  60483365  60513165  60541634  60541636  60541653  60541657  60541659  60541660  60576912  60576917  60645724  60645725  60654171  60715101  60715102  60735582  60758101  60789427  60827433  60829848  60833608  60836672  60838310  60838311  61018707  61048025  61048026  61114066  61143356  61147966  61147968  61147969  61172862  61184348  61194285  61199927  61199928  61211712  61217314  61237550  61468141  61509895  61542197  61618127  60219141R  367354  60093420  60098442  60119180  60130296  60140106  60154586  60173908  60187099  60192558  60205403  60231139  60252741  60258764  60278704  60333266  60342211  60443894  60443895  60479048  60536726  60541773  60565836  60581924  60615898  60645716  60645717  60654172  60680625  60789429  60817727  60828363  60836673  61143357  61147970  61147971  61159192  61164612  61164613  61194288  61199931  61211713  61147971R  367878  368224  60154587  60175553  60187101  60192559  60351967  60358064  60443896  60444848  60497048  60513171  60558972  60558973  60558974  60715104  60830568  60836674  60854050  60854051  00368224R  60444848R  60558972R  367001  368225  60583661  60597748  60747408  60822101  60833609  60859070  61476315  00368225R  60525571R  60597748R  367736  368226  60124018  60426977  60588696  60813943  60816213  60833612  00367736R  60124018R  60426977R  60816213R  60833612R  368227  60297193  60462496  60527067  60527068  60530670  60462495R  60527067R  60527068R  367074  367314  367808  369374  60479050  60525539  60525540  60530682  60530683  60680626  60894634  61018695  61018698  61117740  61170153  61172871  61211722  61220910  61259382  61285982  61458946  61502923  60530682R  60530683R  368229  369065  60093421  60093422  60128044  60138560  60140108  60173909  60177004  60197217  60202920  60264273  60278706  60479051  60505083  60505088  60513190  60513191  60530759  60530761  60530763  60530771  60530772  60530776  60530779  60789430  60817728  60973755  60973756  61049152  61114070  61120895  61147972  61191185  61217315  61244764  61413334  61446625  61452762  61476316  61562067  61746157  60530771R  367086  60127692  60128048  60128049  60128050  60130272  60154588  60173910  60177389  60192561  60205404  60208378  60208379  60219217  60264276  60278707  60380246  60462502  60483554  60513198  60513199  60531032  60531035  60531040  60583662  60615899  60615901  60621837  60680627  60732923  60732924  60817729  60822107  60937694  61049135  61085259  61114071  61114072  61147973  61147975  61172884  61199935  61217316  61351681  61413335  61452763  61516763  61542199  61556937  61672744  61693750  00367198R  60822107R  366593  368231  368844  60089227  60093424  60124020  60128052  60130266  60130267  60131273  60131274  60138562  60138563  60140110  60154589  60156397  60192562  60197218  60208380  60208381  60219219  60219220  60264277  60264278  60324265  60338294  60426384  60437870  60513205  60513209  60541790  60541793  60541794  60567093  60574375  60576923  60615902  60615904  60735584  60735585  60807365  60817730  60833614  60838314  60838315  60838316  60838317  60885947  60893397  61018709  61117741  61147976  61147977  61217317  61228310  61237551  61243932  61293416  61359843  61375269  61389799  61452764  61468142  61490740  60219220R  60264278R  60833614R  365830  367199  60098445  60098446  60135452  60135454  60161904  60173911  60192564  60202921  60278713  60338295  60465850  60483559  60536737  60536738  60541805  60564448  60645714  60680628  60691709  60822102  60836676  60843574  61023623  61151161  61151162  61184354  61194287  61220898  61389801  61490741  61556938  60192564R  60338295R  60615905R  367099  367355  368846  369853  60128053  60159593  60173912  60197219  60249944  60342213  60395077  60444851  60444852  60479057  60507223  60525563  60536740  60558976  60645732  60680629  60691710  60836677  60848508  61484895  61676189  61693751  00367355R  00368846R  60128053R  60507223R  60536740R  60558976R  60645732R  60836677R  60848508R  365829  366591  367098  367753  368847  60124021  60525572  60525573  60645729  60721966  60836679  60859073  61434765  00366591R  00368847R  60525572R  60859073R  367356  368848  60483090  60665006  60858677  00367356R  60483090R  368849  60632410  60827841  60827842  60919002  61542201  60827841R  60827842R  367132  60140112  60278715  60787293  60828365  60925398  60925399  60925401  00367132R  60140112R  60973761R  367339  60128056  60219221  60483566  60505099  60505103  60523535  61049151  61172890  61237552  61382371  61415298  365832  60108247  60128057  60131275  60140114  60197220  60482819  60483095  60530568  60530570  60530571  60530572  60541817  60817731  60817733  61075885  61114073  61147978  61191195  61244765  61270623  61285980  61382372  61476317  60131275R  60530572R  60817733R  61191195R  366587  368852  60119183  60124022  60128058  60138564  60175555  60205405  60208382  60247165  60367766  60380247  60462592  60482840  60541827  60541829  60541830  60541832  60576929  60615906  60789431  60807376  60843576  60844203  60844204  60885949  61018710  61234876  61244766  61475202  60175555R  60885949R  367781  368853  369530  60082953  60175557  60205406  60462606  60508780  60536746  60541598  60541605  60565838  60576930  60576931  61023624  61023631  61307910  61379626  61452765  61490742  60082953R  60508769R  60541605R  60576931R  367102  368854  369067  60462616  60536747  60561209  60601731  60747405  60821408  60827844  60836680  61344230  61446638  61490751  61700386  00365826R  60462616R  60601731R  60827844R  61700386R  367079  368218  368855  60173913  60513220  60525576  60525577  60854064  60513220R  60525576R  367754  60197222  60691712  60821409  60854065  61428525  61452769  00369068R  60197222R  60691712R  60854065R  61428525R  367782  368857  369376  60386474  60919005  61220907  61284075  61317590  61341137  61428526  00369376R  60386474R  61220907R  61284075R  367881  60562833  61049654  61307911  61328156  61382373  00367881R  60562833R  61049654R  61307911R  61328156R  367779  369540  60192566  60426979  60565051  60602191  60691713  60874456  61117742  61375270  61413336  61446643  61476318  60602191R  60874456R  368859  60159594  60192567  60472544  60583663  60586807  60797046  60833622  61382374  61446647  61484932  60472544R  60583663R  60586807R  60797046R  60833622R  61382374R  367739  368900  60479066  60483572  60536748  60541844  60565839  60565841  60565842  61428527  61450097  00367739R  00368900R  60483573R  367093  367104  368861  368862  369072  60462629  60497051  60541578  60561210  60745751  60854071  61484935  60541578R  60561210R  60745751R  367738  368217  368863  369073  60482805  60513223  60523538  60523541  60858689  60482805R  60523538R  60523541R  60858689R  368207  368864  60602190  60711437  60859071  61428528  00368207R  00368864R  60602190R  60711437R  60859071R  367986  368865  60138565  60426980  60665007  61069825  61266210  61324284  61338985  61468143  00367986R  60423630  60919010  61069827  61243933  61263718  61351675  61446662  60423630R  61069827R  61243933R  61446662R  367091  368867  60140117  61434767  00367091R  00368867R  61434767R  368868  61159193  61276466  61375271  61422220  00368868R  61159193R  61276466R  368869  60477024  60684921  60685551  60854072  61344231  61420486  60477024R  60854072R  368870  369825  60588697  60750985  60854073  61382375  61452772  00368870R  60854073R  368871  60708939  61276467  61338986  61352377  61359844  61476319  00368871R  60708939R  61276467R  61359844R  368872  60187102  60984611  61120897  61306349  61348460  61476320  60187102R  60984611R  61120897R  61306349R  61348460R  367983  368873  60477025  60753357  60859072  61328157  61428529  00368873R  60753357R  60859072R  61328157R  367982  368874  60665008  60753358  61178726  61276468  00367982R  60665008R  369075  61178727  61353281",
      "more_code_info": ""
    }
  ]
}