{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91290",
      "recalling_firm": "Johnson & Johnson Surgical Vision, Inc.",
      "address_1": "31 Technology Dr Ste 200",
      "address_2": "N/A",
      "postal_code": "92618-2302",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: FL, MS, NY, TN, TX, IL, MN, OH.  OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom",
      "recall_number": "Z-1007-2023",
      "product_description": "Single-Use Packs, REF: OPO80",
      "product_quantity": "384",
      "reason_for_recall": "Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.",
      "recall_initiation_date": "20221207",
      "center_classification_date": "20230120",
      "report_date": "20230201",
      "code_info": "UDI-DI/Lot: 05050474573352/60326033",
      "more_code_info": ""
    }
  ]
}