      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1007-2014",
      "product_description": "M/DN R-F IM NAIL 9MM DIA , 10 MM, 11 MM, 12 MM, 13 MM and 14 MM. Item codes starting with 0022400",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "368278  61111052  61159202  61228030  61276758  61307948  61111052R  368279  60275442  60994837  61111053  61255583  61294523  61556944  368280  61098707  61178911  61284930  61298508  61342119  61546150  60158048  60281123  61075888  61159203  61220139  61290248  61332792  61727932  368490  60158049  60506370  60741224  60960320  61111054  61120925  61143374  61220140  61298509  61143374R  368281  60114142  60232597  60518286  60960628  60960818  61220141  61266215  61281313  61317747  60175525  60281124  60958453  60960815  61140792  61159204  61275565  61317749  60960815R  61275565R  368311  368493  60082894  60281333  61075889  61111055  61159205  61220142  61290251  61338952  368112  368494  60186975  60253467  60703526  60994877  61075890  61111056  61143376  61220143  61275566  60186975R  60253467R  60703526R  368284  368285  368948  60082895  60186976  60305055  60958452  60960816  61140794  61271223  61281317  61313785  367853  368496  60376357  60506371  60725289  61075891  61140795  61237562  61281322  61298511  61468151  60506371R  60725289R  368324  368497  60082896  60484441  61020656  61075892  61117767  61220144  61294524  61501046  00368497R  60082896R  368286  368498  60158050  60484442  60718439  61237563  61266214  61294525  61367353  00368498R  60158050R  60484442R  60718439R  61266214R  368287  60186977  61140796  61267607  61290253  61341154  00368287R  60186977R  61140796R  368288  60186978  61178912  61307949  61501047  00368500R  60186978R  61178912R  61501047R  368312  61178913  61317751  61328162  61317751R  368313  60186979  61143378  61211142  61228027  61290256  61328163  00368313R  61211142R  61290256R  368289  60158051  60919031  61275568  00368289R  60158051R  368113  60165694  60677567  60919032  61117768  61140797  61317753  61338953  61428547  367341  368114  60591981  61075893  61117769  61140799  61155653  61237564  61249763  61281323  61453188  61501049  61249763R  61501049R  368115  60591982  61049751  61075894  61114076  61140801  61206020  61214454  61220913  61290257  61501052  61049751R  367183  367833  60214950  60338296  60771516  60931843  61075895  61114077  61117770  61140802  61155654  61214458  61220922  60771516R  368116  60114143  60136484  60186980  60422981  60553621  60640168  60735561  61111057  61143379  61206022  61303817  61338954  368117  60114144  60154554  60338297  60348768  60573173  60758081  60815082  60972651  61112102  61133048  61206021  61220914  61453189  60348768R  368118  368508  60104629  60177401  60214951  60325612  60621819  60931805  60958455  60960817  61075896  61111058  61140803  61228032  00368508R  60104629R  60123507R  60958455R  367771  367854  60093774  60139921  60177402  60401281  60487360  60537981  60582853  60747471  60806257  60931901  61049745  61049746  61075904  61111059  61120926  61206024  61220920  61313786  60806257R  61111059R  365769  60123509  60173849  60219049  60253469  60367703  60506372  60573175  60591986  60684961  60735562  60944278  60973785  61049748  61049750  61075905  61111060  61114078  61228031  61237565  61271219  61313792  60123509R  367184  367344  367823  60119119  60138857  60173850  60219050  60269621  60401282  60582854  60621822  60679608  60820242  60931806  60931807  61049737  61049739  61049742  61111061  61155655  61172928  61220924  61276759  61281324  61458979  61556949  60173850R  60582854R  61155655R  365770  367824  367825  60119120  60241736  60338298  60534411  60561204  60582856  60621823  60679607  60684962  60820244  61111062  61172913  61234890  61260903  61276760  61290262  61468152  61542211  00365770R  00367824R  60679607R  61172913R  61260903R  367307  367856  60123511  60241740  60338299  60348773  60566317  60582857  60591988  60645760  60784479  60784480  61049729  61111063  61133050  61143380  61220923  61275569  61281325  61303819  61338955  61382441  61468153  61764008  00367307R  60501479R  60582857R  60591988R  61468153R  368119  368120  368513  60621824  60732930  60789383  60815080  61172927  61234889  61237566  61260886  61290282  61328164  61453195  61556955  00368120R  60621824R  61234889R  368121  61140804  61143381  61214468  61276761  61281326  61303820  61341155  61475213  00368121R  61140804R  61214468R  61281326R  368290  368291  368515  61179457  61248170  61260896  61294526  61328165  61428548  61501020  00368290R  61328165R  368292  60138858  61075907  61133051  61140805  00368292R  60138858R  61140805R  367219  61111064  61159206  61188775  61298512  00367219R  61111064R  61159206R  61188775R  368294  368517  60219051  60919033  61140806  61290286  61357280  00368294R  00368517R  60919033R  61140806R  367834  60472313  60919034  61111065  61117771  61214459  61441071  00367834R  61117771R  368518  60621825  61143382  61143383  61214456  61220916  61472029  61533709  61214456R  368295  369251  61114079  61172921  61185463  61244845  61255579  61275570  61294527  61338956  61508365  61632574  368296  368519  60112057  60138859  60139931  60241743  60438408  60611280  60684964  61111066  61140813  61155658  61172932  61214455  61220917  60139931R  60684964R  367857  368520  60104630  60136486  60138860  60294043  60426153  60487361  60677565  60684966  61075908  61114080  61140814  61242953  61260902  61281328  61441072  60487361R  367826  368521  60123512  60158052  60186982  60191072  60363714  60501480  60639916  60817701  60958454  61075909  61143384  61214460  61712235  00367826R  60817701R  368522  60119121  60158053  60186983  60241748  60469961  60566322  60592005  60677566  60747473  60820252  61075910  61111068  61168497  61242952  61260899  61285993  61546151  367828  368523  60119123  60136487  60161921  60294045  60426156  60668268  60820249  60944280  61111069  61172925  61234888  61260898  61303821  61501056  00368523R  367858  60123514  60161095  60177403  60191074  60253470  60363715  60387650  60506373  60645749  60715080  61092056  61111070  61172930  61234884  61271221  61328166  61542213  60161095R  60645749R  61092056R  367224  367783  60082898  60082899  60136489  60158054  60214952  60241749  60253472  60469962  60501482  60668270  60931809  60931810  61049733  61049734  61049735  61114081  61172929  61237567  61255587  61260893  61281329  61338957  61508366  61769261  60241749R  60668270R  367774  368525  60082901  60104633  60119125  60154555  60186984  60208580  60214953  60338300  60426157  60487364  60536360  60583549  60592006  60668272  60732932  60944282  61053260  61053264  61111071  61114086  61172923  61266212  61281330  61307951  61584850  60592006R  61111071R  367859  368526  60123516  60191075  60214954  60241750  60320600  60338301  60426160  60501483  60566326  60582860  60583550  60684967  60747482  60789384  60944283  61111073  61143385  61242957  61244846  61285994  61313788  61508367  61542215  61581161  00367859R  00368526R  61542215R  368527  60422982  60482917  60565858  60583551  60630433  60711472  60711473  60784481  60817703  60944281  61111074  61155659  61172920  61242955  61244864  61260904  61285996  61313791  60422982R  60482917R  60817703R  61155659R  367860  368528  60214955  60566329  60582862  60630434  60691649  61168501  61188777  61255585  61260894  61281336  61328167  61369916  00367860R  60214955R  60691649R  61168501R  367186  367829  60621826  61159207  61214457  61285992  61303822  61332793  61357281  61395133  61475214  61556956  00367829R  61159207R  61285992R  368297  60588691  61172912  61234881  61290287  61382442  61556957  60588691R  61172912R  61290287R  61382442R  368298  60208581  61133054  61220918  61266211  61317759  61133054R  61266211R  61317759R  368531  61140815  61220921  61244866  60919035R  61140815R  61244866R  367861  60414087  60848684  60885974  60919036  61075911  61143386  61317760  61143386R  368124  368533  369047  60583573  60684969  60848685  61117772  61206025  61220915  61314021  61453187  61501061  61614588  61453187R  367836  367837  368534  60832643  60833720  61206027  61220919  61458981  61607682  60833720R  367769  368125  368535  60158055  60214956  60278644  60426161  60487367  60583605  60668274  60822098  61143387  61206023  61214453  61542219  60583605R  60668274R  61143387R  367775  368126  368536  60082902  60112059  60154556  60253474  60297516  60297517  60387652  60454977  60469964  60501484  60611279  60691653  60822099  61120927  61172916  61501063  60454977R  60469964R  60691653R  367838  60104634  60136491  60154557  60186986  60208582  60303618  60469965  60514799  60537984  60592007  60684971  60784482  60838981  60885978  61143388  61328168  61453193  60784482R  368127  368538  60112061  60173851  60253486  60487369  60583616  60684973  60816176  60826022  60833723  60837659  61075912  61143389  61172919  61276763  60583616R  367308  367862  60093775  60154558  60208584  60303619  60324909  60377208  60566332  60789385  60838999  60885979  61040190  61048034  61185466  61206019  61276764  61761374  367839  368539  60104635  60158056  60202927  60253487  60303621  60324911  60393521  60566334  60583624  60758082  60833725  61048038  61111076  61234886  61266213  61285995  60324911R  60758082R  61111076R  368128  368540  60093776  60119126  60158057  60173852  60197241  60231397  60260288  60320594  60333249  60506374  60582863  60583626  60621827  60668275  60684976  60789386  60831075  60832644  60837661  60839001  61237568  61260905  61281337  61501034  365776  368129  368541  60098426  60104636  60138861  60191076  60253495  60260324  60310275  60320595  60324904  60324905  60333251  60621828  60668299  60684977  60789387  60837662  60839002  60844155  61143390  61237569  61281338  61338958  61533711  60191076R  60260324R  60333250R  60333251R  60668299R  60844155R  367223  368130  60093777  60123517  60158058  60197242  60231399  60310276  60320596  60324906  60324907  60333252  60333253  60684979  60758083  60789388  60816177  60844156  61114094  61117773  61172938  61237570  61260901  61281339  61290288  61307953  61313795  61581162  61718250  61114094R  368131  368132  60173853  60202928  60231401  60320597  60333254  60566338  60582864  60592008  60668301  60789389  60833726  60844158  61117774  61172924  61234882  61237571  61260891  61303823  61313787  61542221  61621134  00368131R  00368132R  00368542R  60173853R  60202928R  60320597R  60844158R  61237571R  367863  60231402  60426166  60645751  60668302  60816178  60831076  60832645  60837663  60839003  61341156  61621135  00367863R  60645751R  60816178R  61341156R  367221  368133  368544  369838  60583631  60820251  60822095  60848691  61242954  61255586  61285990  61351415  61351691  61501065  00367221R  00368544R  60820251R  60848691R  61351691R  368300  368545  60197244  60600387  61040199  61048035  61140816  61298513  00368300R  60600387R  368301  61228029  61290289  61341157  00368301R  368302  60919039  60960629  61298515  00368302R  60919039R  60960629R  368303  368546  60451196  60718437  61133055  61228028  61244869  61357282  61501067  60451196R  60718437R  367309  368304  368547  60640166  61255580  61271220  61281340  61303824  61375274  61475215  00367309R  60640166R  368548  60750989  60833727  61117775  61140817  61175873  61220925  61501022  00368548R  61220925R  367345  367864  60154559  60208585  60219054  60478916  60482919  60582866  60822096  60831077  60832646  60839004  61040208  61048036  61101501  61168508  61214467  61303825  61313793  60482919R  60822096R  368137  60119128  60123520  60208586  60282380  60426170  60469966  60592011  60806259  61048041  61048042  61101502  61271222  61290290  61621136  60426170R  60592011R  61048041R  61101502R  368551  60098427  60123521  60158059  60208587  60438407  60621829  60765997  61101538  61191132  61275571  61468154  60158059R  60208587R  60816179R  60960630R  368552  60123523  60154560  60422983  60583639  60679609  60776487  60826024  60837665  61303826  61351417  61453192  61533712  61641830  368553  60098428  60143347  60154561  60214957  60363717  60426171  60472497  60537987  60592012  60684980  60732931  60820240  60826023  61075913  61237572  61260892  61351692  60426171R  367865  368554  60136492  60154562  60324913  60426172  60537988  60621830  60789390  60831078  60833728  61075914  61172931  61175874  61328169  61542227  60426172R  61075914R  367310  368141  368555  60093778  60104637  60123525  60165695  60175527  60219056  60219057  60324908  60387654  60426173  60472501  60487375  60506375  60536362  60654154  60668324  60831079  60832647  61120928  61172914  61234883  61260895  61281341  61313789  00367310R  00368555R  60093778R  60487375R  60506375R  60654154R  60832647R  368142  368556  369834  60119129  60123526  60175529  60208588  60219058  60260325  60426174  60469970  60501485  60570941  60583645  60654155  60668325  60684982  60833729  61040203  61040205  61048037  61172926  61338959  61584853  00368142R  60501485R  60570941R  60668325R  368557  37107401  60093779  60165696  60214958  60247149  60260326  60482916  60537990  60570942  60659302  60711474  60747468  60796722  60839005  60844194  61234887  61255581  61281342  61313794  61581164  00365780R  60260326R  60570942R  60659302R  60844161R  60844194R  367867  367868  368558  60219059  60363718  60426175  60537991  60583650  60789391  60831080  60832648  60833730  60844162  60844195  00367867R  60583650R  60789391R  60844162R  367869  368340  368559  60588692  60684983  60837666  60839006  60844163  60844196  61313784  61428549  00367869R  60684983R  60837666R  60844163R  367189  368143  368560  60646893  60815078  60815079  60816180  61328170  61367359  61501069  00367189R  60646893R  60815078R  60815079R  61328170R  368305  60214959  61140818  61303827  61357283  60214959R  61140818R  61303827R  61357283R  368561  60123527  61120929  61140819  61206026  00368306R  00368561R  61120929R  61140819R  61206026R  60123533  60670026  60967107  61143391  61332794  60670026R  61143391R  61332794R  368144  60214473  60414088  60848697  60854074  61178286  61501071  60848697R  367841  368568  60158060  60797007  60832649  60854075  61294528  61441082  60158060R  368570  60143348  60472502  60621833  60822097  60832650  60848698  61338960  60848698R  368308  368486  60082905  60143349  60275696  60423635  60588693  61172904  61172905  61175875  60143349R  60423635R  367311  367870  60154563  60253496  60426177  60640165  60848699  60859066  60154563R  60426177R  60472503R  60640165R  368571  60154564  60197246  60278647  60615942  60668343  60820241  61317761  61338961  61317761R  368146  368572  60154565  60423636  60691654  60854077  60859065  61275572  61290292  61338962  60423636R  60859065R  367312  60154566  60202929  60472506  60639975  60832651  60854078  61175877  61375275  60472506R  61175877R  367842  60123534  60197247  60423637  60697295  60789395  60832652  60859064  61276765  61290293  61458983  60123534R  60832652R  367812  368574  60158061  60214960  60387655  60482918  60506376  60570943  60592013  60703530  60817708  60833731  61360845  61441098  60817708R  367809  368576  60119130  60161922  60247150  60351968  60472507  60506377  60570944  60583658  60668346  60747476  60820253  60822094  61172908  61175878  00367809R  60161922R  60247150R  60570944R  60822094R  368148  368575  60143350  60186988  60219061  60351969  60472508  60570945  60583659  60640163  60668348  60715081  60848701  61453194  60143350R  60186988R  60583659R  60848701R  368149  60150964  60177452  60216322  60253502  60294046  60472509  60583660  60677568  60711471  60747462  60816181  60854079  61294529  61338963  00368577R  60150964R  60253502R  60426179R  60472509R  60677568R  60816181R  60854079R  367871  60123537  60154568  60275697  60539946  60630346  60645750  60691655  60747477  60859062  00367871R  60275697R  60539946R  60645750R  60691655R  367810  60123544  60186989  60484443  60640162  60754804  60854080  00367810R  60123544R  60186989R  60484443R  60640162R  60754804R  60854080R  365785  368309  368580  60154569  60725139  61244870  61281344  61332795  61338964  61453190  00368309R  00368580R  60154569R  60725139R  61244870R  368310  368581  61172906  61172907  61175879  61172907R  61175879R  368150  368582  60668665  60859063  61255582  60668665R  60859063R  61178288R",