  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1005-2014",
      "product_description": "VERSA-FX II KEYLESS TUBE, VERSA-FX II KEYLESS SHORT, VERSA-FX II SHORT TUBE PL, VERSA-FX II STD TUBE PLAT and VERSA-FX II SUPRACONDYLR     Item codes starting with 0011930, 0011931 and 0011941",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "366026  60252546  60386429  60387234  60387236  60664570  60746113  60820564  368890  60131310  60161887  60208595  60337755  60386430  60387237  60674436  60674437  60746114  60820565  62063880  60674437R  366027  60165190  60351204  60406618  60598293  60699479  60714970  60820566  61753280  366028  368891  60189692  60510769  60699480  60714938  60820568  60189692R  366029  368892  60154515  60413139  60740818  60812406  62161712  366030  60844072  60868450  60844072R  368894  60154516  60173867  60175580  60206848  60292114  60366271  60379893  60424741  60452501  60476633  60523009  60523011  60568292  60598291  60598292  60610993  60619081  60646238  60726314  60746117  60911330  60931155  60944509  61076064  61169404  61505367  61523575  61640872  62098503  60082818  60082819  60119113  60129285  60144960  60147708  60175581  60182119  60241838  60278773  60328084  60410030  60435440  60476635  60510770  60539704  60581931  60619076  60664573  60726313  60763191  60787023  60877481  60890087  60890088  60900735  60911331  60928243  60931156  60932380  60938399  60943332  60944510  61140151  61163257  61264711  61294823  61374274  61505374  61505375  61558652  61590985  61640873  61748725  62045196  62063886  368896  60127957  60129286  60156372  60189693  60215781  60252548  60292116  60353487  60383970  60405451  60452504  60502812  60539715  60539717  60568287  60581933  60619073  60646239  60726315  60726316  60726317  60726318  60890089  60911332  60918552  61114103  61120930  61344223  61505376  61558653  61596975  61602061  61677240  62050758  62057618  366505  60129287  60175582  60260273  60410031  60433745  60482396  60510771  60619078  60726319  60763195  60959966  61031789  61128189  61527193  61558654  61602062  61748726  61753014  62045197  368373  369962  60402976  60424743  60445069  60510772  60619079  60968719  61118961  61198403  61264715  61693779  62045195  62063881  368898  370907  60812398  61577082  62279718  00368898R  60165193  60482973  60527776  60533011  368916  60598295  60746132  60820576  60598295R  60746132R  60699481  61895937  60074372R  60699481R  368960  60136494  60189695  60266339  60353489  60370812  60383972  60416018  60473692  60489349  60510773  60548437  60699482  60746134  60746137  60803571  60820577  60959967  60962101  61041079  61076065  61118962  61140152  61341979  61503009  61529227  61596976  61780109  366037  368961  60115491  60129288  60136496  60165202  60202542  60231320  60270737  60278775  60325673  60337763  60353490  60359553  60360711  60383974  60393527  60405454  60420022  60452508  60473693  60510507  60568294  60598296  60628437  60630043  60664579  60699483  60746151  60746152  60803572  60803573  60897286  60911333  60932382  60943333  60972786  60998281  61027493  61041080  61076066  61114104  61118963  61140153  61151189  61198406  61264716  61286854  61341980  61375159  61499393  61509877  61523576  61538708  61565471  61621778  61640874  61677268  61677278  61748727  61748729  60998281R  370385  60118087  60119115  60129289  60129290  60136497  60154517  60156373  60165205  60165206  60165207  60189696  60208598  60215782  60215784  60252550  60252551  60269204  60269207  60278777  60325674  60337769  60353491  60366275  60383975  60383976  60387238  60393528  60405455  60420024  60435441  60445071  60452511  60473694  60476637  60485409  60489351  60510508  60510509  60533026  60548395  60568303  60572660  60598297  60610994  60619085  60629728  60629729  60630044  60664586  60664589  60746153  60787024  60803574  60803575  60845769  60845770  60868454  60868456  60874078  60877482  60911334  60918557  60938403  60959968  60962102  60998282  61011698  61027495  61041081  61086812  61109156  61114105  61140154  61151190  61163260  61198407  61198408  61264717  61294824  61375160  61499395  61499396  61499397  61509878  61511594  61516746  61538709  61565472  61621780  61640875  61677279  61677280  61677281  61693783  62330816  60129289R  60845770R  366039  368918  60129291  60136395  60136498  60197427  60229002  60270738  60366276  60387241  60431508  60473695  60510776  60510777  60746183  60787025  60885155  60911335  60962106  61140155  61198409  61523577  61538710  61621784  61640876  61677285  61681374  60746183R  369830  60115492  60119116  60173868  60219026  60270739  60420025  60445072  60473696  60510779  60664591  60746184  60803576  60803577  60938406  60944512  61198410  61509879  61523578  61621786  61681375  61681376  60270739R  60746184R  368919  369863  60136396  60136499  60301526  60359554  60433693  60530497  60628438  60844063  60844064  61009046  61067763  61092188  366042  368920  369945  60154518  60614796  60674422  60868459  61067764  61738425  62089478  366043  366508  368963  60189698  60685006  60834850  61203357  61222979  61294825  60834850R  61294825R  368964  60189699  61027504  61079827  61118964  61140156  61151191  61352928  61352928R  61140159  61151192  366045  368908  369807  60107394  60136397  60156374  60292117  60321101  60325675  60353492  60383977  60405456  60435442  60473698  60510791  60539721  60628439  60664594  60746185  60803579  60820601  60938408  60968723  61031792  61076067  61140161  61151193  61465687  61493896  61529229  61677287  61749793  61758077  366046  60115493  60337770  60383978  60406619  60746186  60803580  61076068  61151194  61465688  61748731  60115493R  60513666  60831411  61020306  61076069  61681388  61076069R  369623  60115494  60115495  60129292  60136398  60136501  60156375  60165249  60168039  60168040  60208601  60219027  60219028  60252553  60252554  60269208  60269210  60278778  60328046  60337771  60351205  60355843  60373570  60379894  60387242  60405462  60420027  60424744  60435443  60452518  60473702  60485410  60510510  60532223  60532224  60532229  60561635  60572853  60598260  60598261  60632485  60666567  60666568  60666569  60666570  60679348  60679349  60704350  60704351  60726367  60726368  60726369  60787026  60787027  60787028  60820607  60820608  60820609  60897287  60923649  60923650  61062345  61066535  61092191  61109157  61118965  61128190  61132952  61140163  61169405  61198411  61222981  61222982  61260778  61264718  61348694  61377325  61386061  61402002  61417095  61420329  61427109  61427345  61432413  61440680  61516747  61565473  61596977  61672111  61672112  61672115  61681390  61681397  62045192  366047  367048  367049  367407  60072659  60074377  60074383  60093366  60098016  60098019  60098022  60109711  60109712  60118088  60123980  60123981  60123982  60129293  60129294  60129295  60131311  60131312  60136502  60147709  60147710  60147711  60148675  60148676  60163980  60163981  60168041  60175583  60175584  60182121  60182122  60187156  60187157  60208602  60219029  60219030  60219031  60231321  60231322  60231324  60241840  60249948  60252555  60269213  60269214  60270741  60270742  60270743  60292118  60307974  60307975  60324513  60328047  60334031  60337772  60337773  60337774  60340368  60351206  60353494  60355845  60373571  60373572  60379895  60387244  60393535  60393536  60394641  60401179  60402977  60402978  60405469  60413144  60416564  60420029  60420030  60424745  60431509  60433694  60435444  60440757  60445073  60452519  60452521  60459385  60476638  60482397  60485411  60489356  60494309  60498020  60502953  60510554  60517267  60532225  60532226  60532227  60532228  60548460  60551531  60551532  60561632  60568309  60568310  60568311  60598262  60598263  60598265  60630089  60630090  60632497  60666550  60666551  60666553  60666555  60666556  60679351  60679353  60679354  60679357  60685030  60685031  60704352  60704353  60704354  60704355  60714932  60714933  60714934  60714935  60714936  60726364  60726365  60726366  60746189  60746190  60746193  60779868  60779869  60779870  60779871  60787029  60812092  60812093  60812095  60812096  60874080  60877483  60885156  60890078  60897288  60897289  60900737  60911336  60923651  60932383  60943334  60944513  60944514  60959969  60968725  60972788  60997748  60998283  61004981  61009470  61031795  61037481  61041083  61059148  61059149  61062346  61066536  61076070  61086813  61092203  61092206  61104845  61109159  61109161  61114117  61118966  61118968  61118969  61120931  61128191  61132953  61136348  61151196  61151198  61169406  61245976  61245977  61245978  61260779  61264719  61274514  61286856  61286857  61304464  61304465  61304466  61340963  61340964  61347112  61347115  61352929  61357155  61364135  61377326  61382041  61382042  61386062  61391477  61391478  61420330  61427110  61427346  61427347  61432415  61440681  61464978  61473798  61516748  61533692  61542594  61551558  61558656  61565474  61578702  61584962  61596978  61596979  61596980  61621788  61677289  61677291  61677292  61677294  61677296  61693774  61693775  61716224  61716225  61716226  61716228  61753017  61753018  61753020  61964341  62042371  62053151  61677296R  366048  366510  367589  368966  60082822  60093367  60093368  60093370  60093371  60093372  60093373  60093374  60098025  60098026  60098028  60098029  60098030  60098031  60098032  60098033  60106382  60106384  60106385  60106386  60106387  60106389  60106390  60108201  60108202  60108203  60108206  60108207  60108208  60108209  60118089  60118090  60118091  60119117  60123983  60123985  60123986  60123987  60129296  60129297  60129298  60129299  60129300  60129301  60129302  60129303  60129304  60131313  60131314  60131315  60131316  60131317  60136399  60136401  60136402  60136403  60136503  60143357  60143358  60143359  60143360  60144961  60144963  60144964  60148677  60148678  60148679  60148680  60148681  60155264  60155265  60155266  60155267  60155268  60155269  60155270  60155271  60156376  60156377  60156378  60156379  60159595  60159596  60159597  60165263  60165264  60165266  60165271  60165273  60165279  60165285  60168042  60168043  60168044  60168045  60168046  60168047  60173869  60173870  60173871  60175585  60175586  60175587  60175588  60175589  60175590  60182127  60182128  60182129  60187158  60187159  60187160  60187161  60197429  60197430  60197431  60202543  60208604  60208605  60208606  60208607  60208609  60215785  60215788  60215789  60219032  60219033  60219034  60219035  60219036  60219038  60229004  60229005  60231325  60231326  60231327  60231328  60231329  60231330  60231331  60231332  60231333  60231334  60231335  60231336  60241844  60241845  60241846  60241847  60241848  60241849  60241850  60249949  60252556  60252557  60260274  60260275  60260277  60260279  60260281  60260282  60266340  60266341  60266342  60266344  60266345  60266346  60266347  60266348  60269216  60269220  60269222  60278780  60292119  60307976  60307977  60311272  60311273  60314208  60322338  60322339  60324515  60324516  60328048  60328049  60334036  60337775  60337776  60337777  60337778  60340370  60340371  60343527  60343534  60343535  60348275  60351207  60351208  60351209  60351210  60353495  60353639  60355846  60355847  60355848  60355849  60359555  60366277  60367581  60373573  60373575  60373578  60373581  60373583  60373586  60373587  60379897  60379898  60379899  60379901  60379902  60379903  60379904  60383980  60383981  60387246  60387247  60387248  60387249  60387255  60389512  60393537  60393538  60393539  60393540  60394642  60394643  60398864  60398867  60398870  60401181  60403716  60403717  60405470  60405472  60406621  60406623  60406626  60410033  60410034  60416020  60416565  60416566  60420031  60420034  60420038  60424746  60424748  60424749  60427677  60429826  60429828  60429829  60429830  60431352  60433747  60435445  60435446  60435447  60435448  60435449  60440752  60440754  60445074  60445075  60449709  60452531  60452532  60452535  60452538  60454909  60454910  60459388  60463418  60467559  60473707  60473709  60476639  60476640  60476641  60476642  60482399  60482401  60485412  60485413  60485414  60489360  60494312  60494313  60494314  60498021  60498068  60510516  60510518  60510519  60517249  60517250  60517251  60532243  60532247  60532252  60537451  60537461  60537462  60537464  60537465  60537466  60537467  60537468  60543094  60543095  60543097  60543098  60543104  60548661  60548662  60548663  60548693  60548694  60557374  60561633  60563227  60563228  60563229  60568314  60568315  60568316  60568317  60568318  60572662  60581941  60581942  60581943  60581945  60598268  60598270  60598271  60598272  60598273  60611018  60611019  60611020  60611021  60611022  60611023  60611032  60630091  60630092  60630093  60630094  60630095  60630096  60630098  60632504  60646243  60666539  60666540  60666541  60666542  60666543  60666546  60666547  60666548  60679360  60679361  60679362  60679363  60685033  60685034  60685035  60685036  60685057  60704357  60704359  60704360  60704361  60704362  60704363  60704364  60704365  60714922  60714923  60714924  60714926  60714927  60714928  60726327  60726328  60726330  60726332  60726361  60746266  60746267  60746268  60746269  60746270  60746271  60746272  60746273  60746280  60746281  60763196  60763197  60763198  60763199  60763201  60763202  60763203  60763204  60763205  60763206  60779873  60779876  60779877  60779879  60779880  60779881  60779882  60787030  60787031  60787032  60787033  60787034  60787038  60812098  60812099  60812100  60812101  60812102  60812103  60812104  60812106  60820611  60820613  60820614  60820615  60820616  60820620  60820621  60820623  60820624  60820626  60820627  60820628  60820631  60820632  60820634  60855078  60855079  60874081  60874082  60877485  60877486  60885157  60885158  60885159  60885161  60897290  60897292  60900738  60911337  60911339  60911340  60918558  60931162  60932384  60943335  60943336  60944515  60944516  60944517  60944518  60959970  60959971  60959972  60962113  60962115  60968727  60968736  60972789  60972790  60972791  60997749  60997750  60998284  60998285  61004983  61004984  61009048  61009475  61009476  61009477  61009478  61009479  61009480  61009481  61009482  61010334  61010339  61010340  61010341  61011701  61011702  61020312  61031801  61031803  61037482  61037483  61059150  61059151  61076071  61076072  61076074  61086816  61092208  61092209  61092212  61092213  61104846  61109162  61109163  61109164  61109166  61109167  61114126  61114127  61114128  61114130  61118970  61118971  61118972  61118973  61120932  61120933  61120934  61128192  61130024  61132954  61136349  61136350  61136351  61140165  61140166  61148123  61148124  61148125  61148126  61151203  61151205  61151206  61169407  61179732  61179733  61198412  61198413  61198415  61198416  61222983  61222986  61222987  61228323  61228324  61228325  61228326  61245980  61245983  61245984  61245985  61245986  61260780  61264721  61274515  61274516  61274518  61293477  61316180  61316182  61340965  61340967  61340969  61347116  61347117  61347118  61347119  61348695  61348696  61352930  61357156  61357157  61357158  61357160  61359992  61364137  61382043  61382946  61386063  61387470  61387471  61391479  61408294  61412171  61417097  61420331  61420333  61420334  61427111  61428555  61432416  61433930  61437986  61440255  61440686  61447015  61457771  61464979  61464980  61464981  61464982  61473799  61473801  61516749  61533693  61542595  61551560  61565475  61565476  61565477  61578705  61578706  61578707  61584964  61584965  61584966  61596981  61596982  61596983  61596984  61607684  61607685  61613170  61640878  61659695  61659696  61659697  61659698  61659705  61659765  61659775  61659776  61677302  61677303  61681398  61681399  61681401  61681402  61681404  61681406  61681407  61681408  61681409  61681410  61681411  61681412  61681414  61681419  61681420  61684789  61684790  61758093  61758101  61758102  62045199  62049341  62049342  60098026R  60598268R  60779879R  366533  60098034  60098035  60118092  60129305  60129306  60136506  60143362  60143363  60144965  60144966  60159598  60168048  60168049  60173872  60182132  60182133  60202544  60208608  60215792  60219039  60219040  60241852  60241853  60249950  60252558  60266349  60266350  60278781  60337779  60353496  60373592  60373593  60379905  60379906  60387284  60394644  60402359  60405473  60416021  60416567  60420040  60429831  60440755  60452539  60489364  60498071  60517262  60517263  60532265  60532266  60548696  60548697  60568312  60581947  60598275  60598276  60630101  60632506  60666557  60666558  60666560  60666561  60714965  60714966  60714967  60714968  60714969  60726370  60787039  60787040  60787041  60820686  60820690  60897295  60923652  60944519  60972792  60998286  61011703  61020314  61130027  61245988  61245989  61245990  61309050  61309051  61316192  61347122  61352931  61382044  61387473  61412172  61427112  61432417  61437987  61440687  61533694  61551561  61635175  61640879  61640880  61640881  61640882  61748732  61758130  366050  60118093  60129307  60129309  60147712  60147713  60159599  60168050  60168051  60173873  60219041  60219042  60229006  60229007  60252559  60266351  60307978  60322345  60337780  60373594  60387287  60393541  60394645  60420041  60445076  60452540  60473711  60482402  60482974  60510532  60527783  60548706  60572663  60581948  60625420  60646250  60699487  60699488  60699489  60699490  60787042  60787043  60820695  60820697  60820698  60923653  60923654  60932388  60932389  60944520  60959973  60997751  61011704  61062356  61062359  61066537  61076076  61151208  61169410  61198420  61260782  61311406  61311407  61316194  61316195  61316196  61316197  61420341  61432418  61453110  61473802  61621811  61677305  61693792  61693794  61716245  62042565  60572663R  61062359R  369441  60147714  60168052  60231337  60241854  60252561  60266352  60266353  60387288  60406628  60429833  60452541  60476651  60489367  60510784  60530507  60530508  60598299  60699492  60699493  60787074  60820704  60923655  60923656  60959974  61031807  61037484  61062360  61109168  61118974  61169411  61245998  61311408  61311409  61671650  61672107  61681364  62042566  62054582  62054583  60429833R  61031807R  366536  367591  368967  60098036  60155272  60159600  60215793  60266354  60387295  60406630  60476654  60482976  60489370  60510785  60572669  60699495  60787078  61114135  61140168  61148127  61163276  61169412  62045238  366052  367590  369946  60123988  60154519  60154520  60359556  60387301  60402980  60482977  60494315  60510588  60528408  60572721  60581999  61745918  62056352  366513  60112020  60182134  60219043  60389513  60401182  60482979  60482980  60510584  60543108  60582003  61260791  60163982  60402981  60482981  60528426  60482981R  60482982  60528430  61020321  61067765  61086830  60528430R  61020321R  61067765R  60482983  60528433  60582008  60482983R  60528433R  366053  366515  60098037  60109713  60109714  60123989  60129310  60129311  60147715  60154521  60173875  60175591  60182135  60182136  60241855  60241856  60252562  60266355  60278782  60337781  60355850  60373596  60379907  60393542  60405474  60420044  60435450  60454911  60473713  60489375  60494317  60532268  60532269  60548714  60582010  60582011  60598277  60632513  60678765  60678768  60678771  60685060  60704345  60704346  60704347  60704348  60714918  60714919  60714920  60787120  60787122  60845772  60877488  60885162  60911342  60932390  60932391  60943337  60944521  60968734  60972795  60998288  61031810  61037485  61041087  61076077  61086835  61114136  61120935  61140171  61143395  61151210  61198424  61198425  61228329  61245999  61341981  61348699  61357163  61484356  61484357  61484358  61512535  61512536  61538713  61551563  61558658  61590986  61607686  61621818  61640884  61672108  61681380  61681384  368925  60082825  60115496  60143364  60173876  60215794  60266356  60359557  60405475  60433695  60445077  60489378  60548732  60679345  60704349  60787126  60885164  60903904  61004997  61020324  61037486  61086839  61309053  61326721  61344224  61484359  61565478  61964337  60143364R  366055  368968  369560  60118094  60241857  60415265  60473715  60646251  60787129  60820829  61311410  61484360  61607687  61681422  60118094R  60241857R  366521  60136404  60136507  60292120  60351212  60405479  60473717  60482984  60489379  60528465  60582014  60482984R  366056  369947  60093375  60093377  60165182  60175592  60260284  60266357  60337782  60366281  60387302  60394648  60420045  60482985  60489380  60494318  60528461  60533078  60726320  60763210  60763211  60931163  60962120  61004999  61027505  61086844  61114137  61118975  61163278  61264724  61375161  61512223  61565479  61621824  61745100  60093377R  366534  370397  60093379  60115497  60129312  60129313  60136405  60136508  60165183  60168053  60173877  60206851  60219044  60219045  60266358  60266359  60278783  60324517  60337783  60337784  60351214  60366282  60367582  60379908  60393543  60401183  60405480  60420046  60435451  60452542  60454912  60476655  60482986  60489381  60494319  60510628  60528458  60561638  60582016  60646255  60646257  60726321  60763212  60763213  60820715  60820716  60820718  60918559  60943338  60972796  61005002  61086847  61092218  61140174  61148128  61163281  61264726  61512222  61512224  61523580  61558659  61565480  61693787  62042752  60435451R  60646255R  60065133  60093378  60148682  60173878  60266360  60482987  60489382  60510582  60533088  60726322  60938414  60962122  61534555  61745109  368928  369949  370401  60129314  60215795  60252563  60278785  60482988  60528473  60726323  60726324  60938415  60962124  61076078  61109169  61163283  61534556  62080175  366522  367636  369950  371019  60165184  60278787  60482989  60489383  60528475  60533094  61026664  366540  60143366  60247712  60255065  60482990  60510577  60528477  60533104  369003  60255066  60482991  60528479  60533124  366523  368910  60255068  61020326  61104850  61114142  61151211  368766  61163286  61163286R  366524  61625238  00369477R  366525  60052915  60052915R  366059  369808  60123991  60175593  60292121  60321103  60351215  60405481  60473718  60510566  60674432  60787130  60868462  61067766  61264729  61294826  61523981  61729212  60123991R  368135  60143367  60405482  60787131  60879858  61067767  61533695  61748918  60754394  60905699  61041088  61076079  61533696  366421  366526  368911  369482  60165185  60182137  60255069  60292122  60337785  60962129  61105464  61228330  61516364  366060  368926  60165186  60202545  60241858  60266362  60611029  60646260  60787135  60938416  60962135  61005004  61163288  61264733  61294827  61375167  60165186R  368927  370399  60129315  60173879  60215796  60266363  60278788  60322347  60366283  60473719  60510561  60582072  60664611  60787141  60845773  60868464  60918560  60962137  60998289  61352932  61516365  61538714  61621829  61748919  369004  60136406  60136509  60252565  60420048  60489384  60510560  60664616  60972798  61286872  368912  60219047  60278790  60787146  60962155  61105466  61286873  61745914  62076944  366539  367638  369005  60189701  60266364  60944522  60962156  61020328  62076945  60266364R  368970  60787150  60874083  61352933  61361653  368971  371020  61309343  366066  366529  368929  61748920  369558  61314528  61314528R  366067  60095674  61246966  61326723  366068  368930  60118095  60219048  60403718  60473720  60528483  60674435  60962157  61027512  61294828  61309344  61506160  61681423  366069  60403719  60582074  60905698  61105467  61506161  366070  61309345  61506162  368931  60115498  60136510  60292123  60334059  60568321  60787178  60885166  61081609  61222988  61228331  00366071R  366072  60136511  60646261  60902827  61179736  61179737  61216918  61216920  61222989  61228332  61228333  61179736R  366073  60266365  60473721  60489385  60510559  60685081  60685083  61179740  61197962  61216921  61216922  61216923  61228334  61228335  368933  60119118  60168054  60266366  60685086  60685089  61081612  61197963  61197966  61198431  61198435  61216924  61216925  61216926  61216928  61222997  61223001  61228336  61228337  61228338  00368933R  60136512  60266367  60787182  60918568  61179750  61197969  61198437  61216929  61246055  61542602  62076935  366530  60646262  60787185  60820743  60902835  60918570  61197972  61198445  61216934  61542603  62080114  366076  366538  369007  369554  60820755  61223002  61264736  61264739  61375169  61565481  366537  368972  61105469  61163291  61542604  61163291R  366531  60231338  61353282  61353282R  366532  61309346  368935  61309347  60321105  60356350  60435452  60530511  60903913  60968746  61441995  61464983  61482340  60845775  61027515  61244920  61309348  61449272  61482341  61738441  61027517  61342272  61446259  61493897  61645758  61342272R  365862  368907  60116178  60129318  60154506  60206852  60225356  60266580  60325676  60366284  60373597  60387309  60393544  60405487  60420049  60445079  60452543  60510673  60551542  60598298  60619089  60685096  60763214  60763215  60787188  60803584  60885169  60938417  60972801  61020335  61027502  61086851  61092223  61103196  61169415  61264742  61341982  61357166  61459335  61511595  61542606  61596985  61635177  365863  60124023  60231284  60253443  60322349  60325677  60373598  60403720  60454913  60714913  60803585  61101547  61375171  61464382  60189710  60334061  60387313  60539728  60787189  61101549  61458224  62271934  60330804  60335489  60458145  61311464  62310748  60335489R  60458145R  365864  366423  60093665  60098401  60098402  60098403  60098405  60109720  60109721  60109722  60124024  60129319  60129320  60143353  60143355  60154507  60154508  60165682  60173816  60173817  60175540  60187162  60187163  60187164  60187165  60208590  60208591  60231285  60241931  60253456  60253458  60260492  60266584  60266586  60311275  60322352  60324518  60337788  60353499  60355851  60366285  60373603  60373604  60379909  60379910  60387319  60389514  60398925  60402361  60405490  60420050  60420051  60431510  60432600  60433696  60452544  60458146  60473722  60482403  60489386  60532270  60532271  60532272  60548745  60548747  60548748  60569272  60598287  60611038  60611039  60619098  60668241  60679344  60685066  60704366  60704367  60704368  60714914  60714915  60714916  60787190  60787191  60787192  60787193  60868468  60877489  60885171  60911343  60938419  60943339  60972802  60998290  61020337  61037487  61041089  61076083  61092237  61092239  61103197  61103198  61140177  61162619  61162620  61169416  61179752  61198454  61228339  61264745  61264746  61286877  61286878  61304472  61357171  61427113  61427114  61427115  61474332  61474333  61523581  61538715  61542781  61565482  61578708  61596986  61619259  61640888  61672132  61672134  61749009  61749012  62310747  61198454R  365911  369442  60129321  60175541  60314209  60322354  60373607  60379911  60402982  60420052  60539730  60598303  60726326  60932392  61037488  61041090  61294830  61474334  61493898  60314209R  60109723  60208592  60311276  60387321  60420053  60569284  60598304  60714917  60787195  60932393  60972803  61037489  61041091  61223003  61286879  61475572  61640889  61749017  60714917R  365866  60208593  60292128  60402983  60510690  60943340  60972804  60112039  60124025  60165683  60219025  60266588  60271118  60386431  60387324  60387325  60704370  60787196  60938421  61037490  61041093  61103199  61162621  61286880  61357172  61446407  61453113  61456386  61619260  61640890  61749027  62105419  60112039R  60154509  60231288  60370813  60402984  60420055  60548751  60763216  61092241  61359993  60154510  60388852  60389515  61309349  61551564  365867  60098406  60109724  60134446  60143356  60175542  60206853  60319383  60334063  60366287  60367586  60372179  60393545  60405493  60420056  60431511  60485415  60530513  60611043  60685102  60685104  60845778  60868471  60944523  61020339  61027507  61103201  61169417  61341983  61482163  61565483  60093666  60134447  60189711  60314210  60321108  60370814  60389516  60410044  60510696  60726325  60787197  61482164  61613171  369509  60187198  60402985  60519718  60744850  60807846  61309350  60807846R  365869  366541  371036  60586902  61005006  61374271  61631420  62230199  60586902R  365870  60093667  60109725  60161888  60208594  60321109  60367587  60402986  60458147  60523196  60619099  60787198  61027510  61169418  61476801  61590988  61640895  61729213  60109725R  60208594R  365871  370697  60116179  60165684  60413146  60539732  60674425  60787199  61005007  61027511  61476802  62271936  60165684R  365872  60539733  60815493  61476803  61631410  365873  60351218  60553727  60787008  61493899  61619261  62215597  369483  60175544  60295687  60337792  60393546  60420057  60669598  60807848  60844062  61516366  61748334  60165687  60337793  60420059  60787010  61027516  61584968  60165687R  365874  60231290  60754395  61027523  61516367  62271938  60187201  60674421  60890072  61027539  61580009  61619262",
      "more_code_info": ""
    }
  ]
}