      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1004-2014",
      "product_description": "Item codes starting with 001930    LAG SCR 12.7D X 55MM SHOR up to 155 MM, LAG SCR 12.7D X 55 MM SHO up to 100MM,  LAG SCR 12.7D X 55MM STAN up to 155 MM, LAG SCR 15.8D X 55MM LARG up to 155 MM",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "61484349  62124275  62157179  61077209  61337135  61484350  60555611  60747379  60874833  61142854  61142854R  60325552  60293083  60879683  61284330  61286943  60085527  60436025  60513856  60879684  61284331  61286947  60145167  60186855  60210921  60815940  60879685  60152929  60293084  60483144  60523344  60483145  60523345  60483145R  60440756  60483146  60523346  61142856  60483147  60523347  60536820  60483148  60523350  60483149  60523353  60483150  60523355  60536821  60483151  60483152  60523360  60483153  60523364  60483154  60523365  61067755  60483155  60523369  60132504  60165597  60206470  60293086  60387521  60452004  60488952  60572992  60653075  60671959  60812149  60984699  61004201  61040107  61092997  61099399  61104820  61142860  61224794  61283207  60127012  60140321  60152930  60208488  60265677  60281409  60293087  60323242  60425131  60452005  60488953  60523371  60561146  60653078  60671960  60812150  60848590  61002194  61002195  61026422  61026423  61026424  61083094  61092998  61099402  61142861  61144218  61195105  61195106  61002194R  61002195R  60096566  60118985  60127013  60137590  60165601  60172576  60234054  60254715  60264345  60317521  60385457  60393375  60436028  60483156  60488954  60513859  60538635  60572993  60576792  60585224  60621346  60653083  60671961  60714813  60736759  60848591  60848592  60973703  61004197  61004199  61083095  61092999  61099403  61142862  61144219  61195108  61275457  61616168  61737706  61774929  62116937  60127013R  60106616  60112849  60127014  60132507  60137591  60140322  60145168  60150159  60165604  60172578  60181949  60195046  60234055  60254716  60258448  60258449  60281411  60281412  60293088  60293089  60317522  60319984  60329109  60425132  60430657  60448001  60452006  60528727  60538637  60572994  60585225  60621347  60653085  60653088  60671962  60671963  60702010  60736760  60794263  60830601  60848593  60848594  60970191  60970192  60970193  61026416  61026417  61026418  61037609  61037611  61099405  61117601  61133098  61136511  61142863  61142864  61195109  61224796  61224797  61224798  61296709  61296710  61562050  61578287  61616170  61737707  61774932  62116938  60106617  60132509  60135161  60137592  60145169  60150161  60159613  60172579  60175630  60181950  60185158  60195047  60208489  60242275  60245531  60249740  60254717  60258450  60281413  60281414  60281415  60293090  60293091  60317523  60319985  60329111  60338353  60338354  60342154  60344023  60406106  60448002  60452007  60452008  60523375  60528729  60538638  60550137  60572995  60585226  60612818  60653095  60653098  60653100  60671964  60671965  60702011  60714814  60736761  60782054  60812151  60830602  60848595  60848596  60879687  60970194  60970195  61004191  61004192  61004193  61037621  61040109  61040110  61040111  61069134  61077215  61099407  61104822  61104823  61128194  61133110  61136512  61136514  61142866  61209058  61224799  61224801  61224802  61224803  61256752  61286949  61344678  61420618  61453204  61473804  61546143  61562051  61562052  61616171  61629788  61686789  61737708  61737709  61774940  61982443  62124277  62124278  60159613R  60714814R  60096568  60101358  60103446  60106618  60109515  60115148  60118986  60127018  60130184  60132511  60135162  60137593  60137594  60145170  60148169  60150162  60155297  60157578  60159614  60165606  60172581  60175631  60181951  60185159  60195048  60195054  60208490  60234056  60242277  60242278  60242279  60245532  60249741  60254718  60258451  60258452  60260176  60271179  60272607  60281416  60281418  60293093  60293094  60309941  60309942  60317524  60319986  60329112  60329113  60332476  60333601  60338355  60338356  60342155  60344024  60345905  60372728  60382493  60382494  60401257  60406107  60436029  60436030  60448003  60452009  60452010  60488955  60513864  60513866  60513869  60528730  60536823  60538639  60550138  60561149  60572996  60576794  60585227  60585228  60597788  60612821  60621349  60621350  60621351  60653104  60653107  60653109  60653110  60671966  60671967  60690730  60702012  60702013  60714815  60714816  60736763  60782055  60782056  60794245  60812152  60812153  60830603  60830604  60848597  60848598  60848599  60848602  60848603  60848604  60848605  60973707  60973708  60973709  61002201  61002202  61009648  61009649  61009650  61009651  61037629  61037630  61037631  61037632  61063239  61063240  61063242  61063243  61072850  61104824  61104825  61104826  61104827  61110039  61117606  61120996  61128195  61133111  61136517  61142878  61171568  61195112  61203643  61203644  61209059  61224804  61224805  61224806  61226637  61251141  61286951  61296712  61304095  61310086  61310088  61420620  61453205  61473805  61491348  61516715  61524256  61538699  61562053  61562054  61589576  61595820  61616172  61629790  61686791  61700055  61737710  61737711  61752960  61774941  61774942  62116939  62116940  62146188  62146190  62146192  60452010R  60094948  60101359  60101360  60101361  60106619  60111968  60112851  60115149  60127019  60130185  60132515  60135164  60137595  60137596  60140324  60145171  60150163  60152931  60155298  60157579  60159615  60165608  60172583  60175632  60181952  60185160  60186856  60195059  60195062  60195065  60208491  60230762  60230786  60230787  60234057  60242280  60242281  60245533  60245535  60249742  60254719  60258453  60260177  60265678  60271183  60272608  60281419  60281421  60293096  60293097  60293098  60309943  60309944  60309955  60309963  60317525  60319987  60323243  60329115  60329116  60332478  60332479  60333602  60333603  60338358  60342156  60342157  60344027  60345911  60372729  60401258  60406108  60422415  60423604  60430658  60436031  60436032  60452011  60452012  60513871  60513872  60513873  60528731  60536824  60536825  60550139  60550140  60572997  60572998  60585229  60585230  60585231  60585232  60612822  60621352  60621353  60653111  60653114  60653115  60653116  60653117  60671968  60671969  60671970  60702014  60702015  60702016  60714817  60736764  60782057  60782058  60782059  60812154  60812155  60830605  60848606  60848607  60848608  60848609  60973704  60973705  60973706  61002191  61002192  61009644  61009645  61009646  61009647  61037612  61037613  61037614  61037615  61037616  61040104  61040105  61040106  61063252  61063253  61069136  61072851  61077233  61079821  61104830  61117608  61117610  61117613  61128196  61133118  61133119  61136525  61136527  61142879  61144220  61168217  61168218  61171571  61179648  61195113  61195114  61203645  61209060  61224807  61251142  61256756  61275458  61275459  61284332  61284333  61286954  61304096  61304097  61326933  61420621  61421217  61421218  61453206  61473806  61491349  61496582  61516716  61542359  61555651  61562055  61565498  61565499  61565501  61589577  61589578  61686792  61686793  61718876  61718877  61737712  61737713  62002274  62137451  62137452  62146203  60344027R  60671970R  369229  369230  60094953  60099549  60101363  60108505  60109516  60111970  60112854  60123252  60127023  60130187  60132520  60135170  60137599  60140327  60148173  60152933  60154173  60157581  60159616  60165612  60172586  60175635  60181954  60185162  60195074  60195075  60208493  60215950  60222846  60222847  60230803  60230804  60230805  60242288  60245540  60245541  60245542  60249746  60254721  60258455  60260179  60272610  60281426  60293103  60309969  60309970  60317527  60319989  60323245  60329119  60329120  60332482  60332507  60333606  60333607  60338361  60338362  60342161  60342162  60344030  60345917  60372731  60406110  60422417  60423606  60430661  60436054  60436056  60479116  60483157  60488957  60488958  60513886  60528733  60536829  60536831  60550143  60573002  60573003  60585236  60585237  60597792  60612825  60612826  60621356  60653128  60653129  60653130  60653131  60653132  60671974  60671975  60702021  60702023  60714819  60714820  60736769  60782064  60782065  60782066  60812157  60830627  60879690  60931355  60931356  61002187  61002188  61002189  61037592  61037593  61037594  61037596  61040092  61040093  61040094  61040095  61063245  61063246  61069138  61077235  61077236  61110042  61110043  61110044  61110045  61133126  61133127  61136542  61136544  61168221  61168222  61179650  61195118  61203649  61203650  61224813  61224814  61226641  61256758  61284335  61286959  61304101  61310090  61378429  61396169  61420622  61421222  61453207  61473808  61491351  61496583  61510153  61524259  61565504  61565505  61589579  61589580  61737717  61737718  62142285  62142288  60272610R  61040093R  367050  367051  60096570  60099550  60103449  60111971  60112855  60115152  60127026  60132521  60135171  60140329  60150166  60152934  60172588  60175636  60181955  60185163  60195077  60195078  60208494  60230794  60230799  60230801  60245543  60245545  60249747  60254723  60260180  60265681  60271187  60272611  60281429  60281430  60309971  60317528  60319990  60329121  60332485  60333608  60342163  60344031  60345924  60382496  60422418  60423607  60430662  60436057  60436058  60488959  60513887  60513888  60536832  60550144  60573004  60585239  60612829  60621357  60653133  60653135  60653136  60653139  60671976  60690731  60702025  60714821  60714822  60736770  60781220  60782067  60812158  60830643  60879691  60925029  60925030  60925031  61002196  61002197  61009652  61009653  61037625  61037626  61037627  61037628  61040113  61040114  61077240  61104836  61104837  61117627  61128199  61133129  61136548  61168223  61168224  61176332  61176334  61179651  61195119  61203651  61224815  61224816  61226642  61226643  61274847  61286960  61296715  61378430  61420623  61453208  61491352  61516717  61546145  61565506  61578288  61595821  61629791  61666118  61686804  61700059  61737719  61752959  60088621  60099551  60112856  60115153  60127029  60135173  60148174  60150168  60152935  60165614  60172589  60181956  60185164  60242292  60254725  60265682  60271189  60272612  60323246  60323247  60332487  60342164  60344032  60345925  60372733  60455013  60472816  60488960  60513889  60536833  60550145  60573157  60653140  60671977  60702029  60714823  60736771  60782068  60812159  60970197  61004183  61037598  61037599  61037601  61040096  61040097  61110048  61117628  61121002  61133131  61136551  61136553  61144224  61195120  61251144  61275460  61286961  61378431  61420625  61491353  61565507  61595822  61686811  61700060  61752961  60550145R  60085984  60096571  60123254  60132523  60140330  60165616  60172591  60181957  60218943  60218944  60242294  60254728  60265683  60272613  60293106  60325559  60372734  60382497  60455014  60472817  60487352  60513890  60550146  60573162  60585245  60612830  60653142  60671978  60702031  60781219  60782069  60830648  60931242  60931243  61026419  61026420  61026421  61063251  61083099  61099409  61117629  61136556  61142887  61195121  61224818  61256761  61491354  61609885  61666126  61741397  60140330R  60115154  60135175  60172592  60208495  60281435  60329122  60523379  60573165  60612831  60690732  60781222  60895465  60970188  60973695  60973696  61037622  61037623  61093001  61099411  61117630  61142888  60573165R  60970188R  60085533  60107379  60172593  60208496  60323248  60455015  60550148  60653143  60690733  60812160  60970190  60973697  61026426  61026427  61093002  61099412  61117631  61117632  60690733R  60096572  60165617  60208497  60550149  60690734  60705049  60973701  61002193  61063237  61095774  61095775  61095777  61104839  60690734R  60152936  60265684  60702037  60781221  60984701  61004195  61093004  60159617  60258989  60467319  60523380  60781214  60812161  60920806  61037624  61093005  61104840  60159617R  60142909  60150169  60258990  60472821  60576795  60721965  60733781  60895438  60970189  61004196  61063133  61095748  61110049  60154174  60348766  60585246  60794237  60842559  60843730  60970196  60973699  61063250  60142910  60159618  60671979  60705050  60817012  60843731  60895436  60931237  60984698  61004187  61093008  61095793  61095796  61095799  61104841  61110051  60159618R  60671980  60781215  60897411  61110052  60496364  60560744  60615110  60799737  60902366  60115155  60150170  60203992  60348150  60585248  60781216  60794238  60902383  61878426  60150171  60208498  60348152  60513891  60536835  61142895  61382853  61500139  61382853R  60123257  60142911  60165618  60203230  60293107  60348153  60513893  60561151  60629948  60671981  60736772  60799738  60856589  60902384  60132525  60152937  60185165  60196504  60208501  60265685  60348154  60488962  60531323  60629949  60653145  60855065  60902385  61142896  61496584  60083102  60115156  60132527  60165620  60185166  60203231  60230809  60265686  60293109  60348155  60363738  60445170  60445171  60479117  60488964  60523382  60539407  60578622  60680440  60799739  60855074  60902386  61142898  61274848  61453209  60855074R  61274848R  60085534  60112858  60140331  60155299  60196505  60210922  60245566  60281439  60323249  60348157  60425133  60445172  60452015  60479118  60488965  60523383  60538642  60629951  60653146  60671982  60714825  60803570  60855075  61142901  61700061  61878382  62160329  60281439R  60714825R  60855075R  60076436  60108506  60115157  60142912  60175638  60186858  60203232  60203993  60230811  60254731  60348158  60363739  60425134  60445173  60445174  60487354  60488966  60513894  60523385  60528734  60612832  60653147  60671983  60714826  60803569  61037619  61040108  61067756  61142902  61195122  61275461  61700062  61775798  60096574  60135182  60154175  60186859  60208502  60242295  60281441  60323250  60385458  60425135  60445175  60487356  60488967  60523388  60539410  60585249  60629756  60671984  60736777  60803568  60879692  60879697  61142903  61184411  61195123  61200599  62046413  60879692R  60123261  60135184  60155301  60203994  60242296  60309973  60363740  60425136  60445176  60479119  60488968  60513896  60523390  60539411  60680441  60794244  60799740  61040116  61525198  62137457  60680441R  60154176  60186860  60210923  60323251  60365814  60393378  60442552  60445177  60509588  60536836  60539412  60727241  60799741  61040115  61063128  61063129  61063130  60115158  60127032  60150173  60186861  60235165  60323252  60509589  60714827  60781218  60984697  61117633  61484351  60323252R  60235166  61026425  61144226  62137456  60150174  60281444  60661882  60827545  61002198  61484352  60142913  60293111  61069139  61337136  61500141  60112861  60142914  60281446  60826967  60902395  61083101  61133146  61142904  60843732  60856772  61069140  61484353  60799742  61072853  61367601  61484354  61367601R  60902396  60028119  61484355  60028119R  60041351  61121003",
      "more_code_info": ""
    }
  ]
}