  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1003-2014",
      "product_description": "FREE-LOCK LAG SCREW, 12.7, FREE-LOCK LAG SCREW, 15.8, FREE-LOCK SUPRACOND TUBE and FREE-LOCK TUBE & SCP PLAT. Items starting with codes 0011810 and 0011811.",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "60879678  61278171  61625237  60564708  61026414  60564708R  60564710  60915347  61141512  61142759  60639474  60782037  62046409  60155294  60206463  62120939  60206463R  60108502  60152927  60206464  60862361  60862362  60862361R  60142899  60165580  60206466  60487350  60671954  60736748  61491355  62120941  60208478  60815946  61278172  61502705  60142901  60172569  60281392  60538633  60705043  60736749  60830649  61026415  60162132  60639475  60781210  61110037  60782040  60782041  61345993  61399023  61045321  61400219  60817011  61268917  60782042  60931208  61400222  61400223  61393376  60782043  61141513  61278174  61284322  61026411  61286935  60106612  60572988  61286938  61491356  60071404  60099545  60107375  60218931  60218932  60653031  60782044  61026409  61195093  61420626  61502706  61616173  60080046  60109509  60115145  60126996  60154170  60195018  60208481  60234047  60275162  60472815  60536817  60653032  60781212  60879681  61026406  61141514  61195094  61274845  61284324  61345995  61629793  61700052  60111966  60112840  60123244  60126997  60142902  60165583  60175627  60189035  60190841  60218933  60218934  60222827  60222828  60234048  60275163  60281396  60290741  60293063  60323238  60653043  60701995  60705044  60782045  60812141  60862367  61037587  61037588  61141515  61195095  61209053  61251129  61278175  61284325  61420628  61466805  61614765  61663369  61278175R  60097895  60103441  60106613  60112843  60123245  60126998  60132497  60142903  60154171  60157576  60172572  60181946  60186847  60195020  60208482  60215946  60222829  60222830  60234049  60242264  60242265  60254701  60275164  60281399  60290742  60293064  60293065  60323239  60572989  60621339  60653053  60671955  60701998  60736750  60782046  60812142  60931209  61026403  61026404  61077251  61110038  61120992  61141516  61171562  61184409  61195096  61209054  61251130  61428485  61534196  61609887  61644214  61663371  61195096R  60088618  60097896  60099546  60103442  60107376  60107377  60123246  60126999  60135159  60142904  60150153  60155295  60165584  60172573  60186849  60195021  60208483  60215947  60222831  60222832  60234050  60236419  60242267  60242269  60254706  60275165  60281401  60290743  60293066  60323240  60381062  60401298  60572990  60621340  60621341  60653057  60671956  60702002  60705045  60782047  60812143  60830650  60862366  60931212  60931213  61037589  61037590  61069128  61083091  61142760  61171564  61184407  61195097  61209055  61420629  61466807  61589581  61762401  62106371  62197640  60186849R  60097897  60103444  60111967  60115146  60123247  60127001  60140305  60142905  60150155  60165585  60181947  60186850  60195022  60208485  60215948  60234051  60236421  60242270  60242271  60254707  60275166  60281404  60290744  60293067  60381063  60401299  60621342  60653063  60671957  60705046  60736752  60782048  60812144  60862369  61026401  61026402  61037586  61069129  61083092  61141517  61171565  61184408  61195098  61209056  61278176  61405652  61466808  61663373  61762403  62197642  60705046R  60097898  60106614  60109511  60112844  60123248  60127002  60135160  60150156  60165586  60181948  60186851  60195023  60222833  60222844  60234052  60264008  60281402  60290745  60323241  60452002  60513841  60536818  60550134  60621343  60653066  60671958  60714808  60782049  60812145  60931215  61037591  61063230  61063232  61069130  61083093  61142761  61195099  61251131  61275452  61284326  61379219  61420631  61466814  61609889  61663375  61762486  60106615  60112846  60123249  60142906  60154172  60175628  60206468  60254708  60264010  60452003  60550135  60653069  60702006  60781213  60830652  60931211  61026407  61069131  61142762  61195101  61286940  61405653  61428488  61663376  60123250  60142907  60186852  60218935  60218936  60234053  60275167  60621345  60714809  60973693  60973694  61077253  61179647  61184410  61195102  61466840  60137579  60208486  60680439  60705047  61026405  61284327  61466842  62050378  60782050  61120993  61284328  61762492  61026399  60781209  61310083  60812146  60812147  60812148  60879682  61114907  61769257  60781211  60830670  62124273  60210917  60186853  60203228  60210918  61209057  61275454  60108503  60186854  60203229  60436016  60862368  60155296  60208487  60210919  61141518  61142763  61184405  61195103  60169718  60203991  60210920  60203991R  60572991  60705048  61184406  61195104  60218937  61769259  60714810  61069132  61749240  60354631  60736754  60736756  60736758  367595  368998  60098363  60123970  365697  367594  368901  369490  60136409  61169419  60351665  61493555  61603996  365698  368902  369535  60098368  60136410  60154522  60175505  60202532  60337741  60405494  60435453  60467561  60674428  365694  368995  60147698  60206854  60714885  60735261  60735263  365696  367599  369566  60182071  61546075  60182071R  366705  367603  60123971  60175506  60206857  60253361  60370811  60398926  60420066  60473723  60569302  60714897  60803586  60885175  61048115  61162614  61264748  61286923  61375175  61453111  61464379  61578712  61621833  61714071  60473723R  367602  369849  60093380  60131307  60143342  60154523  60165720  60172660  60202536  60241926  60307972  60322322  60337742  60351219  60379925  60393547  60416012  60432631  60433697  60532279  60598305  60625536  60685115  60803587  60902841  60969382  61092247  61103202  61169420  61286925  61294831  61512406  61585377  61635176  61761524  365673  367604  369438  369624  60098372  60098373  60129322  60129323  60136411  60136412  60165721  60165722  60175507  60197183  60197184  60215775  60218978  60218979  60253362  60322324  60330803  60337743  60351220  60366262  60379926  60393548  60402988  60405495  60420070  60445080  60452479  60473724  60510737  60539742  60582142  60625542  60714872  60714875  60714877  60803588  60803589  60803590  60988402  61092249  61103203  61162615  61198455  61264749  61359994  61453112  61464380  61565484  61621834  61752594  369478  369684  60093381  60136413  60163976  60241927  60473725  60510744  60563187  60803591  61294832  61752596  365686  367605  369507  60260485  60803592  365687  368904  60314207  60416013  60668258  60962159  61934147  60962159R  368915  61534188  61488452  61030739  61101550  365685  367601  369546  60109726  60161889  60202537  60452480  60494321  60625544  60803593  60944525  61529230  60107395  60108185  60107395R  365691  366416  367628  369508  369756  60098374  60109727  60129324  60165723  60206855  60208526  60253369  60266524  60322327  60337744  60348274  60405496  60416014  60452481  60485416  60532277  60532278  60598310  60657046  60657047  60807853  60807854  60874093  60911345  60944551  60988403  61037492  61092258  61101551  61103204  61198456  61304496  61379890  61402004  61457772  61565485  61621838  61700444  61752599  60807853R  365670  367627  368886  369479  369861  60082872  60093382  60098375  60098376  60098377  60098378  60109728  60109729  60109730  60123972  60129325  60129326  60136414  60136415  60143343  60144957  60144958  60154524  60165724  60165725  60172661  60175508  60182083  60182086  60187206  60187208  60187210  60208527  60208528  60218980  60218981  60231385  60231386  60253370  60260486  60266525  60266526  60337745  60351221  60359550  60366263  60379920  60384016  60393549  60405497  60405498  60410018  60416015  60420072  60424763  60445081  60452494  60473727  60473730  60489389  60494322  60510745  60532274  60532275  60532276  60569304  60582094  60582097  60619105  60624518  60657049  60657051  60657052  60763218  60763219  60763220  60763221  60820807  60820809  60820810  60885177  60902842  60911346  60931164  60938423  60968739  60972805  60988404  61020343  61027552  61048116  61076085  61092273  61101552  61103206  61108242  61198457  61209572  61264750  61294833  61352934  61359995  61377328  61379891  61402005  61420343  61437988  61440688  61542608  61558660  61590989  61621842  61640885  61719308  61719312  61719314  61753041  61753042  60208527R  365671  366417  367626  368815  369547  369757  60034670  60034674  60072680  60093383  60093384  60093385  60093386  60093387  60093388  60093389  60093390  60098379  60098380  60098381  60098382  60108186  60108187  60108189  60108190  60108191  60109731  60109732  60112055  60112056  60112058  60112060  60112062  60112063  60114621  60114622  60116183  60123974  60123975  60129327  60129328  60129329  60129330  60129331  60129332  60129333  60129334  60129335  60129336  60143344  60143345  60147699  60147701  60154525  60163977  60163978  60167251  60167254  60172662  60172663  60172664  60175509  60175510  60175511  60175512  60175513  60175514  60182088  60182090  60187212  60187214  60187215  60187217  60197185  60197186  60197187  60197188  60202538  60202539  60202540  60208529  60208530  60208531  60208532  60208533  60218982  60218983  60218984  60218996  60218999  60219001  60231388  60253371  60260488  60260489  60260490  60266527  60266528  60266529  60266530  60266531  60266532  60337746  60337747  60351222  60353670  60356352  60357422  60359551  60366264  60366265  60366266  60367589  60379921  60379922  60384017  60384018  60393551  60393552  60393553  60394655  60401177  60405499  60405501  60405502  60406633  60416016  60420077  60420078  60420080  60424765  60431513  60432636  60433698  60433699  60452482  60452483  60452484  60452493  60473732  60473733  60476656  60482404  60489390  60494325  60510746  60510747  60510748  60532280  60532281  60532282  60532283  60532284  60532285  60532286  60557400  60557401  60569306  60569307  60582089  60582090  60582098  60598307  60598312  60598313  60619107  60619109  60619879  60624520  60657071  60657072  60657073  60657075  60657077  60657078  60657079  60714890  60714891  60714892  60714893  60714894  60714895  60714896  60763223  60763224  60763225  60763226  60763227  60763228  60763229  60763230  60807855  60807856  60807857  60807858  60807859  60807861  60868474  60874097  60877491  60885178  60885179  60902843  60911348  60911349  60931165  60932394  60938427  60944574  60959978  60968742  60972806  60998291  61020344  61027553  61030740  61048117  61076086  61079834  61086854  61092280  61092281  61101553  61103207  61108243  61108244  61162616  61162617  61179754  61179757  61198458  61209573  61209574  61246085  61264751  61264752  61341158  61352935  61359996  61377330  61379892  61402007  61408293  61412173  61420344  61436467  61437989  61440257  61440689  61457773  61464381  61542610  61558661  61584969  61621843  61640886  61672121  61682542  61682544  61682545  61682546  61682547  365690  367630  368816  369943  60098383  60098384  60109733  60109734  60123976  60123977  60147702  60163979  60172665  60175515  60175516  60197189  60215777  60322328  60337748  60351223  60379923  60384019  60393554  60405503  60420085  60432639  60433701  60459389  60532288  60582133  60582134  60657083  60807865  60885180  60931166  61030741  61037493  61048118  61101554  61294834  61440690  365689  367622  368905  369952  60093391  60129337  60131308  60161890  60253373  60366268  60389517  60402989  60405504  60416017  60435454  60452495  60476672  60532290  60611045  60657087  60657088  60807868  61294835  61565486  61753043  62271890  60657087R  368817  369655  370960  60167255  60206856  60582112  60874101  365692  368818  369953  60074350  60165726  60410029  60452496  60553405  60885181  368819  369954  370922  60175518  60402990  60668263  368820  60730290  61406970  365701  367625  369944  60098385  60109735  60116184  60165727  60197192  60266533  60393555  60405505  60406636  60445082  60467560  60489391  60539743  60646266  60763231  60911350  61086857  61162618  61264753  61264754  61523859  61538716  61565487  61753045  365693  367620  369937  60177352  60489392  60082877  60177353  60241928  60528660  60598308  60988405  61114143  61286928  61375177  61590993  60598308R  365672  368821  369491  60109736  60136416  60154526  60172666  60197193  60241929  60322331  60337749  60393556  60452497  60473734  60510749  60646267  60763232  60931167  60988407  61048119  61198459  61463554  61565488  61756193  365702  368822  369439  369758  60034675  60034676  60098386  60107396  60107397  60108193  60108195  60123979  60154527  60167256  60167258  60175519  60187219  60187228  60197195  60231390  60253377  60266534  60337750  60351224  60366269  60379924  60393557  60401178  60432644  60433702  60433703  60510750  60532293  60598309  60625545  60674430  60674431  60714850  60714851  60820796  60820798  60820801  60931168  60968744  60988408  61030742  61048120  61079836  61092283  61103208  61140178  61198460  61286930  61352936  61463555  61473803  61542611  61565489  61752602  61990140  365681  368823  369489  60119109  60182093  60253378  60539746  60815017  61463557  61613172  365678  368887  369685  60108196  60165728  60260491  60714886  368824  369526  368825  370001  365688  368997  61075747  368808  369480  60098387  60147704  60175520  60868477  60944606  61746531  61753040  365675  368888  369527  60136417  365679  60231391  61269949  61602063  365676  368877  369492  60109737  60143346  60144959  60322332  60337751  60435456  60452498  60473735  60572731  60868478  61037494  61198461  61682532  60109737R  60337751R  368889  369548  60107398  60108197  60167261  60182098  60215780  60359552  60393558  60405506  60452499  60494335  60510768  60611049  60763233  60763234  61198462  61523582  61590994  61640887  365674  368826  369528  369625  60154528  60241930  60668304  60885182  61902191  365684  368827  369539  60165729  60841203  368765  369459  60997531  365680  368878  61527963  368906  61114144  368879  369951  60165730  60668305  60885184  369481  61288126  61602064  61628735  60905697  61115872  61118976  365704  368996  60154529  60253379  60602194  60815019  365700  368880  369820  60116185  60165731  60197196  60962160  60998292  61269944  61717286  365703  368136  368881  369549  369956  60136418  60885185  61048121  61326116  368882  369457  60147706  61720251  368883  369550  369957  60175521  61027583  61086858  61782207  368884  369460  60175522  60175522R  368958  369836  60175523  368959  61115873  61118977  368885  369493  61264756  365695  60093392  60098388",
      "more_code_info": ""
    }
  ]
}