  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64708",
      "recalling_firm": "Philips Healthcare Inc.",
      "address_1": "3000 Minuteman Road",
      "address_2": "N/A",
      "postal_code": "01810",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution including USA (nationwide)",
      "recall_number": "Z-1003-2013",
      "product_description": "Philips Healthcare Computed Tomography X-Ray System.    These devices are whole-body computed tomography (CT) x-ray  systems or sub-systems, each with a continuously rotating x-ray tube  and multi-row detectors enclosed by a gantry. X-ray transmission  data acquired and taken at different angles can be reconstructed into  cross-sectional images. Each device also includes signal analysis  and display equipment, patient and equipment supports, components, and accessories.",
      "product_quantity": "2089 US  4036 ROW",
      "reason_for_recall": "Philips Healthcare discovered  the customer was performing CTDI measurements for facility accreditation and received results that were 15 to 25% higher than Philips protocol parameters stated.",
      "recall_initiation_date": "20100405",
      "center_classification_date": "20130417",
      "termination_date": "20210308",
      "report_date": "20130424",
      "code_info": "Gemini Dual, Gemini 16 Power, Gemini GXL, Gemini LXL,  Gemini TF 16, Gemini TF Base/Ready, Gemini TF 64,  Gemini TF Big Bore, Precedence SPECT/CT, Brilliance CT (6, 10,16, 16P, 40, 64, and Big Bore), Brilliance iCT, Brilliance iCT SP"
    }
  ]
}