      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1002-2014",
      "product_description": "HERBERT MINI BONE SCREW 2 and HERBERT/WHIPPLE C ANN BONE. Items starting with 0011500 and 0011520.",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "60135114  60152390  60235174  60253671  60303868  60318915  60406045  60469637  60538425  60629952  60708626  60713650  60845570  60885012  61156076  61290771  60097888  60126180  60155283  60235175  60270647  60278995  60393280  60425089  60472771  60538426  60555772  60557775  60588046  60591903  60603145  60715075  60845571  60885014  61316528  61616158  61749182  60111959  60140238  60162141  60175614  60208384  60218727  60278996  60285802  60291544  60406046  60435612  60487168  60550156  60555773  60557774  60576766  60591904  60603146  60859714  60885015  61075639  61103192  61166254  61290772  61650519  61750812  60093352  60101703  60115126  60140239  60141562  60186799  60195341  60253673  60265299  60270648  60291547  60380281  60393281  60406047  60472772  60478980  60536890  60555774  60557776  60588047  60591905  60708627  60845573  60845574  60859715  61019223  61103193  61117518  61144243  61434451  61473776  61635184  61650520  61750818  60094821  60115127  60123222  60126182  60135115  60142528  60169707  60189030  60195345  60208386  60222271  60231248  60240379  60265301  60278998  60318916  60323216  60380282  60422419  60429154  60435617  60451941  60536891  60550158  60557777  60576768  60603147  60603150  60671881  60708628  60756056  60756071  60756072  60788112  60845575  60845576  60859716  60885016  61080555  61103194  61141227  61156079  61266929  61287003  61420555  61434452  61487520  61564719  61640830  61660576  61729755  61753841  60101704  60123223  60142529  60169708  60195352  60195392  60195393  60195394  60195395  60291561  60367373  60401265  60435620  60472773  60514749  60557778  60576769  60588048  60598747  60708630  60845577  60845578  60868431  61019224  61061013  61076451  61080556  61144244  61224620  61290773  61538664  61729756  62304810  60097889  60140240  60165147  60195353  60253674  60270649  60278999  60291562  60367380  60435623  60491779  60550161  60671882  60709315  60713651  60845579  60859730  61076452  61111263  61166257  61179536  61206335  61316530  61487522  61616159  60152391  60190823  60231245  60240380  60285801  60323217  60406048  60435626  60478981  60514751  60538427  60598749  60603151  60708631  60788113  60845581  60885017  61019225  61061014  61075640  61076453  61111264  61166259  61206336  61245932  61276825  61276826  60123224  60130226  60140241  60154142  60167096  60185142  60195373  60208388  60231246  60244026  60265302  60363741  60393282  60422420  60538428  60591906  60629953  60652951  60699585  60735906  60756996  60800729  60807190  60885018  60930482  60938158  60944473  61025864  61033142  61080557  61087333  61129845  61141229  61144245  61185817  61395114  61457379  61538668  61729143  61774728  60112824  60130227  60135116  60141563  60155284  60162143  60181935  60195374  60218728  60231247  60244027  60265303  60303871  60325537  60363743  60380283  60396394  60422421  60448279  60550163  60585148  60591907  60652952  60699586  60735907  60756997  60756998  60800730  60829352  60845582  60911367  61033143  61040086  61091802  61144246  61156086  61185816  61282029  61287021  61378418  61380162  61437190  61457380  61484342  61551151  61578282  61616160  61635185  61729757  62258979  60105510  60111960  60127519  60132464  60139614  60141564  60159621  60172478  60179262  60190824  60195375  60195377  60235176  60265304  60265305  60325539  60340792  60363744  60367382  60378417  60391510  60435632  60448281  60501144  60501145  60528750  60536892  60538429  60557779  60563845  60589761  60591908  60640214  60652954  60699587  60714748  60735908  60756999  60800731  60807191  60829353  60829354  60845583  60859805  60885021  60911368  61067739  61091803  61141231  61144247  61156087  61206337  61224629  61282030  61287022  61357125  61378419  61425832  61438777  61451143  61484343  61523640  61578283  61602051  61640831  61728778  61728779  61774746  62040567  62271582  60105512  60123225  60130228  60132465  60135117  60139615  60141565  60145078  60152392  60165148  60172479  60179263  60186801  60195384  60195386  60195387  60195389  60201205  60235177  60235178  60244028  60253675  60303872  60308469  60325538  60340793  60367383  60378419  60387523  60391511  60406049  60422423  60435635  60448280  60487169  60491780  60501146  60501147  60550164  60585152  60589763  60591909  60629955  60640215  60652955  60690621  60699588  60714749  60735909  60757001  60757002  60800732  60807192  60829355  60829356  60829357  60845584  60859806  60885023  60885024  60911369  61087336  61091804  61141233  61141240  61179537  61206338  61233880  61245933  61256430  61276828  61276829  61354689  61378420  61380164  61414320  61425834  61439799  61457381  61465678  61484344  61523641  61567404  61578284  61589558  61616161  61653954  61713682  61729144  61750820  61754463  61774744  61804521  61804523  62033047  62033048  62160887  60094823  60101705  60105513  60123226  60126185  60130229  60135118  60139616  60145079  60150046  60152394  60159622  60165149  60172480  60179264  60186802  60208390  60208393  60210897  60218729  60231249  60235179  60235180  60244029  60253677  60265307  60265308  60285804  60303873  60303874  60308470  60325540  60340794  60351959  60367384  60367385  60378422  60378426  60391512  60406050  60422424  60422425  60451942  60472774  60487170  60491781  60501148  60501149  60528751  60536895  60536896  60550165  60557786  60561933  60563847  60566846  60589764  60591914  60640216  60652956  60690622  60699589  60714750  60714751  60735910  60757003  60757005  60807194  60807195  60810888  60829358  60829359  60845585  60859807  60868432  60885025  60885026  60911371  61025865  61033144  61060982  61067740  61080558  61087340  61129850  61129851  61141253  61144248  61149932  61179539  61200590  61224630  61245934  61256431  61262510  61282031  61287023  61354690  61357126  61378421  61414324  61425836  61439801  61451144  61457382  61465679  61496573  61523642  61542647  61567405  61635187  61640833  61650522  61681023  61707779  61729760  61729761  61777803  61804526  60070210  60105515  60123227  60127520  60132467  60135119  60139617  60140242  60141567  60145080  60152395  60159623  60172481  60179265  60185143  60186803  60201206  60208394  60208395  60218730  60231250  60231251  60235182  60244030  60253679  60253680  60265309  60265310  60285805  60303875  60308471  60325541  60340795  60340796  60340797  60351960  60367386  60378453  60378454  60391513  60406051  60422426  60435637  60451943  60472775  60478982  60491782  60501153  60501155  60528753  60536897  60536901  60538431  60557787  60561934  60563848  60585156  60589765  60591916  60640217  60690623  60699591  60714752  60735911  60757006  60757007  60800737  60800740  60800744  60800746  60800747  60810889  60829360  60829361  60845586  60859808  60859816  60859817  60885027  60885028  60885029  61024793  61025866  61060983  61072081  61080559  61091805  61135980  61141254  61144249  61144252  61200591  61224631  61233882  61262511  61282032  61290774  61316543  61357127  61378422  61395115  61396161  61421213  61438778  61439802  61457383  61484345  61513883  61523643  61542349  61542648  61589559  61616162  61653955  61685399  61729139  61729140  61777804  61804527  62304864  60094825  60111961  60115128  60123228  60126186  60135120  60142530  60145081  60150048  60155286  60159624  60172482  60179266  60186804  60195396  60195405  60195407  60195408  60195409  60218731  60231252  60231253  60244031  60253682  60265311  60265312  60285806  60303876  60308472  60325542  60340798  60340800  60340854  60367387  60367388  60380284  60396395  60396396  60448282  60451944  60472776  60478983  60491783  60501159  60501164  60528755  60536902  60536903  60550166  60557789  60561936  60563849  60566847  60589773  60640218  60652957  60690624  60699593  60714753  60714754  60735912  60757008  60757009  60800751  60800752  60807196  60829362  60829363  60845587  60859819  60859820  60885030  60885031  60885032  60911372  61033146  61060984  61067741  61080560  61120674  61141256  61141257  61141258  61144253  61144254  61179540  61200592  61233883  61238542  61245935  61245941  61305258  61316544  61359713  61378423  61380165  61425840  61437192  61451145  61465681  61513888  61523644  61531071  61542649  61555649  61564721  61589560  61616163  61620160  61653957  61713683  61729141  61750822  61754465  62271585  60094826  60112825  60123229  60126188  60132468  60135121  60139618  60140243  60159625  60172483  60185144  60186805  60195398  60208397  60218732  60231254  60244032  60253684  60265313  60285808  60303877  60325543  60340802  60351962  60367389  60380285  60396397  60401266  60406053  60435640  60469640  60472777  60491784  60514753  60536904  60536905  60566848  60566849  60629956  60640219  60652958  60690625  60699594  60735913  60757010  60800753  60800754  60807197  60829364  60845588  60859822  60868433  60885033  60911374  60911375  61025867  61033147  61075641  61087343  61091807  61120675  61135992  61144255  61156088  61238543  61262512  61282033  61290775  61354691  61357128  61378424  61425843  61465682  61513890  61523645  61531073  61589561  61640835  61685401  61707787  61729769  61754466  60127521  60130230  60135122  60139621  60141568  60169710  60185145  60186806  60208399  60218733  60231255  60244033  60265314  60285809  60308473  60340803  60351964  60380286  60396398  60406054  60422429  60435643  60455018  60472778  60478984  60514754  60538432  60561201  60585159  60591917  60652960  60690626  60699595  60735914  60757011  60757012  60807198  60829366  60845589  60859823  60885035  61024792  61025868  61067742  61080561  61120676  61144256  61144257  61224632  61270323  61276830  61378425  61380167  61437193  61457384  61496574  61602052  61640836  61729142  61729770  62296468  60092666  60112827  60123230  60126189  60130231  60135123  60145082  60154143  60159626  60167098  60181936  60185146  60186807  60201207  60218734  60231256  60244034  60265315  60265316  60308474  60325544  60340804  60340805  60367390  60380288  60396399  60422430  60422431  60478985  60487171  60514755  60536906  60566850  60566851  60629957  60640220  60652963  60699596  60714755  60735915  60757013  60800761  60807199  60810890  60829367  60845590  60859831  60868438  60885036  60944475  61024790  61024791  61033148  61075642  61080562  61120677  61144258  61144259  61156089  61200593  61262513  61276831  61276832  61395116  61425844  61437195  61484346  61513891  61523646  61602053  61616165  61653960  61681022  61728790  61728792  61750824  61750825  62258986",
      "more_code_info": ""
    }
  ]
}