{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Foothill Ranch",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80520",
      "recalling_firm": "Ossur Americas",
      "address_1": "27051 Towne Centre Dr",
      "address_2": "N/A",
      "postal_code": "92610-2804",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA;    Foreign: Australia, Germany, New Zealand, Belgium",
      "recall_number": "Z-1001-2019",
      "product_description": "Ossur, Sterile,    Part/ Description:   505300T/ CB RING KIT TI SM;  505300V2/ OB V2 RING KIT TI SM;  505400T/ CB RING KIT TI STD;  505400V2/ OB V2 RING KIT TI STD;  505500T/ CB RING KIT TI LG;  505500V2/ OB V2 RING KIT TI LG;  515300V2/ OB V2 RING KIT TI SM-No App;  515400V2/ OB V2 RING KIT TI STD-No App;  515500T/ CB RING KIT TI LG-No App;  515500V2/ OB V2 RING KIT TI LG-No App;  515300T/ CB RING KIT TI SM-No App;  515400T/ CB RING KIT TI STD-No App;  515301T/ CB RING KIT TI SM 8-No APP;  515301V2/ OB V2 RING KIT TI SM 8-No App;  515401T/ CB RING KIT TI STD 8-No APP;  515401V2/ OB V2 RING KIT TI STD 8-No APP;  515501T/ CB RING KIT TI LG 8-No App;  515501V2/ OB V2 RING KIT TI LG 8-No App;  540T/ STERILE TITANIUM COMP TRAY;  540V2/ STERILE V2 COMPONENT TRAY",
      "product_quantity": "182",
      "reason_for_recall": "Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.",
      "recall_initiation_date": "20180621",
      "center_classification_date": "20190311",
      "termination_date": "20200904",
      "report_date": "20190320",
      "code_info": "All lots of the products manufactured from 06/30/2009 - 12/13/2018"
    }
  ]
}