  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Malvern",
      "address_1": "51 Valley Stream Pkwy",
      "reason_for_recall": "The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed.  Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor.",
      "address_2": "",
      "product_quantity": "81",
      "code_info": "Model numbers 10094135, 10094137, 10094139, 10094141 and 10280959.",
      "center_classification_date": "20130326",
      "distribution_pattern": "Nationwide Distribution including  AL, CA, CO, FL, IA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, ND, NJ, NY, OH, PA, SC, TX, UT, VA, WA, and WI and to Puerto Rico.",
      "state": "PA",
      "product_description": "Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed.    Angiographic x-ray system",
      "report_date": "20130403",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-0997-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64662",
      "termination_date": "20141106",
      "more_code_info": "",
      "recall_initiation_date": "20130308",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}