{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89915",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, IL, IN, KY, MD, MO, ND, NH, NY, OH, OR, PA, SD, TX, VA and the countries of Argentina, Austria, Bahrain, Belgium, Brazil, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Switzerland, United Kingdom.",
      "recall_number": "Z-0996-2022",
      "product_description": "syngo.via RT Image Suite  upgraded to VB60 running on SOMATOM go.Sim Model Number 11061660. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.",
      "product_quantity": "15 (US)",
      "reason_for_recall": "Upgraded software version makes an automatic change in laser configuration settings for \"markerless workflow\" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.",
      "recall_initiation_date": "20220307",
      "center_classification_date": "20220422",
      "termination_date": "20231219",
      "report_date": "20220504",
      "code_info": "Partial UDI: 04056869249230, Serial Numbers 128123, 128106, 128109, 128144, 128124, 128112, 128132, 128135, 128114, 128125, 128113, 128142, 128107, 128136, 128133",
      "more_code_info": ""
    }
  ]
}