  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mentor",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82035",
      "recalling_firm": "Steris Corporation",
      "address_1": "5960 Heisley Rd",
      "address_2": "N/A",
      "postal_code": "44060-1834",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed to the following US states:  AL, AR, CA, CO, FL, GA, IL, LA, MA, MD, ME, MI, NC, NJ, NM, NY, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV.    The products were distributed to the following foreign countries:  Bahrain, Brazil, Canada, China, Saudi Arabia, South Korea, and Taiwan.",
      "recall_number": "Z-0996-2019",
      "product_description": "AMSCO 5000 Series Washer/Disinfector Model # 5052",
      "product_quantity": "68",
      "reason_for_recall": "The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.",
      "recall_initiation_date": "20181220",
      "center_classification_date": "20190308",
      "termination_date": "20190722",
      "report_date": "20190320",
      "code_info": "Serial Numbers 3627418005  3601518006  3612718013  3633817009  3632318011  3626718014  3625114008  3600515028  3600515029  3600515030  3628516022  3626718013  3631618008  3627418007  3624514006  3628818001  3608518007  3634517006  3630218008  3605316016  3600917004  3628218007  3628818003  3629518009  3626718009  3626718011  3627418001  3627418002  3630218012  3630218013  3630918013  3632318012  3626718016  3634517005  3631414004  3603618005  3607217008  3630318013  3626817008  3628317012  3616615026  3630918014  3633018006  3628218009  3632318009  3628218011  3623615011A  3601518012  3612817007  3627418009  3628218006  3627418008  3631618007  3617315011  3614818009  3624017008  3629215012  3614516007  3628218010  3610216025  3634313010  3611715016  3628818005  3615617008  3624017009  3616516020  3627418004  3627418006"
    }
  ]
}