{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mentor",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82035",
      "recalling_firm": "Steris Corporation",
      "address_1": "5960 Heisley Rd",
      "address_2": "N/A",
      "postal_code": "44060-1834",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed to the following US states:  AL, AR, CA, CO, FL, GA, IL, LA, MA, MD, ME, MI, NC, NJ, NM, NY, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV.    The products were distributed to the following foreign countries:  Bahrain, Brazil, Canada, China, Saudi Arabia, South Korea, and Taiwan.",
      "recall_number": "Z-0995-2019",
      "product_description": "AMSCO 3000 Series Washer/Disinfector Model # 3052",
      "product_quantity": "11",
      "reason_for_recall": "The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.",
      "recall_initiation_date": "20181220",
      "center_classification_date": "20190308",
      "termination_date": "20190722",
      "report_date": "20190320",
      "code_info": "Serial Numbers 3602014006  3631618006  3628818009  3629518008  3633018010  3626718010  3630918008  3630918010  3617717013  3620016019  3623516006"
    }
  ]
}