  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Saginaw",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67076",
      "recalling_firm": "The Standing Company",
      "address_1": "5848 Dixie Hwy",
      "address_2": "",
      "postal_code": "48601-5967",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including AL, CA, DE, IA, IL , IN, KS, KY, MI, MO, NJ, OH, OK, TX, WA, WI,and  WV.",
      "recall_number": "Z-0993-2014",
      "product_description": "Superstand stand-up wheelchair  HPS-2    A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.",
      "product_quantity": "approximately 119",
      "reason_for_recall": "Power Superstand Standing Wheel chair, flex shaft (the flexible wand which supports the joystick control module) may fracture if it is bent repeatedly and /or beyond 90 degrees. A newly designed Flexshaft which corrects this problem is available at no charge.",
      "recall_initiation_date": "20130516",
      "center_classification_date": "20140219",
      "termination_date": "20140625",
      "report_date": "20140226",
      "code_info": "Serial No: 1306, 1313, 1342, 1349, 1362, 1358, 1359, 1392, 1391, 1406, 1402, 1407, 1426, 1441, 1451, 1018, 1453, 1399, 1523, 1477, 1480, 1482, 1497, 1494, 1505, 1516, 1533, 1537, 1536, 1535, 1542, 1544, 1549, 1543, 1550, 1562, 1563, 1566, 1580, 1572, 1573, 1574, 1571, 1578, 1583, 1641, 1593, 1591, 1592, 1599, 1604, 1600, 1615, 1616, 1614, 1622, 1631, 1637, 1643, 1635, 1640, 1634, 1648, 1666, 1652, 1657, 1663, 1664, 1676, 1689, 1665, 1683, 1691, 1688, 1701, 1703, 1711, 1698, 1712, 1719, 1728, 1731, 1726, 1740, 2165, 2172, 2175, 2177, 2257, 2182, 2197, 2205, 2258, 2259, 2262, 2270, 2283, 2279, 2281, 2298, 2306, 2310, 2312, 2330, 2334, 2503, 2515, 2328, 2514, 2523, 2525, 2550, 2561, 2568, 2581, 2573, 1684, and 1427.",
      "more_code_info": ""
    }
  ]
}