  "results": [
    {
      "status": "Terminated",
      "city": "Alpharetta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67154",
      "recalling_firm": "Perrigo Diabetes Care",
      "address_1": "3780 Mansell Rd Ste T50",
      "address_2": "",
      "postal_code": "30022-1112",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including GA & TX",
      "recall_number": "Z-0990-2014",
      "product_description": "Orbit soft Infusion sets, 9 mm soft cannula, P10 (42\") tubing; Single-Use Set  Orbit Infusion Sets are sterile, non-pyrogenic, single-use devices intended to be used for subcutaneous delivery of medication , for example insulin, from an external infusion pump.",
      "product_quantity": "7924 units",
      "reason_for_recall": "Perrigo was notified by our supplier, Ypsomed AG, that they are recalling these batches due to being manufactured with a non-conforming female luer-lock connector. The luer-lock connectors are incompatible with standard male luer-lock connectors for use in connection to compatible infusion pumps.  Perrigo became was notified of a quality complaint related to this issue on November 20, 2013 from a",
      "recall_initiation_date": "20131220",
      "center_classification_date": "20140219",
      "termination_date": "20140805",
      "report_date": "20140226",
      "code_info": "400064, 400069, 400070, & 400072  UPC: 7613117009502",
      "more_code_info": ""
    }
  ]
}