{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Somerville",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84639",
      "recalling_firm": "Ethicon, Inc.",
      "address_1": "Us Highway 22 West",
      "address_2": "N/A",
      "postal_code": "08876",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution nationwide. US and Puerto Rico only.",
      "recall_number": "Z-0989-2020",
      "product_description": "Coated VICRYL (Polyglactin 910) Suture, 18\" (45cm) Product Code J497H. The device is a synthetic absorbable sterile surgical suture.",
      "product_quantity": "23688 units",
      "reason_for_recall": "Some sutures in the lot may exhibit a small section of braid fraying, which can lead to reduced tensile strength at that point.",
      "recall_initiation_date": "20200103",
      "center_classification_date": "20200204",
      "termination_date": "20210813",
      "report_date": "20200212",
      "code_info": "lot PGZ908"
    }
  ]
}