  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bolzano",
      "state": "",
      "country": "Italy",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76102",
      "recalling_firm": "HEALTH ROBOTICS S.R.L.",
      "address_1": "Via Sebastian Altmann 9 A",
      "address_2": "39100",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Shipped throughout United States.",
      "recall_number": "Z-0989-2017",
      "product_description": "i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System  Device Listing Number E454448",
      "product_quantity": "121 systems",
      "reason_for_recall": "The firm has initiated a product correction due to a bag carousel motor error causing misalignment and resulting in a mislabeling of final IV bag preparations.",
      "recall_initiation_date": "20161123",
      "center_classification_date": "20170110",
      "termination_date": "20170315",
      "report_date": "20170118",
      "code_info": "All serial numbers are affected.",
      "more_code_info": ""
    }
  ]
}