  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89870",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0987-2022",
      "product_description": "Artis pheno- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use-single and biplane diagnostic imaging and interventional procedures.  Model:  10849000",
      "product_quantity": "20 units",
      "reason_for_recall": "Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens   Healthineers table or Trumpf/MAQUET table (OEM).  The potential software errors are related to the block movement\u0019s function and manual rotation of the flat detector.  Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed",
      "recall_initiation_date": "20220225",
      "center_classification_date": "20220421",
      "report_date": "20220427",
      "code_info": "Serial Number:   164719  164712  164740  164722  164763  164735  164723  164743  164741  164728  164711  164718  164739  164703  164721  164726  164724  164725  164760    UDI: 04056869046877",
      "more_code_info": ""
    }
  ]
}