      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89870",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0986-2022",
      "product_description": "Artis icono floor  in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures.  Model:   11327700",
      "product_quantity": "35 units",
      "reason_for_recall": "Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens   Healthineers table or Trumpf/MAQUET table (OEM).  The potential software errors are related to the block movement\u0019s function and manual rotation of the flat detector.  Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed",
      "recall_initiation_date": "20220225",
      "center_classification_date": "20220421",
      "report_date": "20220427",
      "code_info": "S/N:  Serial   170319  170026  170343  170367  170368  170305  170360  170307  170331  170349  170309  170310  170361  170351  170383  170023  170304  170328  170334  170336  170025  170346  170051  170358  170357  170333  170365  170332  170312  170027  170303  170378  170322  170053  170364      UDI: 04056869149325",
      "more_code_info": ""
    }
  ]
}