  "results": [
    {
      "country": "United States",
      "city": "Broomfield",
      "address_1": "310 Interlocken Pkwy Ste 120",
      "reason_for_recall": "The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve.",
      "address_2": "",
      "product_quantity": "14 units",
      "code_info": "LOT Numbers: TU00053A, TU00053B and TU00118  Product Code: 8630\u00100205",
      "center_classification_date": "20150115",
      "distribution_pattern": "US Distribution to states of: AZ, CA, NY, TN and TX including PR.",
      "state": "CO",
      "product_description": "Timberline Anchored Lateral Fixed Retractable Sleeve Model No. 8630-0205.  Timberline Anchored Lateral Fixed Retractable Sleeves are used with the Timberline MPF System to help establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae. The Timberline Retractable Drill and the Timberline Retractable Awl fit within this Sleeve.   Spinal Surgical Device.",
      "report_date": "20150121",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Biomet Spine, LLC",
      "recall_number": "Z-0983-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70070",
      "termination_date": "20150313",
      "more_code_info": "",
      "recall_initiation_date": "20130822",
      "postal_code": "80021-3464",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}