  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67343",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including AL, AZ, AR, CA, CO, FL, GA, IL, IN, IA, KY, LA, MN, MO, NE, NJ, NM, NY, NC, OH, OK, PA, SD, TN, TX, UT, VA, WA, WV, WI and Internationally to:    AUSTRALIA, AUSTRIA BELGIUM BOSNIA AND HERZEGOVINA, BRAZIL, CANADA, CHILE, COLOMBIA, CHINA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FRANCE, FINLAND, GEORGIA, GERMANY, HONG KONG, IRAN, ISRAEL INDIA,M ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, MALAYSIA, MEXICO, MOROCCO, NEW ZEALAND, NORWAY, RUSSIA, REPUBLIC OF KOREA, POLAND, ROMANIA, SAUDI ARABIA, SPAIN, SWEDEN SWITZERLAND, TAIWAN, TUNISIA, THAILAND, TURKEY, UNITED KINGDOM.",
      "recall_number": "Z-0981-2014",
      "product_description": "GE Healthcare, PET Discovery 610, Discovery 710, Discovery 600 and Discovery 690.    The GE Discovery XR and XA Systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery XR and XA are to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. These devices can also assist in radiotherapy planning. The Discovery XR and XA can also be used as stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging systems.",
      "product_quantity": "376 (101 US, 275 OUS)",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue due to scatter overcorrection associated with the List Mode Replay feature of your Discovery 600, 610, 710 and Discovery 690.",
      "recall_initiation_date": "20131120",
      "center_classification_date": "20140214",
      "termination_date": "20140729",
      "report_date": "20140226",
      "code_info": "FMI   13715  Serial #                     System ID  00000012391PT9 530UCVETPET      00000012525PT2 RHCDISC710       00000012544PT3 856247PETUPS     00000012561PT7 505216D710       00000012574PT0 858805PCT        00000012615PT1 630432D710       00000012674PT8 801581D710       00000012723PT3 602839D710       00000012732PT4 402327D610       00000012752PT2 719365PETCT      00000012771PT2 414805D710       00000012834PT8 904710PET        00000012843PT9 918505PETCT1     00000012849PT6 205975PET1       00000012865PT2 205975PET2       00000012896PT7 985542D610       00000012915PT5 256386HKPETCT    00000012924PT7 318675D710       00000012937PT9 504842D600       00000012961PT9 303673PETCT      00000012975PT9 605328D710       00000013017PT9 608WIMRD710      00000013057PT5 847872D610       00000013066PT6 4056793  00000013067PT4 212288DST710     00000013072PT4 002902086  00000013081PT5 253968D710       00000013101PT1 270688PETCT      00000012473PT5 910452029  00000012749PT8 910456003  00000012837PT1 83095572097  00000012888PT4 83092720723  00000013026PT0 83092879582  00000012626PT8 AH0132PE01       00000012774PT6 306655D710  00000012533PT6 CT428868CN4      00000012900PT7 EGL067  00000012538PT5 HC1992PE01  00000012582PT3 DK1043PT02       00000012819PT9 DK1054PT17  00000012963PT5 EG1017PT01  00000012993PT2 EG1525PT01  00000012822PT3 A4166968         00000012970PT0 M9886853         00000013082PT3 M4145026  00000013099PT7 EGL110  00000012717PT5 8.30458E+13  00000012722PT5 8.3045E+13  00000012738PT1 8.3045E+13  00000012871PT0 4065738  00000012881PT9 4108707  00000012909PT8 83045062157413  00000012921PT3 8.30452E+13  00000013020PT3 8.30452E+13  00000013127PT6 4119154  00000012600PT3 A5124541         00000012638PT3 A5367810         00000012644PT1 A5366305         00000012655PT7 B5801736         00000012756PT3 A51142135        00000012877PT7 A51051131        00000012887PT6 A51273167        00000012930PT4 A5330720 , A5330718 , A5330719  00000012940PT3 B5315353         00000013023PT7 B5811440  00000012553PT4 GQF002           00000012568PT2 SQF701           00000012585PT6 GQF001           00000012649PT0 SQF001           00000012735PT7 GQF003           00000012766PT2 GQF004           00000012807PT4 SQF710           00000012811PT6 GQ0001           00000012951PT0 SQF003  00000013048PT4 EGL098  00000013078PT1 EGL106  00000013112PT8 EGL116  00000013118PT5 EGL117  00000012599PT7 850450166  00000012665PT6 850450167  00000012788PT6 850450169  00000012801PT7 PL1674PT01       00000012996PT5 PL3057PT01       00000013008PT8 PL2711PT01  00000013038PT5 PL3040PT01       00000013075PT7 EGL105  00000013054PT2 RO1252PT01  00000013014PT6 PET13014PT6  00000012589PT8 34339PET01       00000012868PT6 34367PET01       00000012913PT2 00060DST05       00000012925PT4 00060DST04       00000012931PT2 00060CTS02  00000013007PT0 00040DST01        FMI 13716      Serial #                    System ID  00000001240PT5 262574DST3  00000011146PT8 717657CTPT  00000011218PT5 631MSKPCTD600  00000011233PT4 845695PCT600  00000011344PT9 718370DST6  00000011428PT0 386586FHD600  00000011472PT8 928788PETCT  00000011491PT8 405271PET2  00000011531PT1 386943FHDD600  00000011556PT8 262434D600  00000011565PT9 417347PET  00000011603PT8 479936PETHOG  00000011628PT5 813745D600  00000011636PT8 937771PCT  00000011650PT9 562826D600  00000011653PT3 201858PT600  00000011713PT5 425952D600  00000011714PT3 813396D600  00000011807PT5 626AHMC560  00000011815PT8 520AZO560  00000011870PT3 626COH560  00000011877PT8 RADNETPT1826  00000011912PT3 805739PET1  00000011919PT8 219879PET  00000011923PT0 201858PT2  00000011934PT7 704225PET  00000011937PT0 256308PET  00000011938PT8 RADNETPT619  00000011939PT6 985649D600  00000012244PT0 510537PET  00000012440PT4 770400PCT  00000010847PT2 520UMID690  00000011014PT8 515643D690  00000011240PT9 214370D690  00000011243PT3 312996D690  00000011297PT9 212639PETD690  00000011339PT9 713563D690  00000011435PT5 3154PCT690  00000011460PT3 309672PCT  00000011463PT7 951704PTCT  00000011508PT9 212639PET2D690  00000011553PT5 901595D690  00000011570PT9 480256D690  00000011655PT8 609730D690  00000011757PT2 215464PETCT  00000011851PT3 LARKIND690  00000011878PT6 404712PCT  00000011885PT1 570824PCT690VCT  00000012087PT3 201222PET690  00000012117PT8 312440VCT690  00000012190PT5 916UCDPETCT2  00000012205PT1 681342PCT690  00000012227PT5 716859690VCT  00000012254PT9 614NCH690  00000012257PT2 570476PCT690E  00000012382PT8 970945PET  00000012413PT1 713SCD690  00000012495PT8 804VCUPET690  00000012897PT5 518626PCT  00000010622PT9 650724PET60  00000010952PT0 310888D600  00000010058PT6 NM178700  00000010546PT0 MSK68PET690  00000010816PT7 717242D690PET  00000010902PT5 262928D690  00000011071PT8 816234PETCT  00000011082PT5 574364PET  00000011905PT7 504883D690  00000010454PT7 Shipped  00000010964PT5 Shipped  00000011377PT9 Shipped  00000011427PT2 Shipped  00000013032PT8 Shipped  00000011735PT8 910452027  00000012456PT0 910454013  00000010110PT5 910454012  00000010571PT8 AH5689PE02  00000011227PT6 270617PT02  00000011429PT8 030359PT03  00000011932PT1 BA4073PT01  00000011325PT8 PT256897HM0  00000011497PT5 PET265111HM5  00000011511PT3 PET266034HM8  00000011523PT8 1365454  00000011634PT3 PET269197HM0  00000011728PT3 PET11728PT3  00000011743PT2 2831318PT  00000011783PT8 PET11783PT8  00000011822PT4 PETSXPT0004  00000011846PT3 PET11846PT3  00000011855PT4 CTSXPT00002  00000011911PT5 PETSXPT00001  00000012247PT3 PET12247PT3  00000012642PT5 PET12642PT5  00000012189PT7 PET12189PT7  00000011027PT0 2834425  00000011587PT3 PETSXPT00005  00000011872PT9 PET274985HM1  00000012009PT7 PET12009PT1  00000012448PT7 PET12447PT9  00000012489PT1 PET12489PT1  00000012817PT3 Shipped  00000011431PT4 604877D600  00000011527PT9  613761HIPET2  00000011399PT3 514848PET  00000011298PT7 604877D690  00000011981PT8 82445030011  00000012307PT5 82445160014  00000012431PT3 82445010008  00000012705PT0 82445070004  00000012708PT4 82445160018  00000012741PT5 82445040016  00000012785PT2 82445100018  00000012984PT1 82445140007  00000012136PT8 0828450016_  00000012796PT9 082445100017  00000012804PT1 082445120008  00000012912PT2 82445210009  00000012918PT9 082445050009  00000013051PT8 082445120009  00000011831PT5 PET11831PT5  00000011503PT0 HR4199PE01  00000012065PT9 CS1338PT01  00000011120PT3 CS1318PT01  00000010017PT2 DK1054PT15  00000010275PT6 DK1047PT01  00000011117PT9 FI1015MN02  00000011216PT9 FI1003MN08  00000012361PT2 Shipped  00000011744PT0 X256517201  00000011902PT4 V7405216  00000012174PT9 M256517201  00000012182PT2 X257514801  00000011350PT6 X6010802  00000011356PT3 M1631413  00000011371PT2 B4155424  00000011494PT2 B1768006  00000011516PT2 M5173626  00000011551PT9 M4505295  00000011602PT0 M7417003  00000011627PT7 M4160468  00000011750PT7 A4166958  00000011754PT9 A2943401  00000011888PT5 X4211008  00000012176PT4 M251400604  00000012304PT2 M54606120  00000012347PT1 M4135181  00000012352PT1 M171582005  00000010046PT1 M40180107  00000010648PT4 M4143978  00000011021PT3 M227138301  00000011741PT6 RU1195PT02  00000011829PT9 RU1195PT01  00000011935PT4 NG1127PT01  00000012659PT9 RU2527PT02  00000012777PT9 QA1064PT01  00000012364PT6 GE1004PT01  00000010985PT0 HC0881PE01  00000011190PT6 125521PE01  00000011199PT7 HLAH01PE01  00000012206PT9 HC7946PE02  00000010955PT3 HC0881PE01  00000012342PT2 828450017  00000011808PT3 83045803345211  00000012331PT5 83045202930312  00000011770PT5 83045826760711  00000012689PT6 83045822047612  00000011092PT4 Shipped  00000011388PT6 83045861981111  00000011462PT9 83045002714311  00000011809PT1 83045241979711  00000011901PT6 83045840038411  00000011933PT9 83045842087012  00000012698PT7 83045029088212  00000012704PT3 83045200514712  00000011244PT1 IR1081PT02  00000010509PT8 IL1020NM35  00000011612PT9 A5402105  00000012340PT6 A51272119  00000010846PT4 A5129989  00000011136PT9 F5119164  00000011196PT3 A51043178  00000011534PT5 A5292313  00000011774PT7 A5107783  00000012212PT7 A5691213  00000012255PT6 A5639606  00000010198PT0 B5140608  00000010616PT1 A5216840  00000011147PT6 MF0305  00000011181PT5 MF0303  00000011191PT4 MF0301  00000011219PT3 MFV304  00000011329PT0 MF0313  00000011345PT6 MF0304  00000011418PT1 MF0338  00000011528PT7 MFV305  00000011641PT8 MFV310  00000011796PT0 MF0101  00000011847PT1 MFV306  00000011976PT8 MF0306  00000012118PT6 MFV307  00000012188PT9 MF0016  00000012245PT7 MF0400  00000012348PT9 MFV308  00000012373PT7 MF0012  00000012471PT9 MFV351  00000011379PT5 FT0051  00000011661PT6 FTV621  00000012184PT8 FTV002  00000012386PT9 FTV051  00000012396PT8 FTV003  00000012398PT4 FTV004  00000012405PT7 FTV005  00000010799PT5 FT0101  00000011871PT1 JO1139PT01  00000011179PT9 850450127  00000011387PT8 850450136  00000011814PT1 850450157  00000011936PT2 850450160  00000012202PT8 850450170  00000011228PT4 850450130  00000011268PT0 850450129  00000011571PT7 850450146  00000011662PT4 850450137  00000011931PT3 850450155  00000012057PT6 850450161  00000012146PT7 850450154  00000012357PT0 850450162  00000012197PT0 000850450163  00000012236PT6 KW1020PT01  00000011977PT6 LB1331PT01  00000012374PT5 834450015  00000011991PT7 PET11991PT7  00000011992PT5 CT11992PT5  00000012466PT9 PET12466PT9  00000012720PT9 CT12720PT9  00000011224PT3 2901300DISC  00000012857PT9 PET12857PT9  00000012015PT4 PET12015PT4_  00000012339PT8 MA1135PT01  00000011260PT7 920459004  00000011319PT1 920459004  00000011355PT5 920459005  00000010603PT9 920459002  00000011011PT4 NO1260PE01  00000010848PT0 PL1791PT01  00000011186PT4 PL1803PT01  00000012256PT4 PL1541PT01  00000011731PT7 RU2527PT01  00000011581PT6 RU1712PT01  00000011606PT1 RU1712PT02  00000012564PT1 RU8370RP01  00000012985PT8 PET12985PT8  00000011622PT8 SA2253PT01  00000011881PT0 SA1054PT01  00000012266PT3 SA1013PT01  00000010962PT9 SA1058PT01  00000011773PT9 MPX29917  00000012081PT6 FPG12206  00000011420PT7 M001PE06  00000010990PT0 120810PE01  00000011554PT3 529150PE03  00000011632PT7 260010PE01  00000012066PT7 547090PE01  00000011385PT2 0853450112  00000012514PT6 853450121  00000012511PT2 856450003  00000012632PT6 856450005  00000011727PT5 TN1010PT01  00000011642PT6 34372PET01  00000011668PT1 23011PET01  00000011975PT0 58012PET01  00000011398PT5 34329PET01  00000011611PT1 41023PET01  00000011880PT2 34344PET01  00000012290PT3 Shipped  00000012335PT6 Shipped  00000011436PT3 00860DST02  00000012054PT3 IE1510MN19  00000010083PT4 70001DST02  00000010220PT2 70065DST01  00000010715PT1 00139DST01",
      "more_code_info": ""
    }
  ]
}