  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89860",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide.",
      "recall_number": "Z-0979-2022",
      "product_description": "Artis Q.zen ceiling, Model Number 10848354",
      "product_quantity": "81 systems",
      "reason_for_recall": "If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message \"TUBE HOT, have a break\". Several minutes later the system will block X-ray to prevent further damage and display the message \"NO XRAY: TUBE TOO HOT\".",
      "recall_initiation_date": "20220223",
      "center_classification_date": "20220421",
      "report_date": "20220427",
      "code_info": "UDI: 4056869010021; Serial Numbers: 111160 111162 111163 111137 111502 111139 111174 111185 111503 111005 111002 111145 111176 111510 111155 111179 111140 111507 111122 111168 111170 111173 111512 111141 111508 111521 111133 111003 111111 111103 111101 111011 111182 111184 111143 111150 111164 111183 111504 111149 111109 111151 111129 111121 111130 111517 111513 111017 111010 111115 111116 111117 111119 111104 111148 111127 111175 111506 111112 111131 111132 111134 111158 111167 111157 111161 111142 111156 111500 111501 111169 111505 111166 111107 111514 111000 111106 111123 111128 111135  Additional Serial Numbers as of 10/4/23: 111125",
      "more_code_info": ""
    }
  ]
}