  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukegan",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91322",
      "recalling_firm": "Cardinal Health 200, LLC",
      "address_1": "3651 Birchwood Dr",
      "address_2": "N/A",
      "postal_code": "60085-8337",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia, Singapore, Hong Kong.",
      "recall_number": "Z-0978-2023",
      "product_description": "Cardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2215",
      "product_quantity": "812 units",
      "reason_for_recall": "Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains,  and broken wound tips.",
      "recall_initiation_date": "20221214",
      "center_classification_date": "20230118",
      "report_date": "20230125",
      "code_info": "UDI/DI 50885380182523(cs), 20885380182522(ea), Lot Numbers: N210278, N210368, N210423, N210460, N210468 , N210469, N210543, N220050, N220099, N220139, N220163"
    }
  ]
}