  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89860",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide.",
      "recall_number": "Z-0978-2022",
      "product_description": "Artis Q.zen floor, Model Number 10848353",
      "product_quantity": "33 systems",
      "reason_for_recall": "If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message \"TUBE HOT, have a break\". Several minutes later the system will block X-ray to prevent further damage and display the message \"NO XRAY: TUBE TOO HOT\".",
      "recall_initiation_date": "20220223",
      "center_classification_date": "20220421",
      "report_date": "20220427",
      "code_info": "UDI: 4056869010014; Serial Numbers: 105120 105110 105033 105032 105500 105105 105028 105503 105018 105114 105115 105026 105025 105135 105016 105119 105137 105501 105132 105128 105142 105017 105121 105122 105123 105100 105112 105124 105029 105030 105031 105125  Additional Serial Numbers as of 10/4/23: 143500",
      "more_code_info": ""
    }
  ]
}