{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Broomfield",
      "address_1": "310 Interlocken Pkwy Ste 120",
      "reason_for_recall": "The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve shaft.",
      "address_2": "",
      "product_quantity": "12 units",
      "code_info": "Model 8630-0204 Lots TU00052, TU00052A, and TU00117",
      "center_classification_date": "20150114",
      "distribution_pattern": "US Distribution to states of: AZ, CA, NY, TN, and TX including PR.",
      "state": "CO",
      "product_description": "Timberline Anchored Lateral Variable Retractable Sleeve LITPCR8600-2101-07, Model No. 8630\u00100204     Timberline Anchored Lateral Variable Retractable Sleeves (the Sleeves) are used with the Timberline MPF System to help establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae.",
      "report_date": "20150121",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Biomet Spine, LLC",
      "recall_number": "Z-0978-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70060",
      "termination_date": "20150313",
      "more_code_info": "",
      "recall_initiation_date": "20130822",
      "postal_code": "80021-3464",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}