  "results": [
    {
      "status": "Ongoing",
      "city": "Waukegan",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91322",
      "recalling_firm": "Cardinal Health 200, LLC",
      "address_1": "3651 Birchwood Dr",
      "address_2": "N/A",
      "postal_code": "60085-8337",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia, Singapore, Hong Kong.",
      "recall_number": "Z-0977-2023",
      "product_description": "Cardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2214",
      "product_quantity": "3550 units",
      "reason_for_recall": "Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains,  and broken wound tips.",
      "recall_initiation_date": "20221214",
      "center_classification_date": "20230118",
      "report_date": "20230125",
      "code_info": "UDI/DI 50885380182516(cs), 20885380182515(ea), Lot Numbers:  N210278, N210347, N210368, N210423, N210460, N210468, N210469, N210543, N220050, N220099, N220139, N220163"
    }
  ]
}