  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82026",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution to nationwide US.    Foreign distribution to Australia, Belgium,  Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic,  Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India,  Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia,  Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto  Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan,  Trinidad and Tobago, United Arab Emirates, and the United Kingdom.",
      "recall_number": "Z-0975-2020",
      "product_description": "Sof-Flex Double Pigtail Ureteral Stent  Set, Global Product No. G15076  G14906  G14840  G14865  G17128",
      "product_quantity": "13388 total",
      "reason_for_recall": "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.",
      "recall_initiation_date": "20190104",
      "center_classification_date": "20200204",
      "termination_date": "20200529",
      "report_date": "20200212",
      "code_info": "NS7949093  NS8077953  NS8077954  NS8077955  8070691  8183342  8183343  8070693  8172134  8187189  NS8070692  8424304  7990841"
    }
  ]
}