  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82026",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution to nationwide US.    Foreign distribution to Australia, Belgium,  Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic,  Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India,  Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia,  Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto  Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan,  Trinidad and Tobago, United Arab Emirates, and the United Kingdom.",
      "recall_number": "Z-0974-2020",
      "product_description": "Kwart Retro-Inject Ureteral  Stent Set, Global Product No. G15016  G14836  G14885  G14887  G17151",
      "product_quantity": "13388 total",
      "reason_for_recall": "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.",
      "recall_initiation_date": "20190104",
      "center_classification_date": "20200204",
      "termination_date": "20200529",
      "report_date": "20200212",
      "code_info": "8011027  8183322  8037660  8030233  8018054  7994816"
    }
  ]
}