      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82026",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution to nationwide US.    Foreign distribution to Australia, Belgium,  Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic,  Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India,  Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia,  Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto  Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan,  Trinidad and Tobago, United Arab Emirates, and the United Kingdom.",
      "recall_number": "Z-0973-2020",
      "product_description": "C-Flex Double Pigtail Ureteral Stent Set, Global Product No. G14364  G14365  G14367",
      "product_quantity": "13388 total",
      "reason_for_recall": "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.",
      "recall_initiation_date": "20190104",
      "center_classification_date": "20200204",
      "termination_date": "20200529",
      "report_date": "20200212",
      "code_info": "NS7953136  NS8169661  NS8180857  NS8169662  NS8180858  NS8180859  NS8180860"
    }
  ]
}