{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89860",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide.",
      "recall_number": "Z-0972-2022",
      "product_description": "Artis zee biplane MN, Model Number 10094143",
      "product_quantity": "7 systems",
      "reason_for_recall": "If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message \"TUBE HOT, have a break\". Several minutes later the system will block X-ray to prevent further damage and display the message \"NO XRAY: TUBE TOO HOT\".",
      "recall_initiation_date": "20220223",
      "center_classification_date": "20220421",
      "report_date": "20220427",
      "code_info": "UDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402",
      "more_code_info": ""
    }
  ]
}