{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82026",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution to nationwide US.    Foreign distribution to Australia, Belgium,  Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic,  Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India,  Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia,  Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto  Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan,  Trinidad and Tobago, United Arab Emirates, and the United Kingdom.",
      "recall_number": "Z-0972-2020",
      "product_description": "Bander Ureteral Diversion Open-End Stent Set, Global Product No. G14822  G14823",
      "product_quantity": "13388 total",
      "reason_for_recall": "The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.",
      "recall_initiation_date": "20190104",
      "center_classification_date": "20200204",
      "termination_date": "20200529",
      "report_date": "20200212",
      "code_info": "8011030  8033568  8063430  8063431  8150491  8417631  8417632  8587162  8587163  8590532  8590533  7990830  8003084  8033571  8063428  8590520  8590522  8629475  NS8629473"
    }
  ]
}