  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Raanana",
      "state": "",
      "country": "Israel",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79243",
      "recalling_firm": "Medimop Medical Projects Ltd.",
      "address_1": "4 Hayetzira Street,",
      "address_2": "P.O. Box 2499",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Distribution",
      "recall_number": "Z-0972-2018",
      "product_description": "Vial2BagDC, Product Number 6070112.   The 20mm Vial2Bag DC device, is intended for use in  healthcare facilities or in-home environment by the patient  or care-giver to aid and support prescribed treatment and  therapy. The device consists of the Vial2Bag piercing  spike and cover, the IV Port connector and an integrated  Vial Adapter (20mm) for access to the drug/solution vial.",
      "product_quantity": "453,600 units",
      "reason_for_recall": "Medimop is recalling certain lots of this product Vial2Bag DC due to the potential of the IV Port separating from the Via12Bag  DC body during the drug reconstitution I IV Set attachment process.",
      "recall_initiation_date": "20171006",
      "center_classification_date": "20180312",
      "termination_date": "20240123",
      "report_date": "20180321",
      "code_info": "Lots: 9507,9508,9509,9510,9511,9512,9513.",
      "more_code_info": ""
    }
  ]
}