  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Malvern",
      "address_1": "40 Liberty Blvd",
      "reason_for_recall": "Siemens initiated a corrective action to address two possible, mutually independent causes of a system defect related to the following:  - In Artis Systems with A100 Plus or A100G generators, an attempt to resume operation following detection of a fault can result in the failure of a module in the high-voltage generator.  - For biplane systems delivered with software version VD11B since April of this year, software problems in conjunction with graphics card may, in rare cases, result in the loss of image display in the examination room.",
      "address_2": "",
      "product_quantity": "1,500 distributed Worldwide",
      "code_info": "Model # 10848280, 10848281, 10848282, 10848353, 10848354, 10848355",
      "center_classification_date": "20170109",
      "distribution_pattern": "Nationwide Distribution",
      "state": "PA",
      "product_description": "Artis Q/Q.zen,  Angiographic x-ray system  Stand alone system  The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee/ zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee/ zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis",
      "report_date": "20170118",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-0972-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "76016",
      "termination_date": "20190409",
      "more_code_info": "",
      "recall_initiation_date": "20161208",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}