  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Franklin Lakes",
      "address_1": "1 Becton Dr",
      "reason_for_recall": "A limited portion of the lot was manufactured with less than the required amount of K2EDTA additive, an anticoagulant, which may cause erroneous results that could lead to specimen recollection and potential delay of treatment or misdiagnosis/mismanagement of treatment and serious complications, such as failure to detect thrombocytosis or postponing surgery.",
      "address_2": "",
      "product_quantity": "8,494,000",
      "code_info": "Catalog No. 367841  Lot No. 6279849  UDI: (01)30382903678410",
      "center_classification_date": "20180312",
      "distribution_pattern": "Nationwide Distribution",
      "state": "NJ",
      "product_description": "BD Vacutainer¿ K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL).  Catalog Number: 367841.    Used for whole blood hematology determinations, immunohematology testing and blood donor screening.",
      "report_date": "20180321",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Becton Dickinson & Company",
      "recall_number": "Z-0968-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79041",
      "termination_date": "20190516",
      "more_code_info": "",
      "recall_initiation_date": "20170829",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}