{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Atlanta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82072",
      "recalling_firm": "Elekta, Inc.",
      "address_1": "400 Perimeter Center Ter NE Ste 50",
      "address_2": "",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution to Wisconsin, Texas.      International distribution to Denmark, England, Germany, Netherlands, Sweden.",
      "recall_number": "Z-0965-2019",
      "product_description": "Monaco Radiation Treatment Planning (RTP) System",
      "product_quantity": "14 devices",
      "reason_for_recall": "If  Improve Target Dose  was chosen as an optimization model in a previous treatment session, Monaco¿ will  automatically use this optimization model again when proceeding with the online plan adaptation of a completion plan when it should not.",
      "recall_initiation_date": "20190208",
      "center_classification_date": "20190227",
      "termination_date": "20240819",
      "report_date": "20190306",
      "code_info": "Software version 5.40",
      "more_code_info": ""
    }
  ]
}