  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hercules",
      "state": "CA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66947",
      "recalling_firm": "Bio-Rad Laboratories, Inc.",
      "address_1": "4000 Alfred Nobel Dr",
      "address_2": "N/A",
      "postal_code": "94547-1803",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Latin America, Asia Pacific, Canada, Australia, New Zealand and India.",
      "recall_number": "Z-0965-2014",
      "product_description": "D-10 Dual Program, which contains Update Kit D-10 A1c program diskette.    United States; Bio-Rad Laboratories,  Hercules, CA 94547    The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, Ale} in whole blood. The separation is based on the principles of high performance liquid chromatography.",
      "product_quantity": "2906 kits (318 Kits in US; 2588 kits internationally)",
      "reason_for_recall": "D-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the D-10 Hemoglobin Testing System.",
      "recall_initiation_date": "20131122",
      "center_classification_date": "20140210",
      "termination_date": "20140320",
      "report_date": "20140219",
      "code_info": "Catalog number: 220-0201:  Reorder Pack Lot No./Expiration date:  20030322 3/31/2014  20030337 3/31/2014  64000081 5/31/2014   64000331 5/31/2014   64000332 5/31/2014  64000888 08/31/2014  64001344 04/30/2014  64002239 11/30/2014     (  64002240 08/3112014   64002241 11/30/2014"
    }
  ]
}