  "results": [
    {
      "status": "Ongoing",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91392",
      "recalling_firm": "Smiths Medical ASD Inc.",
      "address_1": "6000 Nathan Ln N",
      "address_2": "N/A",
      "postal_code": "55442-1690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ, CL, LB, MY, HU, JP, TW, and CN",
      "recall_number": "Z-0963-2023",
      "product_description": "CADD Administration Set, List Numbers:  1) REF 21-7300-24,  2) REF 21-7301-24 3) REF 21-7302-24,  4) REF 21-7321-24,  5) REF 21-7322-24,  6) REF 21-7323-24,  7) REF 21-7324-24,  8) REF 21-7333-24,  9) REF 21-7336-24,  10) REF 21-7339-24,  11) REF 21-7343-24,  12) REF 21-7349-24,  13) REF 21-7359-24,  14) REF 21-7383-24,  15) REF 21-7390-24,  16) REF 21-7391-24,  17) REF 21-7394-24,  18) REF 21-7395-24,  19) REF 21-7301-24JP,  20) REF 21-7302-24JP,  21) REF 21-7321-24JP,  22) REF 21-7322-24JP,  23) REF 21-7359-24JP,  24) REF 21-7394-24JP",
      "product_quantity": "45285351 units",
      "reason_for_recall": "There are two potential issues with CADD Infusion  System Infusion Sets:  (1) Lack of Delivery or  Underdelivery related to Tubing Occlusion and (2) False \"No Disposable Attached (NDA)\" Alarms",
      "recall_initiation_date": "20221209",
      "center_classification_date": "20230130",
      "report_date": "20230208",
      "code_info": "1) REF 21-7300-24, UDI/DI 30610586027196, Lot numbers:  3630772 through 4334107;  2) REF 21-7301-24, UDI/DI 30610586027219, Lot numbers:  3630747 through 4334119;  3) REF 21-7302-24, UDI/DI 30610586027233, Lot numbers:  3617363 through 4334138;   4) REF 21-7321-24, UDI/DI 30610586029633, Lot numbers:  3773534 through 4308545;  5) REF 21-7322-24, UDI/DI 30610586029640, Lot numbers:  3776375 through 4334324;  6) REF 21-7323-24, UDI/DI 061058603961, Lot numbers:  3776373 through 4334325;  7) REF 21-7324-24, UDI/DI 30610586029657, Lot numbers:  3773527 through 4334328;  8) REF 21-7333-24, UDI/DI 30610586043233, Lot numbers:  3776362 through 3984144;  9) REF 21-7336-24, UDI/DI 30610586027295, Lot numbers:  3776360 through 4025381;  10) REF 21-7339-24, UDI/DI 30610586027301, Lot numbers:  3780565 through 4009665;  11) REF 21-7343-24, UDI/DI 35019517191824, Lot numbers:  3965344 through 4334338;  12) REF 21-7349-24, UDI/DI 35019517184239, Lot numbers:  3926579 through 4339160;  13) REF 21-7359-24, UDI/DI 30610586027400, Lot numbers:   3776315 through 4334339;  14) REF 21-7383-24, UDI/DI 30610586027479, Lot numbers:  3780549 through 3971523;  15) REF 21-7390-24, UDI/DI 30610586027509, Lot numbers:  3780548 through 4334346;  16) REF 21-7391-24, UDI/DI 30610586027516, Lot numbers:  3773276 through 4334351;  17) REF 21-7394-24, UDI/DI 30610586027530, Lot numbers:  3774739 through 4339178;  18) REF 21-7395-24, UDI/DI 30610586027547, Lot numbers:  3808536 through 4308562 19) REF 21-7301-24JP, UDI/DI N/A , Lot numbers:  3630747 through 4334113;  20) REF 21-7302-24JP, UDI/DI N/A , Lot numbers:  3617363 through 4329630;  21) REF 21-7321-24JP, UDI/DI N/A , Lot numbers:  3773534 through 4308545;  22) REF 21-7322-24JP, UDI/DI N/A , Lot numbers:  3776375 through 4334324;  23) REF 21-7359-24JP, UDI/DI N/A , Lot numbers:  3776315 through 4334339;  24) REF 21-7394-24JP, UDI/DI N/A , Lot numbers:  3774739 through 4339178."
    }
  ]
}