  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87087",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "FL, KY, MD, NS, NY, PA, TX  Foreign: Canada, Switzerland",
      "recall_number": "Z-0960-2021",
      "product_description": "Cook Medical Kwart Retro-Inject Ureteral Stent Set 7.0Fr. 22-32cm -Used for retrograde injection during Extracorporeal Shock Wave Lithotripsy (E.S.W.L.) and leaving an indwelling ureteral stent  Reference Part Number (RPN): 003700  GPN: G14837",
      "product_quantity": "42 units",
      "reason_for_recall": "Manufactured with incorrect material, which may result in the stent being stiffer and potentially susceptible to degradation could potentially lead to breakage/fragmentation.",
      "recall_initiation_date": "20210104",
      "center_classification_date": "20210205",
      "termination_date": "20210916",
      "report_date": "20210217",
      "code_info": "Lot Numbers:   13024536  13089895  NS13089897    UDI:  (01)00827002148376(17)230303(10)13024536  (01)00827002148376(17)230327(10)13089895  (01)00827002148376(17)230327(10)NS13089897"
    }
  ]
}