      "status": "Ongoing",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91339",
      "recalling_firm": "Getinge Usa Sales Inc",
      "address_1": "1 Geoffrey Way",
      "address_2": "N/A",
      "postal_code": "07470-2035",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide and Worldwide Distribution:  Domestic distribution to AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA, WV. Foreign distribution worldwide.",
      "recall_number": "Z-0959-2023",
      "product_description": "Getinge Flow-i Anesthesia Systems  Flow-i C20, Part no. 6677200  Flow-i C30, Part no. 6677300  Flow-i C40, Part no. 6677400",
      "product_quantity": "8,016 Total (7,766 OUS; 250 US)",
      "reason_for_recall": "Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.",
      "recall_initiation_date": "20221222",
      "center_classification_date": "20230126",
      "report_date": "20230201",
      "code_info": "Part #       UDI 6677200 7325710001349 Serial Numbers 1202-8027  Part #       UDI 6677300 7325710001363 Serial Numbers 1058-1201  Part #       UDI 6677400 7325710001318 Serial Numbers 2034-8446  Software versions 4.8.0, 4.8.1, 4.8.2, and 4.8.3.",
      "more_code_info": ""
    }
  ]
}