  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81981",
      "recalling_firm": "Shimadzu Medical Systems Usa Com",
      "address_1": "20101 S Vermont Ave",
      "address_2": "",
      "postal_code": "90502-1328",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide Distribution in the states of IN, OH, IL, SC",
      "recall_number": "Z-0959-2019",
      "product_description": "BRANSIST safire    Product Usage:  DIGITAL ANGIOGRAPHY SYSTEM This system is intended to be used for radiologic visualization of the heart, blood vessels or lymphatic system during or after injection of a contrast medium. It is to be used in the diagnosis of the circulatory vascular system while being combined with a positioning device and other devices.",
      "product_quantity": "8 units",
      "reason_for_recall": "Normal operation of the device is to power up the device in the morning, register the first patient, and then perform a fluoroscopy. If, however, the user powers up the device in the morning and makes an error by starting a fluoroscopy while the first patient is still being registered, the device application will abnormally terminate and require service intervention before it can be used again. This event will not occur after the first patient procedure.",
      "recall_initiation_date": "20190114",
      "center_classification_date": "20190226",
      "termination_date": "20240607",
      "report_date": "20190306",
      "code_info": "Serial Numbers:  40AD58C4B001, 40AD58C55001, 0261Q80503, 0261Q80804, 0161Q84501, 0161Q83102, 3M7A16A1A001, 0161Q82302",
      "more_code_info": ""
    }
  ]
}