  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Acton",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64509",
      "recalling_firm": "Allen Medical Systems, Inc.",
      "address_1": "One Post Office Square",
      "address_2": "N/A",
      "postal_code": "01720",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution- USA including the states of TX, NV, KY, GA, OH, PA, NY, AL, WA, IL and the countries of New Zealand and Japan.",
      "recall_number": "Z-0959-2013",
      "product_description": "Allen Traction Boot,  Modular, OT 1000 Traction Device Boot Assembly, Fracture Table Orthopedic Leg Holders, non-invasive medical devices designed to attach to orthopedic fracture tables to aid clinical staff in applying traction to the leg of a patient during surgical procedures.",
      "product_quantity": "24 devices",
      "reason_for_recall": "Some units may have been assembled incorrectly which could result in the boot breaking free from the mounting base and allowing the patient's leg to fall.",
      "recall_initiation_date": "20130228",
      "center_classification_date": "20130318",
      "termination_date": "20130807",
      "report_date": "20130327",
      "code_info": "A260264, A253312, A260989, A260262, 418292, A260991, A260998, A253313, A260993, A260265, 418293, A253309, A260995, A253310, A260992, A253311, 418294, A253323, A260261, A253321, A260994, 418291, 418295, A253322"
    }
  ]
}