{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Orlando",
      "address_1": "12151 Research Pkwy",
      "reason_for_recall": "The MR 200 devices failed the initial power up test due to capacitor short and hi-pot testing; possible loss of Patient Monitoring or User/Service Electrical Shock.",
      "address_2": "Suite 200",
      "product_quantity": "20",
      "code_info": "Model No. 866120; Lot/ Serial Numbers: US33600065, US33600066, US33600075, US33600077, US33600078, US33600079; US33600080, US33600081, US33600082, US33600083, US33600084, US33600086, US33600087, US33600088, US33600089, US33600090, US33600091, US33600092, US33600093, US33600094.",
      "center_classification_date": "20150113",
      "distribution_pattern": "CA, FL, IN, KY, MN, NC, NJ, NY, OK, SD and TX   Netherlands, Poland and United Kingdom.",
      "state": "FL",
      "product_description": "Philips Expression MR200 MRI Patient Monitoring System.",
      "report_date": "20150121",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Invivo Corporation",
      "recall_number": "Z-0958-2015",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "69587",
      "termination_date": "20170109",
      "more_code_info": "",
      "recall_initiation_date": "20141022",
      "postal_code": "32826-3222",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}