  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Completed",
      "city": "Woodbury",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96078",
      "recalling_firm": "Murata Vios, Inc.",
      "address_1": "700 Commerce Dr Ste 190",
      "address_2": "",
      "postal_code": "55125-9239",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of NY, LA and the country of India.",
      "recall_number": "Z-0956-2025",
      "product_description": "muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate",
      "product_quantity": "275 units",
      "reason_for_recall": "During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will \u001cflicker\u001d or \u001cflash\u001d very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.",
      "recall_initiation_date": "20210621",
      "center_classification_date": "20250116",
      "report_date": "20250122",
      "code_info": "Soft ware version 1.49.02 UDI: +B595BSM20500/\\$\\$7V1.49.02/16D20200504S and Software version 1.52.00 UDI: +B595BSM20500/\\$\\$7V1.52.00/16D20210305K",