{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue involving the breakage of the two steel cables which support the bucky device in the wall stand of Proteus XR/a X-ray imaging systems.    No injuries reported.",
      "address_2": "",
      "product_quantity": "3096 (2486 US, 610 OUS).",
      "code_info": "FMI 17107 - Proteus Wall Stand Counterpoise Cables Break   Serial # System ID  00000044028HL9 928656ER  00000063303HL2 719595ORTHO  00000045940HL4 803751PR2  00000041478HL9 812238FPCRAD  00000057994HL6 208359PROT  00000046527HL8 310732TIPRO  00000047064HL1 318255PRO  00000036130HL3 319352PROT  00000053610HL2 336765WFBIRAD  00000021190HL4 360457ORTHO  00000037481HL9 405377PRO  00000031767HL7 406638CATOMO  00000034159HL4 407628PRO  00000031902HL0 409938TC1  00000074625HL5 419578RAD2  00000096491HL6 419636RAD  00000096684HL6 435563RAD  00000041757HL6 435586PROT  00000057200HL8 435658RAD  00000088306HL6 505983ORTHPROT  00000020347HL1 508698NEPATS  00000049345HL2 573882NSCRAD1  00000064320HL5 601788RAD1  00000048917HL9 601968MAERM1  00000016587HL8 602285PROT2  00000106226HL4 608366PRO  00000047063HL3 616752NWPROTEUS  00000033643HL8 623327XR3  00000059646HL0 630312PROT1  00000082840HL0 630856SCPRO  00000057436HL8 702396XR1  00000102867HL9 703771PROT  00000049772HL7 703784RAD1  00000067466HL3 706641CTC  00000024904HL5 708226PROT  00000041647HL9 708226RM2  00000062869HL3 757606PRO1  00000063690HL2 770252GBJ3  00000049128HL2 801IMCXR  00000036909HL0 804594RM4  00000112434HL6 805681RAD4  00000107242HL0 805681RAD5  00000027096HL7 814833PROT  00000046048HL5 818409RAD3  00000081590HL2 870245PROT  00000041480HL5 904264PRO  00000050595HL8 904381RAD  00000068296HL3 912279CCGA  00000062303HL3 912748PROT  00000035901HL8 918352DRAD1  00000061157HL4 MCCPROT1  00000061158HL2 MCCPROT2  103733HL2 918494XRTRAUMA2  00000020349HL7 210297NCRAD  00000044647HL6 252466CPRAD1  00000060369HL6 303269RM2  00000089696HL9 402294EBPROT  00000038564HL1 520202PRO1  00000038565HL8 520202PRO2  00000073629HL8 609878PRO1  00000075649HL4 609878PRO2  00000045941HL2 816282LSPRO1  00000078007HL2 817250RAD  00000061302HL6 856341PROT  00000059336HL8 203932PROT2  00000106552HL3 412232PROT1  00000019948HL9 812256RT1  00000045545HL1 662434RM2  00000046670HL6 808KWAJPROT  00000084213HL8 412315PROT1  00000050315HL1 318368PRO  00000111660HL7 276694PRO  00000068951HL3 256486XR1  00000058725HL3 931766ER1  00000031770HL1 662779RAD1  000000000000NA 706544MA2  00000104753HL9 0004177304A  00000018751HL8 201358ER1  00000042692HL4 201418ER  00000043018HL1 201418ER  00000023401HL3 201418RAD1  00000021897HL4 201418RAD3  00000060463HL7 201894PRO1  00000060864HL6 201894PRO2  00000112815HL6 203319PROT  00000065033HL3 203466PRO  00000016999HL5 203679RAD  00000058726HL1 203688C262  00000065367HL5 203688PROT2  00000074013HL4 203688PROT2  00000063510HL2 203739PROT  00000022750HL4 203797PROT2  00000028477HL8 203852RM5  00000057596HL9 203932PROT1  00000038056HL8 217333MXR1  00000028501HL5 217337CXR3  00000081384HL0 217366CWXR2  00000021397HL5 217366CXR1  00000034089HL3 217383CXR20  00000033499HL5 217383CXR24  00000044381HL2 217431CFAXR1  00000026066HL1 217431CXR1  00000062685HL3 217443UXR2  00000024374HL1 217732LPR1  00000017462HL3 217788PRT1  00000023841HL0 217824PRT  00000065364HL2 217854RAD1  00000052078HL3 217893CRA2  00000052238HL3 218229PROT  00000016191HL9 218759RAD  00000084370HL6 218773RAD  00000080250HL4 218828RAD1  00000061264HL8 218983RAD  00000017830HL1 219284ER  00000016991HL2 219347RAD  00000017822HL8 219492RAD1  00000018607HL2 219492RAD2  00000022756HL1 219757ER  00000028879HL5 219757WP1  00000111658HL1 219836PRO2  00000028783HL9 225761R5  00000033602HL4 228436XR1  00000057198HL4 228875PRXRA  00000041103HL3 229243RAD1  00000023267HL8 229249SSC  00000074884HL8 229257MR1  00000027844HL0 229259OUT  00000083631HL2 229259PR2  00000036119HL6 229276PROT  00000021034HL4 229386TER  00000057595HL1 229434PER  00000019691HL5 229890CR4  00000018991HL0 231547PROTEUS  00000065484HL8 231726PROTEUS2  00000064530HL9 231726PROTEUS3  00000028630HL2 231924ER  00000016391HL5 231924RM3  00000028785HL4 239254RAD  00000033660HL2 239368XR1  00000049934HL3 239403RAD  00000025617HL2 239432XR1  00000022389HL1 240379RAD  00000035861HL4 248652ER  00000025565HL3 248652RAD4  00000030083HL0 248858ER1  00000017618HL0 717733R1B  00000041407HL8 717761PRO1  00000017640HL4 717859CR1  00000016345HL1 718262RAD1  00000047004HL7 718470XRA  00000064049HL0 718960PROT  00000086110HL4 718960PROT  00000092087HL6 718960PROT1  00000021894HL1 719560GD1  00000063302HL4 719767PROTEUS  00000048109HL3 719846XR  00000084573HL5 719LVRAD1  00000074623HL0 724347PROT  00000102140HL1 724357RM1  00000019535HL4 724537PROT1  00000104749HL7 724588PROT  00000050598HL2 724626PROT1  00000079885HL0 724935UCC  00000016985HL4 724983PX  00000025564HL6 724MTVXR  00000023293HL4 727341ER  00000037716HL8 727341PPER1  00000037377HL9 727341RAD1  00000017622HL2 727521ER  00000042380HL6 727528RAD1  00000041883HL0 727588XRA5  00000047372HL8 731422RM2  00000023840HL2 731425RS  00000031156HL3 731425TRAUMA2  00000028102HL2 731425TRAUMA3  00000030902HL1 731584RAD  00000021837HL0 732222ER2  00000019787HL1 732294ER  00000076135HL3 732295RAD  00000035183HL3 732321ERP2  00000021657HL2 732321ERPRO  00000034453HL1 732499PROT5  00000032882HL3 732499PROT7  00000028975HL1 732530ERP1  00000028881HL1 732530ERP2  00000098078HL9 732840PRO5OP  00000015944HL2 732970RAD  00000016594HL4 734475ORTHO  00000021038HL5 734712PROT1  00000027091HL8 734712T2  00000027946HL3 734930UMED  00000018812HL8 740356PROT  00000023540HL8 740368PR2  00000021838HL8 740368PR4  00000080937HL6 740383PROT  00000106549HL9 740393PRO5  00000041829HL3 740393PROT1  00000050599HL0 740393PROT4  00000047232HL4 740420PRO  00000037887HL7 740454BHER  00000031439HL3 740454BHRM6  00000042861HL5 740454SZUC  00000036692HL2 740474PROT  00000028298HL8 740474R2  00000032399HL8 740592PROD  00000044743HL3 740779PRO1  00000043746HL7 740779PRO3  00000080251HL2 740845PRO  00000049771HL9 740941WRM1  00000030011HL1 757398PRO1  00000023173HL8 757827RM3  00000017631HL3 757934RM1  00000017636HL2 757934RM5  00000020776HL1 760241R2  00000093930HL6 760241RAD1  00000089699HL3 760876RAD  00000033672HL7 760934OP1  00000018990HL2 763236RAD  00000028177HL4 763520N2  00000062561HL6 763581PRO  00000037408HL2 763744PROT  00000029622HL8 763780R2  00000025415HL1 765455IUPROT  00000028290HL5 765455PROT  00000100702HL0 770228RM2  00000037888HL5 770267RM4  00000015868HL3 770389HRAD  00000061066HL7 770434PROT  00000040835HL1 770507EL1  00000075398HL8 770533LR1  00000020592HL2 770748PER  00000050596HL6 770918ER1  00000028505HL6 770918ER2  00000021901HL4 770918PRO1  00000020354HL7 772468ER1  00000016983HL9 773525TPR  00000028181HL6 773564WPRO1  00000028504HL9 773IMPRO11  00000066676HL8 775445PROTEUS  00000016599HL3 775738RAD  00000042078HL6 775982FERNLEY  00000016347HL7 775982XR1  00000029493HL4 775SMPROT  00000050602HL2 781581UR3  00000033000HL1 785239PRO  00000043331HL8 785239PRO2  00000046434HL7 785239PRO3  00000078011HL4 785354PROT  00000073846HL8 785354PROT2  00000085135HL2 785354PROT4  00000063305HL7 785357PROT  00000071486HL5 785452PRO  00000034921HL7 785632PRO  00000061519HL5 801256RAD  00000038331HL5 801268CHT  00000017824HL4 801268ER  00000040770HL0 801314RM1  00000063180HL4 801321OSG2  00000045155HL9 801387PRO1  00000062098HL9 801387PRO2  00000035370HL6 801461PROT  00000064816HL2 801461RM2  00000021658HL0 801501RAD  00000049455HL9 801535RM21  00000037748HL1 801535RM23  00000041581HL0 801562RM2  00000035180HL9 801634RM3  00000057038HL2 801639MI  00000059448HL1 801714XR  00000023176HL1 801774ER  00000051941HL3 801786XR  00000060982HL6 801840RM1  00000061156HL6 801840RM2  00000031152HL2 801855RM4  00000081379HL0 802775PRO  00000068442HL3 803502XR  00000013301HL5 803641ER  00000080462HL5 803641XR4  00000021498HL1 803699PR1  00000055683HL7 803713R1  00000057439HL2 803713R3  00000015955HL8 803791LR1  00000070369HL4 804254RMA  00000021898HL2 804541ER  00000024467HL3 804747DHPER  00000078170HL8 804784RAD1  00000024735HL3 805383XR1  00000089900HL5 805418XR1  00000032165HL3 805525OPXR3  00000069561HL9 805563RAD1  00000030961HL7 805648OPXR2  00000019784HL8 805648RM1  00000024370HL9 805681ORTHO  00000059447HL3 806250RAD  00000050592HL5 806647RAD  00000051286HL3 806647RAD  00000047230HL8 808257PROTRM1  00000053316HL6 808473NMCPRTRM1  00000053415HL6 808473NMCPRTRM3  00000024471HL5 808973PROT  00000051940HL5 810648PRO  00000045758HL0 810844AVANT  00000029205HL2 810985ER  00000019150HL2 812237TH2  00000048620HL9 812254ER  00000020772HL0 864459AR1  00000024660HL3 865218RAD1  00000023296HL7 865218RAD2  00000025118HL1 865218RAD3  00000105863HL5 870208RF  00000103416HL4 870208XR1  00000020861HL1 870423CR2  00000065590HL2 870541PROT6  00000031860HL0 901722RAD1  00000029158HL3 903232LRRT  00000042196HL6 903525RAD1  00000036747HL4 903614PROTEUS  00000062686HL1 903654RAD  00000034663HL5 904202CSRAD1  00000102864HL6 904202ER1  00000095427HL1 904202ER2  00000036321HL8 904276RM1  00000024555HL5 904391RAD1  00000026637HL9 904399ER1  00000061301HL8 904399RADWEST  00000068783HL0 904819WGRAD  00000034025HL7 904821CHEST1  00000111678HL9 906337PROT  00000031158HL9 906341RAD1  00000109254HL3 906632PRO  00000048621HL7 906932PROT  00000060865HL3 907580RM2  00000032396HL4 907762SXR  00000044312HL7 907796PROT  00000072092HL0 907874PROT  00000029230HL0 909336RAD1  00000018494HL5 909788R1  00000017637HL0 909788R2  00000067634HL6 910259PMRAD1  00000045157HL5 910343NHR5  00000062560HL8 910609HSPROT  00000025559HL6 910667BFRAD1  00000049454HL2 910907WARM2PROT  00000049456HL7 910907WARM2PROT  00000053614HL4 910907WARM8PRO  00000086589HL9 910938CHPROT  00000025485HL4 912275RM4  00000052232HL6 912283R209  00000104278HL7 912369PROT  00000038209HL3 912466ER05  00000024375HL8 912564PROT  00000056338HL7 912576PROTEUS  00000019534HL7 912691C02  00000040537HL3 912754PROT  00000078010HL6 912754PROT2  00000030012HL9 913588KER  00000028976HL9 913588KR1  00000039752HL1 913588KUIVP  00000025125HL6 913684CPRO  00000042560HL3 913684RM3  00000062199HL5 913ATCHPROT  00000029057HL7 914378PROTEUS  00000041885HL5 914666PRO  00000034685HL8 915521ERRAD  00000075268HL3 915629RAD  00000021840HL4 915735RAD1  00000023036HL7 915853RAD  00000024584HL5 916784PRO  00000028877HL9 916984PRO  00000048519HL3 918307SFSP1  00000077763HL1 918323PROT1  00000060066HL8 918453PROT1  00000084368HL0 918458PROT1  00000087774HL6 918505PROT  00000017004HL3 918540MRAD  00000076133HL8 918579PROTEUS  00000096318HL1 918689PROT1  00000028571HL8 918744TOMO1  00000020207HL7 918825PROTEUS  00000017827HL7 919350NWRAD1  00000019152HL8 919497PROT  00000032494HL7 919662RAD1  00000035099HL1 919663CRAD  00000030644HL9 919781PROT2  00000016408HL7 919781WRAD1  00000108819HL4 919787RAD3  00000035367HL2 919934JA1  00000035369HL8 919934JA2  00000080060HL7 919934JMSER  00000016815HL3 919934JRAD1  00000016737HL9 919934JRAD2  00000018292HL3 920262RAD1  00000018502HL5 920262RAD1  00000080939HL2 920457PROT  00000048357HL8 920622PROT  00000019228HL6 920682PR1  00000034709HL6 920787PROT  00000029881HL0 920926RAD1  00000024737HL9 925734RAD1  00000019881HL2 928348PRO  00000030185HL3 928639RM2  00000024746HL0 928645RM1  00000072091HL2 931388CPXR  00000086104HL7 931752RAD  00000028786HL2 931967RM2  00000089695HL1 936348MVPRO  00000041812HL9 936756RAD1  00000048919HL5 937237PRO1  00000027843HL2 937278ER1  00000027850HL7 937278ER2  00000056090HL4 937298PRO1  00000063512HL8 937298PRO1  00000034780HL7 937352RM1  00000084367HL2 937382PROT  00000061395HL0 937384PRO1  00000053411HL5 937458PRO1  00000032400HL4 937498PRO1  00000062562HL4 940297PROT  00000036469HL5 940384PRO  00000030271HL1 940384RAD  00000004255HL6 940676RM3  00000042455HL6 940676RM3  00000036912HL4 940676RM4  00000043071HL0 941473RM3  00000046761HL3 941629RM5  00000017140HL5 952403RAD  00000088065HL8 952758PRO  00000028631HL0 954201RAD1  00000028637HL7 954201RAD2  00000099705HL6 954355PROTEUS  00000060064HL3 954385RAD  00000028246HL7 954476RAD4  00000087139HL2 956632XREC  00000036352HL3 970242RMORAD  00000030177HL0 970479XR2  00000056596HL0 970624XR1  00000056735HL4 970624XR2  00000041736HL0 970668XRA  00000088856HL0 970675XRA  00000082844HL2 970945ORTHO  00000035366HL4 970945RAD  00000039521HL0 970BASRAD  00000019955HL4 970SILTRAD  00000083637HL9 972230PRO  00000025487HL0 972420LR4  00000028679HL9 972420LR5  00000095074HL1 972566PEDIPRO  00000032207HL3 972566PROER1  00000016343HL6 972566PTOMO  00000032206HL5 972566RAD8  00000017152HL0 972596OPRAD1  00000017466HL4 972596OPRAD1  00000018295HL6 972596PER3  00000046525HL2 972875RAD  00000029127HL8 973365PROTEUS  00000023178HL7 973908RM8  00000083036HL1 973971ARI  00000083636HL1 973971ARI  00000023106HL8 973971OP  00000019157HL7 978249ER1  00000045523HL8 978355PRD  00000042477HL0 978521MVRAD2  00000034841HL7 979764PROT  00000045525HL3 979830BCRAD1  00000057996HL1 979865PRO2  00000032299HL0 985230DCRAD1  00000108376HL5 985230SCRAD1  00000027941HL4 985646RAD2  00000054532HL7 985646RAD3  00000060863HL8 985652ER  00000106224HL9 985867RAD1  00000077526HL2 989224PROTEUS  00000047008HL8 989723RAD  00000028378HL8 989839RAD  00000059952HL2 ALLIANCEXR113  00000114260HL3 FOU719390RAD  00000022588HL8 FWR4112  00000023839HL4 FWR4113  00000023872HL5 FWR4113  00000060273HL0 HUGHESPRO1  00000087775HL3 LMICPRAD1  00000087393HL5 MH589RAD1  00000029686HL3 UVRMCRM2  115529HL0 817BSFWPRO  116076HL1 618436R1  113347HL9 817FCARLPRO  89699HL3 760876RAD  117111HL5 850623NWFXR  1-165721296 519735PRO  114991HL3 303341XR  115530HL8 605842RAD  MISSING_SN_GON4219475 706335PRO1  00000102144HL3 870226PROT  00000103729HL0 0004034785  00000018689HL0 201358ER2  00000075125HL5 203453PROT  00000061517HL9 203688PROTLW  00000060398HL5 208882QCX1  00000044410HL9 210590OWRAD  00000041617HL2 210616PRO  00000045704HL4 210617NEI1  00000086106HL2 210703RAD1  00000019768HL1 212570RAD  00000033899HL6 217443UXR7  00000107243HL8 219836PRO1  00000037153HL4 248615PROT  00000015948HL3 248858ER  00000018690HL8 248937RM3  00000017463HL1 251435I02  00000052564HL2 251847RAD  00000023295HL9 252482CR2  00000035579HL2 252757ORTHO1  00000036131HL1 252757ORTHO3  00000034924HL1 253597XRM1  00000097810HL6 253680PROTEUS  00000035973HL7 253735MS1  00000075650HL2 253815PROT  00000018495HL2 256705XR1  00000028873HL8 256974PR1  00000050245HL0 256990FPR1  00000039750HL5 260373RM1  00000041261HL9 260426RAD  00000027944HL8 262249PROT  00000015888HL1 262268RAD1  00000017464HL9 262387RAD1  00000036992HL6 262642PROT  00000092360HL7 262646PRO  00000018603HL1 262741PR1  00000025421HL9 262741PR2  00000031794HL1 262928ER2  00000054166HL4 262948PRO1  00000061520HL3 269983PRO  00000025486HL2 270745RAD1  00000027947HL1 270745TOMO  00000017823HL6 276546PRORAD  00000019161HL9 276546PRORAD  00000020947HL8 276546PRORAD  00000071636HL5 276628PRO  00000019538HL8 281338RAD  00000106554HL9 281379FCYOPT  00000087392HL7 281395KATYPRO  00000093645HL0 281450FCP  00000064326HL2 281461PROT  00000057195HL0 281566ALCPRO  00000070005HL4 281576PROT  00000029233HL4 281599PRO  00000032395HL6 281837BPROD1  00000075126HL3 281923PWPRO1  00000047001HL3 281929PRORAD  00000076313HL6 281955RAD  00000100471HL2 281FCDPPROT  00000106547HL3 281FCERAPRO  00000113342HL0 281FCERSC220  00000104279HL5 281FCLPPROT  00000060464HL5 281OBMCPRO  00000077760HL7 281SLUTPRO  00000081383HL2 281UTSWPRO  00000054662HL2 281WHSLXRD  00000051586HL6 301209PRO  00000021505HL3 301290RAD1  00000079708HL4 301609RM2  00000022466HL7 301652RAD  00000062198HL7 301744RAD  00000049931HL9 301809RAD  00000036260HL8 301868R  00000103730HL8 301877RAD3  00000017454HL0 301896RAD1  00000017810HL3 302392PRO  00000021700HL0 302651PRO3  00000056259HL5 303426RM1A  00000023022HL7 303426RM2  00000030960HL9 303629RM05  00000023035HL9 303629RM06  00000031549HL9 303629RM08  00000028874HL6 303661RAD  00000022394HL1 303673TXR1  00000023034HL2 303673TXR2  00000105204HL2 303678FCFSXR  00000020953HL6 303699ER  00000070587HL1 303699RM3  00000103413HL1 303AFCXR  00000107786HL6 303FCBMOPTIMA  00000114262HL9 303NWMCXR  00000058405HL2 304253PROT  00000018385HL5 304424PRO  00000039696HL0 304424RM6  00000018323HL6 304424RM8  00000095422HL2 304598PROT  00000020067HL5 304599PRODA  00000019957HL0 304599PRODB  00000034541HL3 305345STRAD  00000034253HL5 307527PROTEUS  00000077287HL1 308425FPROT  00000041989HL5 308534NPOXR  00000099465HL7 308865KRRAD  00000030184HL6 308928RAD  00000036613HL8 309346PRAD1  00000033603HL2 309582PROT  00000016993HL8 309655FXR6  00000019597HL4 309655MORTON2  00000029530HL3 309663MHXR1  00000026059HL6 309664CBN2  00000068297HL1 313369PROT1  00000029288HL8 313640PROT1  00000045760HL6 314251SJBL  00000045792HL9 314251SJBR  00000037249HL0 314251SJMXR1  00000056595HL2 314739P1  00000081135HL6 314919JRAD1  00000062870HL1 315719PRO1  00000055252HL1 316685PRO1  00000049933HL5 316721WWRAD2  00000016409HL5 316722WWRAD  00000029086HL6 317888PRO1  00000039279HL5 317888PRO2  00000073628HL0 318259RM2  00000064174HL6 318443XRM1  00000029335HL7 318445LCXR1  00000070368HL6 318495WDR1  00000034712HL0 318878PRO  00000032393HL1 319235PROT  00000037749HL9 319642PROT  00000017467HL2 320654R1  00000024911HL0 320843RAD  00000086115HL3 321FCR1  00000016816HL1 3271539RAD  00000077530HL4 330297PX  00000031276HL9 330682ER  00000017468HL0 330740PX  00000080463HL3 330841PXD  00000111842HL1 331551DMGLORTHO  00000109667HL6 331551DMGLPROT1  00000107530HL8 331551DMGLPROT3  00000046882HL7 334244RAD2  00000018500HL9 334273RAD1  00000022473HL3 334636TRM1  00000028676HL5 334670RAD  00000050101HL5 336625R5203  00000053612HL8 336625R5207  00000098073HL0 337289SMRAD1  00000089288HL5 337363VR2  00000053199HL6 337531FRM2  00000038831HL4 337531FRM3  00000072363HL5 337531PRM5  00000044039HL6 352597RAD1  00000035182HL5 352689PRO  00000067465HL5 360397BGP  00000096320HL7 360397D8K  00000032496HL2 360412XR1  00000063175HL4 360417JTRAD1  00000063179HL6 360452RAD1  00000053542HL7 360476BANGOR  00000051288HL9 360476PSNS  00000016998HL7 360566RM1  00000028500HL7 360651XR1  00000042690HL8 360992CLK1  00000042384HL8 361698DPR1  00000022501HL1 361980RAP  00000101179HL0 386326ER  00000041259HL3 386FHIRAD  00000043171HL8 386HPORAD  00000031446HL8 386PORAD  00000031225HL6 386TWLRAD  00000029121HL1 401431PR1  00000020211HL9 401727P1  00000035302HL9 401766ARP1  00000064047HL4 402354MRAD  00000087142HL6 402393MERAD  00000088602HL8 402484AMIRAD  00000053200HL2 402745PROT  00000039280HL3 402768RAD1  00000086875HL2 402837CTCRAD  00000030412HL1 402947WRAD  00000037818HL2 404265ER1  00000035240HL1 404321PRO1  00000032761HL9 404466ED  00000016182HL8 404501P1  00000015898HL0 404501P2  00000049343HL7 404785PRO4  00000048257HL0 404785PROT  00000101400HL0 405271UPRO2  00000098076HL3 405271UPRO3  00000031854HL3 405360PRO  00000026060HL4 405379XR1  00000059950HL6 405636PRO  00000059577HL7 405682FTCPRO  00000074270HL0 405734TPR  00000072365HL0 406466TMC1  00000056597HL8 406653WXR  00000023538HL2 406748CXR  00000049347HL8 406768PXR  00000106550HL7 409267RAD1  00000036061HL0 409724RAD3  00000027849HL9 409727MJRAD2  00000063516HL9 409729PROT  00000040637HL1 409883RAD1  00000044038HL8 409899DCRAD1  00000086105HL4 410362ER  00000030726HL4 410601ER1  00000049346HL0 410604KIRAD  00000094526HL1 410787DRS8  00000029494HL2 410871RAD2  00000068786HL3 4109681  00000085826HL6 412372R1WDR  00000020590HL6 412624PROT1  00000087604HL5 412664ED1  00000045882HL8 412782PROT1  00000073181HL0 412784ORTHO2  00000019229HL4 412784PROT1  00000016344HL4 412784PROT3  00000042691HL6 412784PROT4  00000063694HL4 412788PROT  00000022748HL8 413499PROT1  00000023373HL4 413528R1  00000062564HL0 413738RAD1  00000019949HL7 414274PR1  00000022390HL9 414647RAD1  00000017825HL1 414805PRO2  00000070717HL4 414805PRO3  00000018611HL4 417269NX1  00000048625HL8 417354CMX  00000046431HL3 417461MRC  00000064169HL6 417625MSSU  00000090335HL1 417678ERXR  00000033829HL3 417820JX1  00000033716HL2 417820JX2  00000033958HL0 417820JX3  00000029628HL5 417820X1  00000029337HL3 417820X2  00000076948HL9 417847XR2  00000029126HL0 417887SNI2  00000033552HL1 419383RAD1  00000033387HL2 419407GE3  00000079709HL2 419479RAD  00000099140HL6 419547RAD1  00000078967HL7 419627RAD1  00000032885HL6 419843RAD  00000032813HL8 423431SDS  00000032294HL1 423431XR1  00000031273HL6 423431XR2  00000060102HL1 423586RAD  00000030014HL5 423778ERM1  00000030009HL5 423778ERM2  00000028289HL7 423778EXR2  00000025553HL9 423778EXR3  00000017617HL2 425228RC  00000019160HL1 425251VMA1  00000045630HL1 425374RAD1  00000037003HL1 425672RMB  00000074785HL7 425741RAD1  00000071150HL7 425885RAD1  00000022160HL6 432268RAD  00000104758HL8 432362RAD  00000062992HL3 434793RAD  00000087391HL9 435251NSHPC  00000062761HL2 435251OPPRM1  00000096689HL5 435528RAD  00000048801HL5 435635HCPRO1  00000053417HL2 435652PR1  00000086872HL9 435755PRO  00000050998HL4 435792RM1  00000083634HL6 435792RM3  00000100469HL6 435893RM3  00000088304HL1 435FCMCPRO1  00000086590HL7 440748PROTEUS  00000082523HL2 440967RAD1  00000037247HL4 440997PROT  00000031863HL4 443849ER2  00000086877HL8 469764MBPRAD  00000053413HL1 478275RAD2  00000035574HL3 478757CGTC  00000093303HL6 478PCHRAD  00000074786HL5 480345XR1  00000037380HL3 480598PRO  00000031269HL4 480603XR1  00000046336HL4 480730PROT  00000038830HL6 480984XR1  00000111898HL3 480MAYOXR  00000076947HL1 501202BNLER  00000080461HL7 501202BNLER  00000018694HL0 501268PROT  00000070589HL7 501268PROT2  00000059949HL8 501622PROT  00000047374HL4 502484PROT  00000041257HL7 502955RAD1  00000057949HL0 503216PRO  00000034865HL6 503224COPROT  00000074271HL8 503391SKY  00000024731HL2 503994RM2  00000103415HL6 504678PRO  00000018615HL5 504897SBER  000004-1188HL4 505727PROT  00000039602HL8 505763RAD1  00000086874HL5 505786PROT  00000045405HL8 505887RAD1  00000043961HL2 505887RAD2  00000048800HL7 505983PROT1  00000096227HL4 505DZXR  00000065131HL5 505LOBOXR  00000045880HL2 507212PRO  00000047010HL4 507212PRO  00000047428HL8 507274PRO  00000048111HL9 507274PRO  00000036696HL3 507283PROTEUS  00000090333HL6 507372PROTEUS  00000070887HL5 507407PROTEUS  00000020213HL5 507474PR1  00000087776HL1 507537PROTEUS  00000041101HL7 507723RAD  00000032814HL6 507836PROTEUS  00000050094HL2 508430FMC1  00000046763HL9 508548BRAD  00000025614HL9 508676PCR1  00000028374HL7 508676PCR2  00000022558HL1 508862CCH2  00000091771HL6 508996HRAD  00000113757HL9 508SANBR1  00000058225HL4 509247FAR1  00000055874HL2 509247FAR2  00000033900HL2 509276PROT  00000022601HL6 509332PRAD  00000033601HL6 509332PRAD  00000049932HL7 509525MCX1  00000052563HL4 509525MPROTEUS  00000049769HL3 509525VPROTEUS  00000105548HL2 509697SX1  00000063692HL8 509829VVX1  00000106220HL7 509966SX1  00000043859HL8 512887PR01  00000038408HL1 512887PR03  00000107528HL2 512FCBRAD  00000093644HL3 512FCJRAD  00000082841HL8 512LWMOBRAD  00000111894HL2 512SDERCPRAD  00000076310HL2 513233PRO  00000077529HL6 513389PRO1  00000023542HL4 513529XR  00000079182HL2 513853PRO1  00000064534HL1 513862WRPRO  00000040674HL4 513872PRO1  00000040892HL2 513872PRO2  00000063306HL5 513872PROER  00000110790HL3 515282XRA1  00000046673HL0 516674RM1  00000031862HL6 516676ER  00000023302HL3 517788PROTEUS  00000017472HL2 518243RAD1  00000017471HL4 518243RAD2  00000034240HL2 518254RAD3  00000034023HL2 518371RAD  00000113337HL0 518523PROT  00000043862HL2 518618RAD  00000033717HL0 520498PR1  00000033830HL1 520498PR2  00000016393HL1 520669XR  00000093305HL1 530528PROT  00000073848HL4 530BGPRO  00000056334HL6 540283RAD  00000080061HL5 540536UCARE  00000107531HL6 540591RAD  00000042903HL5 540741PRER  00000061396HL8 540741XR  00000041826HL9 540774VO  00000037688HL9 540776ROCLG  00000062868HL5 541276YX1  00000033426HL8 541322RM2  00000033553HL9 541322RM3  00000104755HL4 541342XR  00000080941HL8 541523PROTEUS  00000081136HL4 541548RM2  00000086873HL7 541549RM1  00000047231HL6 541OIT119A  00000047226HL6 541OIT119B  00000047067HL4 541OIT119C  00000047068HL2 541OIT119D  00000046766HL2 541OIT119E  00000047002HL1 541OIT134A  00000046883HL5 541OIT134B  00000046881HL9 541OIT134D  00000047005HL4 541OIT134E  00000046884HL3 541OIT134F  00000062993HL1 559459XR3  00000051287HL1 559683XR1  00000028380HL4 561207PBCC1  00000028638HL5 561207PBCC2  00000041102HL5 561627RAD  00000049590HL3 561650RAD  00000062866HL9 561881RAD  00000070367HL8 563252PR2  00000049131HL6 563387PRT2  00000021706HL7 563422PR  00000078166HL6 563568PROT  00000017821HL0 570409RAD  00000080460HL9 570455PRO  00000021185HL4 570455RAD1  00000104602HL5 570853PRO  00000023905HL3 571223RAD  00000056809HL7 573358R1  00000025061HL3 573442RAD  00000032264HL4 573556JCR4  00000064819HL6 573635JCOR1  00000022500HL3 573686ER  00000036060HL2 573729SRM1  00000028174HL1 574258PRO  00000048362HL8 574753PRT3  00000080732HL1 574753PRT4  00000096526HL9 575624RM1  00000082842HL6 580213VPRO  00000033424HL3 580220ROOM1  00000041993HL7 580371PRO  00000082527HL3 580481XRAY  00000063301HL6 580548BCPRO  00000056044HL1 580928PRO  00000079886HL8 585343RMC  00000028880HL3 586263MO2  00000043499HL3 586263RM1  00000061063HL4 586323PRO  00000066822HL8 586803PRO  00000031790HL9 601484PRO1  00000037589HL9 601583PROT1  00000042860HL7 601765PROT1  00000046429HL7 601783RAD1  00000061399HL2 601792PROT1  00000085130HL3 601876PROT1  00000020217HL6 601968COLRAD1  00000089290HL1 601968ERFP  00000029227HL6 601968PROT1  00000022914HL6 602220RM11  00000028296HL2 602246RM4  00000015953HL3 602249ERRAD  00000067235HL2 602251XR1  00000068441HL5 602251XR1  00000016188HL5 602285PROT1  00000017000HL1 602344RM1  00000018109HL9 602562PROT  00000086108HL8 602CIGNAPROT1  00000020955HL1 603788PROT  00000029229HL2 603882SJ4  00000015878HL2 603882SJ5  00000071925HL2 605328PROTEUS  00000046671HL4 605332PROTEUS  00000032295HL8 605342RAD  00000059948HL0 605347RAD2  00000061064HL2 605348RAD  00000044516HL3 605384PROTEUS  00000040980HL5 605428RAD  00000041210HL6 605462RAD  00000045108HL8 605532PROTEUS  00000035580HL0 605583PROTEUS  00000066385HL6 605665PROTEUSLC  00000038464HL4 605697PROTEUS  00000042902HL7 605697PROTEUS2  00000044646HL8 605698RAD  00000112431HL2 605755RAD  00000056339HL5 605835PROTEUS  00000057438HL4 605867RAD  00000031901HL2 605874RAD  00000029954HL5 605882PROTEUS  00000071487HL3 605996PROTEUS  00000020872HL8 606285PROT  00000047427HL0 606598FCI  00000080252HL0 606598XR1  00000042128HL9 606723PROT  00000050601HL4 606928PROT  00000031897HL2 607274EAST2  00000040769HL2 608258PROT  00000086588HL1 608326RAD  00000021191HL2 608364PR1  00000020588HL0 608796PR1  00000099139HL8 608WTCPRO  00000036792HL0 609426PROT  00000022749HL6 609465PROT  00000037717HL6 609625PROT  00000041135HL5 609646PROT  00000030410HL5 609652PROT  00000088599HL6 610565JOIPRO  00000031545HL7 610861PRO  00000083397HL0 610861PROFP  00000066675HL0 610JOIPROT  00000084009HL0 612313PRO  00000022155HL6 614234WESTPRO  00000045759HL8 614478PRO  00000023598HL6 614868PRO4  00000019232HL8 614898PROER  00000104750HL5 615435WRADOP  00000059848HL2 615446HER  00000104751HL3 615745PROT  00000023298HL3 616292PROTEUS  00000048493HL1 616365PROTEUS  00000023595HL2 616391PROTEUS1  00000030006HL1 616391PROTEUS3  00000042563HL7 616391WPROTEUS  00000050393HL8 616494PROTEUS  00000104281HL1 616685PROTEUS  00000042900HL1 616739PROTEUS  00000016189HL3 616752ER  00000043498HL5 616752M6PROTEUS  00000045408HL2 616752PROTEUS  00000034539HL7 617243PR1  00000080731HL3 617499PRAD  00000040588HL6 617MCOP1  00000069103HL0 618282RBRM2  00000103732HL4 618436R1  00000096155HL7 618664FCRAD  00000057777HL5 620257PRO  00000032300HL6 620285RAD  00000023260HL3 620327PRO  00000090859HL0 620332RAD  00000034686HL6 620724CX2  00000076043HL9 620786PRO  00000076509HL9 620872PRO  00000078551HL9 623876PROT  00000074784HL0 623876XR1  00000028175HL8 623882XR1  00000069843HL1 623925XR1  00000049765HL1 630323PROT  00000089291HL9 630351DOPROT1  00000046531HL0 630545PROT  00000050591HL7 630545PROT2  00000048492HL3 630545PROT3  00000048916HL1 630545PROT4  00000099909HL4 630551PROT2  00000111895HL9 630646PROT  00000049038HL3 630790PROT  00000041373HL2 630856PRO1  00000107790HL8 630871DMGWPROT1  00000075267HL5 630941PROT  00000021036HL9 630953LOBRAD  00000048518HL5 630968PRO  00000071146HL5 630968PROT  00000082186HL8 636239JR2  00000085136HL0 636239MOBR1  00000106555HL6 636390MSDR  00000025488HL8 636498MPR1  00000064171HL2 636530P1  00000098081HL3 636695MWR1  00000091293HL1 636789PDR  00000086817HL4 636866XR1  00000023172HL0 641295RAD1  00000052231HL8 641394PROT  00000064168HL8 641394PROT2  00000035374HL8 641424PROT  00000085133HL7 641446XRA  00000051825HL8 641456PROT  00000062991HL5 641484PROT  00000036911HL6 641664PROT  00000034344HL2 641754ERRAD  00000096319HL9 651254PRO  00000096154HL0 651265PRO  00000086586HL5 651293PRO  00000083218HL8 651439PRO  00000086114HL6 651789EPPA  00000064701HL6 651789PRO  00000098876HL6 651789PROVH  00000060368HL8 660530SFR2  00000088601HL0 660885WDR1  00000041733HL7 660890GVPRO  00000047006HL2 660890PROT3  00000017621HL4 661327RAD1  00000070714HL1 662285PRO  00000023904HL6 662377R1  00000062304HL1 662473R1  00000071147HL3 662726PRO1  00000053611HL0 671USNHPR  00000087607HL8 678289PRO  00000042327HL7 678318PRO1  00000086871HL1 678686PRO  00000045705HL1 701742RAD  00000059648HL6 701854RAD  00000099910HL2 702255XR2  00000038260HL6 702382XR1  00000023261HL1 702696XR  00000019149HL4 702731ORTH3  00000019148HL6 702731ORTHO1  00000018753HL4 702731ORTHO2  00000018814HL4 702731ORTHO4  00000080249HL6 702732MPXR1  00000056594HL5 702869XR1  00000057039HL0 702869XR1  00000056336HL1 702869XR2  00000057948HL2 702869XR2  00000043525HL5 703339RAD  00000083219KL6 703441RAD  00000017151HL2 703481RAD  00000019956HL2 703591RAD  00000034973HL8 703670ER1  00000103731HL6 703697PROT1  00000019159HL3 703751RAD  00000083058HL8 703849RAD1  00000057041HL6 704323PRO1  00000070888HL3 704323UPRO  00000065132HL3 704660PRO1  00000032662HL9 704660RAD1  00000033661HL0 704783PRO1  00000033600HL8 704783PRO2  00000023601HL8 704895RAD1  00000052076HL7 706245TR1  00000022156HL4 706554MSR1  00000053616HL9 706756RAD2  00000085824HL1 708237PROT1  00000107527HL4 708246PROT  00000057043HL2 708478PROT  00000036978HL5 708756ER1  00000083398HL8 708799PROT  00000022237HL2 708873XRA  00000058126HL4 708WSBCPROT1  00000065037HL4 712225XRA  00000086113HL8 712251PROTEUS  00000039491HL6 712737PROTEUS  00000041814HL5 712737PROTEUS  00000036861HL3 713442KSRAD  00000038443HL8 713463TCRAD  00000063805HL6 713547HCPROT  00000065483HL0 713566RAD  00000062473HL4 713566RAD2  00000073179HL4 713636HCRAD  00000106548HL1 713636SITIPRO  00000039753HL9 713662UTPROT  00000042079HL4 713770RAD  00000047488HL6 713948PMCRAD1  00000059773HL2 713982SCPROT  00000071634HL0 713HCALDPROT  00000084575HL0 713UTNWPRO  00000074012HL6 713UTPROTEUS  00000054531HL9 715241PR2  00000029689HL7 715241PRO  00000067055HL4 715284HRAD  00000050249HL2 715294PROT  00000051828HL2 715362PR1  00000042780HL7 715393PR5  00000065369HL1 715424PR1  00000061622HL7 715478PR1  00000063300HL8 715478PR1  00000037592HL3 715539PRO1  00000059576HL9 715588PR1  00000028577HL5 715644RAD  00000035153HL6 715672RAD  00000082524HL0 715793PR  00000057992HL0 715822RAD  00000017155HL3 715833PROD  00000029952HL9 715847PROTEUS  00000023268HL6 716633RAD  00000023405HL4 716633RAD1  00000075127HL1 716636PRAD1  00000077284HL8 716636PRAD1  00000053416HL4 716712RM1  00000054164HL9 716839RAD1  00000107787HL4 717270RAD1  00000086587HL3 717270RAD2  00000039370HL2 717464R1  00000088068HL2 717531MJR1  00000102865HL3 717561PRO1  00000032296HL6 717627RAD1  00000100280HL7 717761PRO2  00000058403HL7 717761PROT  00000029283HL9 717761RAD1  00000023168HL8 718267ER  00000059951HL4 718267PROTEUS  00000023368HL4 718567ER  00000055346HL1 718579RAD1  00000029449HL6 718604ER1  00000029022HL1 718604PROTEUS1  00000046672HL2 718604PROTEUS2  00000098080HL5 718840RAD2  00000107788HL2 719264RAD  00000046762HL1 719346PRO1  00000111469HL3 719487RAD  00000063806HL4 719584XR1  00000090332HL8 719638RAD  00000054935HL2 719667RM1  00000059844HL1 719686RAD  00000057435HL0 719SLVXR1  00000058603HL2 719SLVXR2  00000066674HL3 724224UCC  00000031610HL9 724230PROT1  00000022467HL5 724627PROT2  00000040988HL8 724779PROT  00000049588HL7 724779PROT2  00000080624HL0 724864PROT1  00000040779HL1 731659RF1  00000029724HL2 731784RM1  00000038055HL0 732223PRO  00000017455HL7 732360ER  00000034923HL3 732360PAT  00000021164HL9 732505RAD  00000094363HL9 732530URGW  00000015943HL4 734434PROT  00000043526HL3 734446PROT  00000077286HL3 734712AABJ  00000094199HL7 734712AABJ2  00000079512HL0 740351PROT  00000085134HL5 740506PROT  00000063515HL1 757275PRO1  00000056739HL6 757314RAD1  00000059649HL4 757314RAD2  00000045524HL6 757673ER  00000031898HL0 757673PRO1  00000050996HL8 757673PRO2  00000042562HL9 757673RM2  00000076700HL4 757822TCC118  00000077070HL1 757822TCC122  00000040432HL7 757889MM2  00000017818HL6 757889RM3  00000043858HL0 757889RM5  00000089698HL5 760379RAD  00000045546HL9 760830PRO  00000094656HL6 763587PRO  00000111843HL9 76",
      "center_classification_date": "20150112",
      "distribution_pattern": "USA:  All states in continental US including, Guam, DC.    OUS;  ARAB EMIRATES, AUSTRIA, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, ECUADOR, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, KUWAIT, LATVIA, LITHUANIA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PHILIPPINES, POLAND, POLYNESIA, PORTUGAL, RUSSIAN FEDERATION, SAUDI ARABIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, UKRAINE, UNITED KINGDOM, URUGUAY, VENEZUELA, VIETNAM.",
      "state": "WI",
      "product_description": "GE Healthcare, Proteus XR/a with wall stand model number 600-0301.",
      "report_date": "20150121",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare",
      "recall_number": "Z-0954-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70063",
      "termination_date": "20160203",
      "more_code_info": "3795PRO  00000057199HL2 765254PROT  00000071424HL6 765472RAD1  00000089292HL7 765472VA1  00000050390HL4 765674PROT  00000052237HL5 765674PROT1  00000032881HL5 770229GT  00000070209HL2 770297NDR2  00000023300HL7 770751PRO  00000081880HL7 770838TDVAPRO  00000045614HL5 770965BR2  00000080253HL8 770968PRO  00000034669HL2 770971PROT  00000015893HL1 773265PROT  00000032298HL2 773265PROT2  00000020866HL0 773282TOMO8  00000075399HL6 775SLYONXR  00000071923HL7 785263PRO  00000033377HL3 785434PRO  00000059717HL9 785475PRO  00000015887HL3 785562PRO  00000055684HL5 801273XRAY  00000054412HL2 801294XRAY  00000060396HL9 801302MOBRM1  00000037747HL3 801302RM1  00000045761HL4 801321OSG  00000060061HL9 801341SS1  00000062301HL7 801387RM1  00000041886HL3 801489XR  00000043589HL1 801501XR  00000057774HL2 801525PRO  00000114259HL5 801525PRTS  00000015867HL5 801581RM6  00000070208HL4 801627PRO  00000020584HL9 801734RM1  00000064531HL7 801777RM1  00000068443HL1 801777RM2  00000049039HL1 801779RAD  00000087140HL0 801779RAD2  00000090857HL4 801786XR2  00000085132HL9 801871XR1  00000044042HL0 801972WKMD  00000099141HL4 801AFP1  00000087143HL4 801RCXR1  00000031226HL4 803358UCR1  00000031270HL2 803536R1  00000028519HL7 803541BRAD  00000037576HL6 803581RM2  00000024732HL0 803749R1  00000035917HL4 803895RAD  00000042199HL0 804217RAD1  00000043173HL4 804287PRO1  00000096690HL3 804325RSPRO  00000078167HL4 804595XR1  00000078168HL2 804595XR1  00000041580HL2 804639RAD  00000106221HL5 804893RAD  00000076311HL0 804893RAD1  00000076698HL0 804893RAD1  00000054534HL3 804HDMCRM2  00000059774HL0 805737RAD1  00000036120HL4 805PIXRA  00000020206HL9 806212RAD1  00000079888HL4 806355RAD1  00000095426HL3 806663RAD  00000062019HL5 808448XRA  00000030415HL4 808696RM1  00000098075HL5 808873PROA  00000066963HL0 808876PRO  00000050391HL2 810342ER1  00000025121HL5 810342RMB  00000023870HL9 810762ER1  00000050594HL1 810985PRO1  00000022327HL1 812242APN1  00000022240HL6 812242APN2  00000036987HL6 812242APN3  00000075817HL7 812242PROT  00000086593HL1 812242UAP1  00000089904HL7 812242UAP2  00000053540HL1 812343PROT  00000026638HL7 812450ER1  00000041449HL0 813253PRO  00000031769HL3 813964PRO  00000032888HL0 814255R2P  00000084773HL1 815300PVPROT  00000083635HL3 815395OSFER1  00000107789HL0 815395OSFXRF  00000060366HL2 815462PROT1  00000017456HL5 815842JR1  00000048113HL5 816348PROT  00000087777HL9 816348PROT2  00000033957HL2 816561DICPRO  00000049589HL5 816HILLYARD  00000086816HL6 817336PRO  00000035025HL6 817346RAD  00000034971HL2 817346RAD1  00000025562HL0 817347HPROER  00000040838HL5 817543ACC  00000038258HL0 817605CCHRAD  00000084771HL5 817644FCRAD  00000037346HL4 817685ER2  00000022555HL7 817685HEBTOMO1  00000022753HL8 817685HEBTOMO2  00000046427HL1 817837PRO  00000102866HL1 817995PRODR  00000089902HL1 818360XR1  00000056335HL3 828433JIP  00000044379HL6 830816RAD  00000051829HL0 830896VAPRO  00000057950HL8 843228MCASRAD  00000043812HL7 843228PR1  00000033642HL0 843784RAD  00000107534HL0 843867RAD1  00000109666HL8 843PPCELMSR1  00000034239HL4 844336PROT  00000035578HL4 844336PROT2  00000019876HL2 845703RAD  00000027848HL1 845703RAD2  00000037929HL7 845938RM2  00000078966HL9 847292PROT  00000060062HL7 847543CLCA  00000059335HL0 847543CLCB  00000059574HL4 847543CLCF  00000024744HL5 850431ER1  00000024734HL6 850431ER2  00000104280HL3 850623NWFXR  00000081139HL8 850654PRO  00000062017HL9 850696WXR1  00000101180HL8 850862PRO  00000088855HL2 856234PROT  00000053346HL3 856363PROT  00000070004HL7 856451PROT  00000020353HL9 859226ER  00000043494HL4 859987PROT  00000068620HL4 860667RAD1  00000020768HL8 863419ER  00000113336HL2 864299XRA1  00000037117HL9 864560XRA1  00000039278HL7 864898RM1  00000062100HL3 865525RAD  00000064323HL9 865573XR1  00000036610HL4 870972PRO1  00000037119HL5 870972PRO2  00000106223HL1 901NAVY  00000072093HL8 903886PRO1  00000065129HL9 903TOPRO  00000086107HL0 904230RM1  00000032492HL1 904858MRAD  00000051823HL3 904928RAD  00000111892HL6 904998RM1  00000039954HL3 907234PROT  00000023837HL8 907796ORTHO  00000022331HL3 907835PROTEUS  00000081877HL3 907861VER  00000034093HL5 908522PRO  00000044744HL1 910450CHBAY  00000045629HL3 910450NHCL3  00000044740HL9 910450NHCL6  00000090334HL4 910904CFHPROT  00000046337HL2 910907WA5  00000062990HL7 910907WACC1  00000059341HL8 910907WAJOEL1  00000038444HL6 910907WAPROT  00000059845HL8 910907WARM6  00000052230HL0 910907WARM7  00000065130HL7 912287OTC  00000101038HL8 913397PROT  00000068622HL0 913785FDR1  00000036468HL7 915533RAD1  00000062302HL5 915577ERRAD  00000087605HL2 915869RAD  00000028983HL5 916278PROT1  00000039281HL1 916351PRO1  00000043330HL0 916575PROMN  00000023110HL0 916781PRO  00000032559HL7 916781PROT1  00000041618HL0 916781PROT2  00000028034HL7 916874PRO1  00000028040HL4 916874PRO3  00000028037HL0 916874PRO4  00000045111HL2 916LAGPRO1  00000077285HL5 916MCCPRO  00000019947HL1 918287PRAD  00000043959HL6 918294MPPRO  00000046339HL8 918294MPPROT  00000054533HL5 918392PROTEUS  00000041477HL1 918486PROTEUS  00000094652HL5 918567PROT1  00000048798HL3 918599XR112  00000046626HL6 918599XRER  00000100278HL1 918647PROT1  00000028521HL3 918787GRAD1  00000028980HL1 919572SWSM1  00000020963HL5 919781RAD1  00000016986HL2 919787WRAD2  00000023114HL2 919954PROT1  00000054165HL6 919954RM10PROT  00000079711HL8 919BCPROT  00000017826HL9 920288RAD1  00000018386HL3 920288RAD2  00000018388HL9 920288RAD2  00000028639HL3 920303ER  00000028780HL5 920303PROT  00000022585HL4 920327RAD  00000028787HL0 920451RAD1  00000036790HL4 920720PRT  00000020215HL0 920738RAD1  00000020685HL4 920738RAD2  00000020721HL7 920738RAD3  00000039055HL9 920793PRO1  00000038562HL5 920793PRO2  00000034432HL5 920866PR1  00000016592HL8 920952ER1  00000016736HL1 920952RAD2  00000031900HL4 925210RAD1  00000015952HL5 925210RAD2  00000015949HL1 925210RAD3  00000022058HL2 925370ORTHO1  00000022063HL2 925370ORTHO2  00000087389HL3 928453UCXR1  00000038832HL2 928725RM1  00000105864HL3 928737RAD  00000038463HL6 928758RAD  00000020076HL6 931403RAD1  00000020068HL3 931520PROT1  00000021035HL1 931520PROT2  00000074014HL2 936825GSJRAD1  00000042284HL0 937268R113  00000059269HL1 937440PROH  00000088307HL4 937440PROH2  00000039367HL8 937440PRON  00000038465HL1 937440PROS  00000041136HL3 937440RM3  00000069101HL4 937440RM4  00000084369HL8 937440SPRO1  00000041645HL3 937558PRO  00000050246HL8 937558SCPRO  00000086591HL5 937OVXR  00000081380HL8 940320PROTEUS  00000047519HL4 940898PROD  00000064172HL0 949517RM4  00000086818HL2 952541PRO  00000114263HL7 952738PRO  00000033586HL9 952944RAD  00000046428HL9 956366RAD1  00000028476HL0 956632MCRAD1  00000038035HL2 970453XR1  00000032884HL9 970565RAD1  00000022757HL9 970668XR  00000041648HL7 970DOAXR1  00000035154HL4 972331RAD1  00000055347HL9 972487BGDX2  00000036466HL1 972487BGOP1  00000044035HL4 972487BGPRO  00000041172HL8 972487BGRAD  00000034842HL5 972566DMSRAD1  00000033875HL6 972665BPPR1  00000033898HL8 972665BPPR2  00000034946HL4 972665BPSC1  00000034866HL4 972665BPSC2  00000033671HL9 972665DDA1  00000033714HL7 972665DDA2  00000084774HL9 972744PRO1  00000087608HL6 972899FMPRO  00000043021HL5 972935ERPRO  00000041828HL5 972981PRO10  00000037343HL1 972981PROTEUS1  00000042325HL1 972981TOMO9  00000045282HL1 973331PRO  00000041579HL4 973439PRO  00000038257HL2 973503PRO  00000029129HL4 973625RM2  00000040431HL9 973746PRO  00000029538HL6 973989RM2  00000058404HL5 978372WRAD  00000088303HL3 979245PRO  00000072528HL3 980487PRO1  00000035023HL1 985732RAD1  00000072529HL1 989246RAD  00000045793HL7 989246RM2  00000088067HL4 989343ER  00000018815HL1 989352RMB  00000035981HL0 989802PRO1  00000035371HL4 989839RM5  00000065035HL8 CVSPRAD1  00000109668HL4 FCERPL200  00000111891HL8 FCFRISCOPROT  00000102139HL3 FCGARX-RAY  00000103417HL2 FCMESQPROT  00000106553HL1 FCRICHPROT  00000053315HL8 FHAPKPRO  00000044029HL7 FHEOPROTEUS3  00000022557HL3 FHORLER1  00000023170HL4 FHORLER2  00000051589HL0 FHORLPRO4  00000048624HL1 FHWPPRO3  00000046526HL0 GAFBXRA  00000048005HL3 INSIGHTXR107  00000086592HL3 MC532RAD1  00000112429HL6 PMUC7B  00000081589HL4 WC231RM1  00000077528HL8 WC231RM4  00000022388HL3 WH231RAD1  114269HL4 605322CLIFFRAD  00000053412HL3 205481PRO  00000053538HL5 205592DCPRO  00000045210HL2 205750RAD  00000086876HL0 205795PRO  00000037575HL8 206248RAD1  00000113339HL6 206248RAD3  00000082322HL9 206288XR1  00000035237HL7 206302XRM1  00000104282HL9 207255RAD1  00000045795HL2 207438KRAD  00000016184HL4 207454CRAD1  00000073627HL2 207454RAD2  00000053617HL7 207662PRO  00000074527HL3 207664PROTEUS  00000077762HL3 207695PROXR  00000071635HL7 207794RAD2  00000039695HL2 207795CM3  00000059338HL4 207973PROTEUS  00000018756HL7 208227RM1  00000019410HL0 208227RM5  00000040540HL7 208265PROT  00000020071HL7 208359RM2  00000020723HL3 208373RAD2  00000023269HL4 208373RAD3  00000019231HL0 208381ED2  00000102868HL7 208402RAD1  00000042863HL1 208455RAD3  00000042194HL1 208455RAD4  00000035581HL8 208463RAD1  00000015875HL8 208463RAD2  00000029231HL8 208476CRAD  00000104065HL8 208524RM3  00000037591HL5 208524ROOM2  00000015942HL6 208529ER  00000038334HL9 208667KS  00000071922HL9 208678PROTEUS1  00000028682HL3 208746VX1A  00000608866HL1 208882IX1  00000030965HL8 208962SRAD  00000024470HL7 208983SRAD  00000043397HL9 209223PROTAM1  00000023266HL0 209385RAD1  00000024368HL3 209524GR5  00000022586HL2 209577RAD3  00000042688HL2 209754RAD1  00000029281HL3 209832RAD1  00000047229HL0 209835RAD2  00000020769HL6 209848OVRAD2  00000041213HL0 210292LRCR1  00000028979HL3 210297BMCR5  00000025613HL1 210297NERAD  00000028575HL9 210297SEBR4  00000114264HL5 210403MSPRO  00000105865HL0 210447ER  00000046134HL3 210575LCRAD  00000081382HL4 210590RAD1  00000029536HL0 210616MCTRAD2  00000029813HL3 210616R2  00000056738HL8 210617NWPRO  00000043175HL9 210617PRO  00000028677HL3 210646NEMR1  00000037574HL1 210646RAD2  00000016585HL2 210704SRER  00000102502HL2 210877PRO  00000042478HL8 210921PRO  00000044380HL4 210925KRM177  00000044382HL0 210925KRM185  00000071924HL5 210MIRAD1  00000021398HL3 212241RAD  00000023112HL6 212241RAD  00000021395HL9 212604PROT3  00000021893HL3 212604PROT3  00000019405HL0 213744PROTXR2  00000019596HL6 213744PROTXR2  00000064045HL8 214266IRVPRO  00000036576HL7 214266RAD1  00000051937HL1 214268RAD1  00000016400HL4 214320PTOMO  00000021323HL1 214345RAD1  00000071488HL1 214370BCCPRO1  00000090860HL8 214491RPG  00000059715HL3 214590UROPRO  00000081591HL0 215245PROFP  00000036599HL9 215441PROT1  00000024475HL6 215453PRO3  00000042133HL9 215481PROT1  00000052236HL7 215827PROT1  00000016819HL5 215952PROT  00000043861HL4 215955PROER  00000102504HL8 216445PROTEUS  00000104754HL7 216445PROTEUS  00000047515HL2 216778RAD1  00000024659HL5 217222QR6  00000036243HL4 251368ARAD  00000023400HL5 252209RCR3  00000036063HL6 252436RAD1  00000036096HL6 252436RAD2  00000035974HL5 252443NGER1  00000036550HL2 252443NGER2  00000044033HL9 252493RPCC  00000019877HL0 252522LRAD1  00000036281HL4 252637ECR2  00000032262HL8 252794BR2  00000100701HL2 252809MGRAD  00000031861HL8 252847EDR1  00000031899HL8 252847EDR2  00000037483HL5 252847PRE1  00000042285HL7 252847RAD8  00000042324HL4 252847RAD8  00000080062HL3 253841DINW1  00000033956HL4 253ECN2420RM4  00000036192HL3 253ECN3535RM2  00000065036HL6 253ECN47859RM5  00000036864HL7 254288DAR2  00000044518HL9 254288DAR4  00000021705HL9 254562RAD2  00000016600HL9 254897GRRAD  00000016735HL3 254897GRRAD  00000017809HL5 256235ORT1  00000064043HL3 256332RRM1  00000061620HL1 256356RAD1  00000053535HL1 256386RM4  00000082187HL6 256429CR2  00000073849HL2 256782PRO1  00000090522HL4 256CCP1  00000020716HL7 262334RAD1  00000060367HL0 262569PRO  00000079179HL8 262697PRO1  00000076419HL1 262697PRO2  00000016824HL5 262785RM1  00000047062HL5 262884PRO1  00000049034HL2 262896PRO1  00000063695HL1 269327PROTEUS  00000050597HL4 269381PROTEUSA  00000052225HL0 269381PROTEUSB  00000046049HL3 269463PROTEUS  00000023834HL5 269637PROTEUS  00000051939HL7 269657PROTEUS  00000051664HL1 269683PRO  00000025422HL7 269945PROTEUS  00000046338HL0 269968PROTEUS  00000015954HL1 270259R3  00000030413HL9 270444LR4  00000030408HL9 270444MNPOC  00000090521HL6 270483GPROT  00000045942HL0 270798AHC  00000052239HL1 270798ORTHO  00000059446HL5 270798ORTHO2  00000040539HL9 270798RAD  00000041616HL4 270798RAD3  00000047227HL4 270798RAD7  00000053615HL1 270956LR  00000055873HL4 270956LR  00000057040HL8 270965LR2  00000076042HL1 276666RM4  00000022892HL4 276964PRORAD1  00000046669HL8 281348PROTEUSER  00000017458HL1 281558PRO2  00000024908HL6 281588PROER  00000043329HL2 281897MHPRO  00000107532HL4 281ACERSPROT  00000080730HL5 281FCPROTEUS  00000107536HL5 281HHVPRO  00000072983HL0 281WHGPPRO  00000043070HL2 301609ER  00000025063HL9 301698ER2  00000024681HL9 301698RM3  00000060600HL4 301874UBRAD  00000029280HL5 302421PRO  00000036693HL0 3024307  00000052080HL9 302653RAD1  00000052228HL4 302674RAD1  00000054535HL0 302684RAD  00000042781HL5 302744ER  00000023177HL9 303320ER1  00000016402HL0 303450ED  00000045632HL7 303563XR  00000022157HL2 303730RM5  00000030414HL7 303778EDXR  00000073388HL1 304373R2  00000032394HL9 304598PROT1  00000033385HL6 304624PRT  00000031611HL7 304792PROT1  00000041735HL2 304792PROT2  00000069100HL6 304855PROT  00000041992HL5 304927R2  00000041992HL9 304927R2  00000023115HL9 305295R1  00000040960HL7 305325RM9  00000031150HL6 305682ER1  00000064048HL2 305FISHPRO  00000094364HL7 305MCHDORAL  00000089697HL7 305MRCOPRM1  00000016183HL6 307265GERAD  00000057042HL4 307578MOBRAD  00000038280HL4 307634XRA  00000015874HL1 309343CR2  00000046880HL1 309343CRAD1  00000028520HL5 309454BER  00000060101HL3 309454BX3A  00000064533HL3 309647GMGR1  00000031155HL5 309662JER  00000016406HL1 309944HHPROT  00000015945HL9 313359PROT  00000027950HL7 314577CP1  00000027845HL7 314577CP2  00000019952HL1 314625WER  00000060462HL9 315255XRAY1  00000061886HL8 315255XRAY1  00000036424HL0 315332NR4  00000057437HL6 315332RM3  00000044645HL0 315338ER  00000016593HL6 315782RAD1  00000016590HL2 315782RAD2  00000088600HL2 315787TB1  00000017157HL9 316689WC3  00000017145HL4 316689WC4  00000018489HL5 316689WC4  00000089289HL3 316962ERPRO1  00000064817HL0 316962ORTHO  00000042862HL3 316962PRO  00000054415HL5 316962WCPRAD  00000034303HL8 317621PROT  00000020066HL7 317745HEPR5  00000019408HL4 317745HEPR6  00000021328HL0 317745HEPR7  00000040979HL7 318251PRO  00000029039HL5 318254PRO  00000030645HL6 318335ORM1  00000048112HL7 318456BRM3  00000048522HL7 318648RM2  00000029120HL3 319268PROT  00000025377HL3 319268SER  00000015897HL2 319376PROT  00000029656HL6 319863PROT  00000061263HL0 320243RAD1  00000017150HL4 320632RAD  00000062472HL6 320693RAD2  00000024203HL2 325670PRO2  00000029448HL8 325670PRO3  00000037156HL7 325696RAD1  00000037250HL8 325696RAD2  00000016734HL6 330332R4  00000040636HL3 330363RAD1  00000039369HL4 330363RAD2  00000043396HL1 330386ER  00000029128HL6 3304382  00000041371HL6 330543XRA  00000046532HL8 330707PX  00000040889HL8 330834RAD  00000023409HL6 332047  00000077527HL0 333IMM  00000046759HL7 333PRO  00000033384HL9 336372ARAD1  00000043809HL3 336719NR140  00000082321HL1 336846AMRAD  00000083632HL0 337289RAD  00000039583HL0 337457RAD1  00000034559HL5 337981ER1  00000028578HL3 337981R3  00000033498HL7 337981RAD3  00000077764HL9 33RM4  00000040538HL1 352253RAD1  00000096686HL1 352596PROM  00000040990HL4 352596XRA  00000019688HL1 352861RAD1  00000083056HL2 360346PROTEUS  00000042298HL0 360414OTHO  00000036323HL4 360493PROT  00000046764HL7 361275RM1  00000099137HL2 361293PROT  00000027095HL9 361572CER  00000032556HL3 361857DOR1  00000044409HL1 361888CCOPC  00000052226HL8 373ORTHO  00000023299HL1 386238FHR1  00000024469HL9 386238FHR3  00000028035HL4 386446RM1  00000028984HL3 386446RM2  00000035980HL2 386774RM1  00000034561HL1 386774RM2  00000016581HL1 401274ER1  00000057947HL4 401348WR1  00000022760HL3 401782SR2  00000029727HL5 401782SR3  00000057194HL3 402269RAD1  00000016818HL7 402481BX3  00000016186HL9 402481LGMX10  00000032263HL6 402643SMX1  00000020352HL1 402685RAD  00000073178HL6 402878PHSRAD  00000024428HL5 402925RAD  00000022507HL8 404616PRO1  00000062018HL7 404686CRAD5  00000056737HL0 404686MOT2  00000029226HL8 404778EP1  00000080938HL4 405262EFCIRAD  00000054414HL8 405271ETPRO  00000043639HL4 405271PRO  00000044311HL9 405271UPRO  00000017623HL0 405271UPX1  00000106222HL3 405307PRO  00000069327HL5 405359PBPRO  00000016405HL3 405372SP1  00000016404HL6 405372SP2  00000040891HL4 405527RAD  00000016813HL8 405736MR4  00000034664HL3 405927R1  00000102143HL5 405949ER1  00000018693HL2 406222LR2  00000036549HL4 406228FMRAD3  00000027945HL5 406238BCW1  00000071926HL0 406327PLX1  00000065368HL3 406363PROT  00000101781HL3 406363XRP  00000064535HL8 406543SX3  00000032297HL4 406547MVR1  00000091772HL4 406585PXR1  00000091104HL0 406585PXR2  00000031268HL6 406654PXR  00000023175HL3 406657BCH  00000062563HL2 406665BHR1  00000045432HL2 406728PRT  00000022161HL4 406782SJER  00000028570HL0 406782SX1  00000072530HL9 406787RRAD  00000042783HL1 406842RVXR  00000044310HL1 406863RAD  00000078008HL0 406873PXR  00000023375HL9 407296ERP  00000071318HL0 407296RM2  00000078169HL0 407296RM2  00000049930HL1 407321ERPRO1  00000032763HL5 407518PR1  00000049766HL9 408559RAD12  00000021194HL6 408559RAD2  00000042782HL3 408636PROTEUS1  00000028036HL2 408885EDA  00000025060HL5 408885EDB  00000052077HL5 408885GIL  00000064173HL8 408885MIL  00000056257HL9 408885MOOR  00000038406HL5 408885SVC  00000035372HL2 408885TUL  00000081134HL9 408885VHC1  00000052073HL4 408885VSC2  00000054167HL2 408885VSC3  00000086111HL2 408957ECF1  00000070006HL2 409212RAD1  00000050389HL6 409384RAD1  00000020072HL5 410392R3  00000015889HL9 4105462  00000029688HL9 410641ER  00000061621HL9 410740R1  00000064046HL6 410740R2  00000072984HL8 412232ER1  00000076507HL3 412232ER2  00000020867HL8 412367PROT1  00000111659HL9 412367XR1  00000068950HL5 412623PROT5  00000062099HL7 412623UCC  00000018114HL9 412641PROT1  00000042527HL2 412641R1  00000088302HL5 412647WDR1  00000031904HL6 412692PROT1  00000032706HL4 412692PROT2  00000105546HL6 412942WDR  00000019230HL2 414527PR1  00000029786HL1 414647RAD2  00000029496HL7 414761RAD1  00000100277HL3 414774PRO  00000042454HL9 417256XR2  00000096524HL4 417837XR1  00000026636HL1 417883SG23  00000076312HL8 419228PR  00000112817HL2 419238RAD  00000036577HL5 4193543  00000023369HL2 419383HCRAD  00000034844HL1 419394ER  00000069102HL2 419394RAD2  00000039584HL8 419399RAD  00000079710HL0 419538RAD  00000020957HL7 419578RAD  00000086112HL0 419578RAD3  00000050096HL7 419586PR  00000021704HL2 419592RAD2  00000015947HL5 419666RAD1  00000016739HL5 419692RAD  00000048797HL5 419738PR  00000034455HL6 419891ER  00000034922HL5 419891RM2  00000108377HL3 419946PR1  00000019879HL6 423493ER  00000025067HL0 423542RAD1  00000019690HL7 423778EE1  00000019772HL3 423844R1  00000069330HL9 423844R2  00000052234HL2 432837RAD1  00000039605HL1 434924HW2  00000071315HL6 434CCRAD  00000091295HL6 434SCHRM3  00000057196HL8 434SCHRM4  00000056812HL1 435207RM1  00000028574HL2 435251RM3  00000028977HL7 435251RM5  00000032663HL7 435462XR  00000060063HL5 435628RROPRO1  00000024738HL7 435637RM1  00000029687HL1 435676RAD  00000046528HL6 435688RRPRO1  00000073387HL3 435716RM4  00000072706HL5 4357792RM2  00000023541HL6 435865PRO1  00000023297HL5 435865PRO2  00000088860HL2 435DCMCPRO1  00000088305HL8 440743PROTEUS  00000029725HL9 440827ER1  00000023836HL0 440988RAD1  00000062760HL4 443481OSM1  00000021703HL4 443849ER  00000031856HL8 443849ER1  00000025416HL9 443849RAD4  00000020591HL4 478633EC1  00000038033HL7 502259ER  00000039697HL8 502624R7  00000017153HL8 502726RAD1  00000015892HL3 502893ER  00000024426HL9 502897TOMO  00000031546HL5 503215PROTEUS  00000103418HL0 503399PRO  00000062196HL1 504371PRO  00000034454HL9 504391ER  00000020070HL9 504391R3  00000044037HL0 504447RAD1  00000041448HL2 504464ER  00000036910HL8 504842OUC  00000025177HL7 504853RAD1  00000055872HL6 504897OBED  00000031768HL5 505262PROT1  00000099908HL6 505426PROT  00000020864HL5 505522RAD1  00000020958HL5 505522RAD2  00000021895HL8 505538RAD1  00000024371HL7 505662PROT  00000033713HL9 505662PROT2  00000049768HL5 505727PROT101  00000037116HL1 505743RAD1  00000057440HL0 505893PROT  00000034458HL0 505DULCEXR  00000080623HL2 507385EDRAD1  00000047330HL6 507385NRRAD1  00000035373HL0 507385RAD  00000045364HL7 507385RAD1  00000020074HL1 507434RAD1  00000065133HL1 507526RAD1  00000062471HL8 507837RAD  00000059270HL9 507931RAD1  00000106551HL5 508457FR1  00000039585HL5 508460RAD1  00000041508HL3 508548FR2  00000081879HL9 508833SHRAD  00000044043HL8 508862R11  00000083055HL4 508FHCMRAD  00000105549HL0 508NOFALRAD  00000023108HL4 509525WX1P  00000068953HL9 509575YRX1  00000052081HL7 509633CX3  00000029953HL7 509682LX2  00000056042HL5 509754CX2  00000041260HL1 509758TRAD2  00000076134HL6 509786RAD1  00000041404HL5 509865TCX1  00000048360HL2 509962K2X  00000039057HL5 509SHRINPROT  00000024907HL8 510437PROT1  00000024912HL8 510437PROT2  00000017003HL5 510655SM11  00000019780HL6 510667PROTEUS  00000042197HL4 510797RM1  00000025176HL9 510797W2  00000029447HL0 510797W5  00000033554HL7 512397RAD2  00000033587HL7 512397RAD3  00000036425HL7 512447PRAD  00000082326HL0 512BASRAD  00000075818HL5 512BCRAD  00000074272HL6 512PFERRAD  00000026064HL6 513584PRO1  00000026635HL3 513584PRO2  00000032555HL5 513584RM9  00000019774HL9 513636EGXRAY  00000022324HL8 513636OPAPRO  00000040981HL3 513636PRO  00000028293HL9 513745PRM2  00000028097HL4 513745RM1  00000059449HL9 513BNOPRO1  00000059452HL3 513BNOPRO2  00000030913HL8 515222XRA  00000034540HL5 515222XRA2  00000060983HL4 515278IHPXRA  00000036551HL0 515382XRA  00000034429HL1 515643XRA  00000051938HL9 515832XRA  00000029537HL8 516520RM1  00000028372HL1 516632ER1  00000052235HL9 516632RM2  00000048521HL9 516632RM5  00000113338HL8 517334ER  00000035141HL1 517543PROTEUS  00000024580HL3 517663PROTEUS  00000032495HL4 518382RAD6  00000033386HL4 518382RAD8  00000016198HL4 518583RAD5  00000083633HL8 518697RAD1  00000101402HL6 518725RAD  00000028033HL9 518828RM1  00000035181HL7 520228PRO  00000029119HL5 520432RM2  00000039490HL8 520533PRO1  00000043497HL7 520533XR3  00000064170HL4 520STJOPRO  00000084007HL4 530225PROT  00000020956HL9 530662PRO  00000049127HL4 530662PRO2  00000063176HL2 530BAFPRO  00000016731HL2 540228RAD  00000019543HL8 540462RAD  00000048358HL6 540536ER1  00000049126HL6 540636RAD1  00000034306HL1 540776PR2  00000034304HL6 540776PR3  00000023171HL2 540862RAD1  00000023871HL7 540862RAD5  00000045706HL9 540953PRO  00000020950HL2 540981RX1  00000077288HL9 541276AOPROTEUS  00000038563HL3 541390RM1  00000038441HL2 541390RM2  00000045366HL2 541426WX1  00000097811HL4 541447CRM1  00000025561HL2 541451RM2  00000037746HL5 541677MOIRAD1  00000040780HL9 541768RM5PROT  00000047331HL4 559459XR1  00000048361HL0 559459XR2  00000083059HL6 561246RAD  00000019153HL6 561395RAD  00000030443HL6 561488RAD1  00000030407HL1 561488RAD3  00000042476HL2 561540RAD  00000022503HL7 561798ER1  00000029532HL9 561863RM6  00000088859HL4 561965RAD1  00000031442HL7 562493RM1  00000026063HL8 563589PROT  00000044515HL5 563652PROT  00000025563HL8 570286ER  00000029932HL1 570325PRO  00000046760HL5 570387PRO1  00000033376HL5 570523PRO1  00000060984HL2 570759RAD3  00000070211HL8 571231RAD1  00000083219HL6 571231RAD2  00000023404HL7 573486HR2  00000043528HL9 573582AUDRAD2  00000036574HL2 573756ER  00000065485HL5 573882RAD1  00000041505HL9 573884WXR  00000050387HL0 574535PRO  00000029284HL7 574537RAD  00000082190HL0 574546RAD1  00000028882HL9 574772PRO  00000087606HL0 575521RM1  00000085131HL1 575521RM5  00000030178HL8 580233ER1  00000018988HL6 580233PRO  00000041372HL4 580249SPRO  00000041258HL5 580249SPRO2  00000046530HL2 580323PRO  00000034661HL9 580335RAD1  00000019155HL1 580336PRO  00000029025HL4 580355RAD1  00000029286HL2 580355RAD2  00000029287HL0 580355RAD3  00000042456HL4 580548PRO  00000018499HL4 580822PRO  00000028573HL4 580938SPRO  00000079181HL4 585266RAD1  00000080942HL6 585266RAD1  00000058406HL0 585368RAD  00000022238HL0 585798RAD1  00000028684HL9 586263MO1  00000051942HL1 586366PRO  00000071316HL4 586573EDPROT  00000024582HL9 601200RM1  00000038332HL3 601288PROT1  00000019147HL8 601364RM3  00000015941HL8 601426ERRAD  00000066383HL1 601443RAD2  00000066269HL2 601587PROT  00000024436HL8 601672R1  00000038036HL0 601740PROT1  00000025373HL2 601764RAD1  00000020350HL5 601824ER1  00000067054HL7 601849PROT1  00000088066HL6 601883PRAD  00000081385HL7 601883PRAD1  00000031444HL3 601894RAD1  00000079510HL4 601928PXR  00000020687HL0 601947GCRAD  00000022321HL4 602239ER1  00000021508HL7 602344ER12  00000063514HL4 602344PRORM4  00000037378HL7 602344RM2  00000042198HL2 602588RM8  00000025560HL4 602824RM4  00000036451HL3 602867RM3  00000040502HL7 603292NRAD  00000045109HL6 603335FER  00000055344HL6 603335WMMCRAD  00000020595HL5 603432P3  00000037406HL6 603577R10  00000033832HL7 603577R4  00000032397HL2 603577RT5  00000027092HL6 603752PROT  00000024373HL3 603778RAD4  00000094527HL9 603924PROT  00000059713HL8 605271PROTEUS  00000023534HL1 605322PROTEUS  00000053536HL9 605322PROTEUS2  00000068785HL5 605322PROTEUSCC  00000015883HL2 605322RM10  00000043745HL9 605345RAD  00000050593HL3 605347RAD  00000101404HL2 605353PROTEUS  00000035144HL5 605353RAD  00000041405HL2 605385RAD  00000043020HL7 605432RAD  00000111897HL5 605455RAD  00000087141HL8 605472PROTEUS  00000041450HL8 605578RAD  00000053318HL2 605644RAD  00000064167HL0 605673RAD  00000015869HL1 605685RAD  00000039522HL8 605697RAD  00000046879HL3 605725RAD  00000034456HL4 605854RAD  00000043527HL1 605925PROTEUS  00000073180HL2 605964RAD  00000016988HL8 606325PROT  00000066384HL9 606666PRO  00000113340HL4 606759PRO2  00000076949HL7 606759PROT  00000041642HL0 606833PROT2  00000049770HL1 606833PROT3  00000064529HL1 606878RM3  00000041187HL6 606878XR2  00000088064HL1 606920PROT  00000048921HL1 607535PRO  00000016823HL7 607796RAD1  00000017457HL3 607796RAD2  00000098074HL8 608323PRO  00000051588HL2 608324CLPR2  00000048215HL8 608324PR3  00000065487HL1 608339RAD  00000034560HL3 608342PRT  00000066677HL6 608355PROT  00000033954HL9 608364PROT  00000046529HL4 608524PRT  00000029026HL2 608745CLRAD2  00000028875HL3 608745RAD1  00000063802HL3 608775PR10  00000028518HL9 608775PRE  00000027093HL4 608775PRT  00000023599HL4 608775RM9  00000070885HL9 608776PROT  00000018105HL7 608782PRO  00000096523HL6 608785PRO  00000016193HL5 608873RAD  00000017628HL9 609584ER1  00000017638HL8 609584ER2  00000021702HL6 609584PRO  00000016989HL6 610237ERPRO  00000015865HL9 610237PRO  00000019951HL3 610278PRO  00000041374HL0 610284PRO2  00000021698HL6 612798RAD  00000036259HL0 614257ER  00000037004HL9 614293DIAG1  00000028379HL6 614293ERB  00000068952HL1 614293GPRO  00000064044HL1 614293HC  00000032679HL3 614293RPRO  00000019225HL2 614446CX3  00000029882HL8 614566ER3  00000024429HL3 614566PTB  00000066823HL6 614688PROT  00000029027HL0 614839PROT  00000016820HL3 614861PROT  00000033831HL9 615284ERPROT  00000108820HL2 615341NGPRO  00000083396HL2 615342CRAD1  00000018379HL8 615342PROT1  00000089700HL9 615342RAD  00000068444HL9 615435WRAD4  00000028253HL3 615768SCMCCR01  00000028297HL0 615768SCMCCRER  00000025558HL8 616392PROTEUS3  00000040673HL6 616394HMI  00000041263HL5 616394PROTEUS2  00000021162HL3 616659PROTEUS  00000059575HL1 61763HL9  00000032661HL1 617726RS1Y3  00000029533HL7 617983PR2  00000018383HL0 618395R1  00000023629HL9 618395WR2  00000026065HL3 618465SAR3  00000034711HL2 618549CCR  00000029933HL9 618635RAD  00000085827HL4 618798GWER  00000029955HL2 619644PRO1  00000047228HL2 619644PRO2  00000050099HL1 619644PRO3  00000019883HL8 620223MRAD  00000016987HL0 620272JEDI  00000021836HL2 620272RM3  00000054660HL6 620340ORTHO  00000059847HL4 620356PRO  00000028381HL2 620672PRO  00000098079HL7 620886PRO  00000041990HL3 623433XR1  00000021031HL0 623445PROT  00000038259HL8 630579PROT  00000041173HL6 630652PROT2  00000018108HL1 630833ER2  00000023737HL0 630933PROT1  00000030274HL5 630933PROT2  00000023109HL2 631351ER  00000015891HL5 631476ER  00000020582HL3 631723PRO  00000018110HL7 631726PRO  00000047007HL0 631726PRO1  00000020722HL5 636239JR1  00000029959HL4 636326FRM1  00000029159HL1 636625WRM3  00000052072HL6 641444RAD  00000023371HL8 651232RAD3  00000020684HL7 651982R2  00000075269HL1 660223MRAD2  00000047518HL6 660259PROT  00000038282HL0 660562SPRO  00000015876HL6 660646PRO  00000016394HL9 660886FR3  00000049130HL8 660886FXR2  00000031438HL5 661945HDH2  00000028101HL4 661KRRAD  00000021699HL4 662289RAD  00000100282HL3 662378PRO  00000038210HL1 662615R1  00000070719HL0 662647PRO  00000047066HL6 678583XR  00000052992HL5 678604RM3  00000092361HL5 678741PRO  00000027949HL7 678945PROT  00000105205HL9 701256RAD  00000038283HL8 701567RAD  00000091770HL8 701665RAD  00000029957HL8 702346XR1  00000044030HL5 702614XR1  00000042528HL0 702614XR2  00000016399HL8 702731ERXR  00000015895HL6 702731RAD2  00000015959HL0 702731RAD4  00000039523HL6 703478PROT  00000030186HL1 703478PROT1  00000028292HL1 703689RM06  00000028100HL6 704484MOC  00000055250HL5 704873DPRO  00000049937HL6 706282SR2  00000050600HL6 706282SR2  00000045790HL3 706335BR1  00000044741HL7 706396COM  00000043328HL4 706468XR  00000019406HL8 706494DER  00000031272HL8 706517MMP  00000044742HL5 706544CC1  00000065366HL7 706544HC  00000047371HL0 706544MA3  00000039733HL1 706544MA4  00000044489HL3 706544MA5  00000050392HL0 706544WC2  00000025169HL4 706596PER  00000044065HL4 706632PRO  00000095423HL0 706655PRO  00000071149HL9 706678RAD  00000016822HL9 706754HR1  00000052783HL8 706787R3  00000023735HL4 707429NBER1  00000022469HL1 707429NBGV1  00000047069HL0 707464RAD1  00000039058HL3 707554ERRAD  00000049129HL0 707554PROT1  00000027943HL0 708388PROT  00000039123HL5 708425XR5  00000034948HL0 7087471635  00000035093HL4 7087471635  00000035143HL7 7087471639  00000024649HL6 708756JR3  00000048214HL1 708756PRO3  00000043588HL3 708783PRO1  00000043744HL2 708783PRO2  00000022465HL9 708783RM5  00000016582HL9 708783RM7  00000022591HL2 708783RM9  00000030278HL6 712279PROTEUS  00000107529HL0 712324RAD  00000029726HL7 712423BRAD1  00000045431HL4 712722PROTEUS  00000030007HL9 713285PPPRAD1  00000043019HL9 713359PROT  00000044041HL2 713383MHPROD  00000043172HL6 713778PRO  00000025556HL2 713873B15  00000028376HL2 713873B18  00000029087HL4 713873BPRO1  00000031683HL6 713873QMRAD1  00000101782HL1 714537RM3  00000065592HL8 715243PRO  00000035152HL8 715268RAD  00000082191HL8 715356PR  00000020209HL3 715425R1  00000052229HL9 715635RAD  00000037820HL8 715726RAD  00000022162HL2 715732RAD1  00000036423HL2 715743PR1  00000023602HL6 716675RAD1  00000050243HL5 717225PROT  00000019776HL4 717245R6  00000024202HL4 717290R1  00000053317HL4 717299RAD1  00000054661HL4 717624R1  00000032886HL4 717637ER1  00000073626HL4 717637R1A  00000043333HL4 717637RAD1  00000036862HL1 717653R1  00000040635HL5 717684R1  00000041256HL9 717684R1  00000095424HL8 717733GDSR1  00000093306HL9 717733GR1  00000021389HL2 717733HP  00000094650HL9 812265PROT  00000031686HL9 812265TOMO  00000073390HL7 812268PROT  00000085129HL5 812385RAD  00000023264HL5 812450RT11  00000033953HL1 812482ER2  00000034670HL0 812482TOMO  00000039122HL7 812663PROT2  00000041451HL6 812683RAD  00000022464HL2 812738RAD1  00000107533HL2 812847PROT  00000036694HL8 812858MRAD  00000102503HL0 812932ORTHO  00000086878HL6 812934PROT1  00000067371HL5 813558ERPRO  00000018752HL6 813571XPRO  00000016814HL6 813634RAD  00000082188HL4 813763BRPCER1  00000042901HL9 813873ERPRO  00000040837HL7 813971PPRO  00000018987HL8 814226R1  00000018301HL2 814274R1  00000064532HL5 814450RAD  00000042475HL4 814467R1  00000028572HL6 814533ER1  00000020351HL3 814533PROT1  00000035151HL0 814533R8  00000025615HL6 814643R4  00000017160HL3 814827R  00000020069HL1 814838XR  00000019950HL5 815235RAD3  00000016401HL2 815243RAD  00000038208HL5 815478RAD  00000036241HL8 815664MX1  00000054530HL1 815673PROT1  00000023903HL8 815741RM2  00000019158HL5 815741RM5  00000044407HL5 816276PROT1  00000045283HL9 816276PROT2  00000025419HL3 817255NHTOMO  00000018808HL6 817347MPTOMO  00000078549HL3 817347PRODIG  00000048255HL4 817408PRO1  00000053541HL9 817473PRO  00000060395HL1 817599PREOP  00000061518HL7 817599PREOP  00000077761HL5 817639ALLPRO  00000035576HL8 818502PROT1  00000103414HL9 818563DR  00000025417HL7 818676PROT1  00000028678HL1 818884ER  00000042326HL9 828213AI118  00000042328HL5 828213AI119  00000041643HL8 828213RCR1  00000041615HL6 828213RCR2  00000029539HL4 828315FRC  00000033427HL6 828681PRER1  00000045794HL5 828697LPR1  00000034241HL0 828754CER  00000050603HL0 830298PRO  00000050872HL1 830298PRO  00000075815HL1 830 426MRH2  00000087778HL7 830608DIGPRO  00000018695HL7 831424RAD1  00000078009HL8 84016HL5  00000045156HL7 843236CFR1  00000052227HL6 843355RAD1  00000052071HL8 843549RAD2  00000040638HL9 845279P4  00000016591HL0 845534ER  00000048623HL3 845938RM3  00000033961HL4 847316PROER1  00000023263HL7 847432ER  00000028579HL1 847471RAD  00000025124HL9 850325RAD1  00000015957HL4 850434BER  00000024552HL2 850494WER2  00000023374HL2 850584RAD1  00000041582HL8 850626SR1  00000032182HL8 850747GC2  00000018098HL4 850863WR3  00000018099HL2 850863WR3  00000063513HL6 850875RAD1  00000031859HL2 850995PMC1  00000073389HL9 856246PROT2  00000029124HL8 859572ER1  00000029124HL5 859572ER2  00000023174HL6 859572RM1  00000022508HL6 859962RM1  00000022472HL5 859962RM2  00000028178HL2 859967PROT  00000029625HL1 859986XR1  00000022919HL5 860364RM4  00000029535HL2 860444XER2  00000031441HL9 860444XER2  00000032557HL1 860446EDX  00000032302HL2 860585RM1  00000035975HL2 863298RM1  00000082843HL4 863314PROT  00000019946HL3 863763RM3  00000087138HL4 209895RAD1  00000024745HL2 217333LXR2  00000096683HL8 218935PRO  00000080729HL7 781762RM1  00000034451HL5 303399RAD11  00000023033HL4 336768OSR1  00000023539HL0 205387PRO  00000047450HL2 209578ER1  00000107535HL7 713236PRAD1  00000029820HL8 803407R1  00000052075HL9 912287R608  00000032265HL1 434924KR1  00000028502HL3 602231XR1  00000042282HL4 918756XR1  00000002068HL3 479484PROT  00000020681HL3 479484PROT  00000045707HL7 864255XRA1  00000056810HL5 864400XRA1  00000018496HL0 856488PROT  00000017812HL9 856582PROT  00000020948HL6 251275RAD  00000080063HL1 251275RAD2  00000019233HL6 423744RAD3  00000043496HL9 423745ER  00000041756HL8 518525RAD1  00000074624HL8 AE1059RX01  00000016589HL4 0910072078  00000021506HL1 0910072103  00000021502HL0 0910072104  00000031275HL1 0910072122  00000042530HL6 0910072144  00000042616HL3 0910072146  00000042784HL9 0910072146  00000051665HL8 0910072171  00000076508HL1 0910072216  00000090520HL8 0910072230  00000091292HL3 0910072231  00000021393HL4 0910073062  00000024662HL9 0910073066  00000045406HL6 0910073081  00000046047HL7 0910073082  00000081588HL6 0910073111  00000061397HL6 0910074024  00000091102HL4 0910074035  00000082325HL2 0910075021  00000082324HL5 0910075022  00000076420HL9 0910076047  00000094362HL1 83089838806  00000096687HL9 83091661241  00000100279HL9 83093535827  00000024740HL3 0910073065  00000021396HL7 0910072090  00000063689HL4 BA4106RX01  00000063693HL6 BA4147RX01  00000063800HL7 BA4149RX01  00000063691HL0 BA4151RX02  00000063517HL7 BA4153RX01  00000025123HL1 90700  00000025119HL9 2543552  00000024906HL0 85179  00000022509HL4 85182  00000025068HL8 85186  00000024903HL7 85194  00000025065HL4 85211  00000042385HL5 BG4549RX02  00000042689HL0 BG4549RX03  00000062474HL2 0847070048  00000062476HL7 0847070075  00000072708HL1 306338RAD1  00000048359HL4 306873RAD2  00000051824HL1 709832RAD  00000036190HL7 506422XRA  00000045938HL8 506623PROTEUS  00000040400HL4 204239RAD  00000031440HL1 204765XRA  00000020170HL7 204836RAD1  00000035918HL2 204945XRA  00000031151HL4 250314XRA  00000029880HL2 250365XRA  00000030182HL0 250395XRA  00000024661HL1 250412RAD1  00000029787HL9 250428XRA  00000035026HL4 250491XRA  00000030801HL5 250545XRA  00000018992HL8 250632XRA  00000036242HL6 780945XRA  00000024468HL1 807343BRAD1  00000036261HL6 807343BRAD2  00000020073HL3 807345PROTEUS  00000032266HL9 807626RAD3  00000036189HL9 902427XRA  00000042200HL6 902428ARAD1  00000046335HL6 902597RAD1  00000020344HL8 902625PROTEUS  00000035368HL0 902765XRA  00000041646HL1 902794PROTEUS  00000091294HL9 902794XR1  00000072366HL8 902862RAD1  00000072364HL3 902862RAD2  00000020077HL4 905336RAD1  00000020171HL5 905682RAD  00000018807HL8 905732RAD1  00000016181HL0 905845RAD  00000017461HL5 905845RAD2  00000020357HL0 905895XRA  00000033500HL0 905895XRA2  1-89097903605 905895XRA2  00000065034HL1 306778RAD7  00000028680HL7 780585RAD2  00000039207HL6 250782XRA  00000030962HL5 250945XRA  00000036193HL1 250945XRA  00000038034HL5 250949XRA  00000039751HL3 250956RAD1  00000030957HL5 306332RAD1  00000016584HL5 306352RAD1  00000042195HL8 306359RAD1  00000054815HL6 306435RAD  00000055251HL3 306628RAD2  00000048258HL8 306752RAD1  00000031793HL3 306763RAD1  00000020212HL7 306775RAD1  00000037345HL6 306778RAD4  00000042329HL3 306778RAD5  00000045281HL3 306778RAD6  00000048363HL6 306862RAD1  00000031153HL0 306865RAD1  00000046136HL8 403212RAD1  00000021841HL2 403227RAD1  00000041994HL5 403259XRA  00000022915HL3 403343RAD2  00000021507HL9 403343RM1  00000020862HL9 403556RAD1  00000033876HL4 403777GRAD1  00000023597HL8 403777HRAD1  00000025378HL1 403782RAD1  00000022159HL8 403783RAD1  00000040959HL9 416348XRA  00000089901HL3 418589XR1  00000036132HL9 418677XRA  00000018384HL8 418682PROTEUS  00000031277HL7 418844PROTEUS  00000051826HL6 418844PROTEUS2  00000036282HL2 450358XRA  00000017460HL7 506632RM11B  00000057993HL8 514398RAD1  00000078550HL1 514495XR1  00000083638HL7 514495XR2  00000083668HL7 514495XR2  00000040589HL4 514697PROTEUS  00000040672HL8 519254PRO1  00000040587HL8 519254PRO2  00000017139HL7 519396RAD  00000033662HL8 519426XRA  00000034452HL3 519485XRA  00000003958ML8 519661RAD1  00000032493HL9 519735PRO  00000034024HL0 519842XRA  00000028878HL7 519843XRA  00000015884HL0 604278XRA  00000023594HL5 604775XRA  00000063803HL1 604775XRA2  00000041174HL4 604869XRA  00000017820HL2 604875CXRA  00000031445HL0 613392XRA  00000036244HL2 613541XRA  00000067609HL8 613548BRAD1  00000036262HL4 613687XRA  00000019407HL6 613933RAD  00000016194HL3 705286RAD  00000036191HL5 705424XRA  00000016195HL0 705856RAD  00000016811HL2 709884PROTEUS  00000022918HL7 819864PROTEUS  00000064042HL5 604775XRA3  00000059775HL7 613351RAD1  00000059772HL4 613536RAD1  00000045615HL2 HSRXR1  00000046045HL1 HSRXR2  00000096685HL3 XR24609M30  00000081876HL5 XR98881HL6  00000041133HL0 082407100301  00000045209HL4 082407160217  00000034242HL8 082407030307  00000022923HL7 082407100129  00000032893HL0 082407250034  00000025171HL0 082407010123  00000028634HL4 082407020293A  00000023603HL4 082407040103  00000028038HL8 082407040129  00000028031HL3 082407040130  00000028250HL9 082407040169  00000029534HL5 082407040209  00000023898HL0 082407050043  00000026061HL2 082407050046  00000028039HL6 082407050055  00000029334HL0 082407060018  00000051667HL4 082407060050  00000019403HL5 082407070037  00000024741HL1 082407070048  00000035239HL3 082407070077  00000040504HL3 082407070085  00000054659HL8 082407070095  00000064699HL2 082407070120  00000031857HL6 082407090069  00000066962HL2 082407090136  00000043174HL2 082407140124  00000024206HL5 082407160084  00000038947HL8 082407160198  00000031858HL4 082407180033  00000034161HL0 082407180038  00000025173HL6 082407200025  00000019882HL0 082407270078  00000025174HL4 082407290024  00000030280HL2 082407020303  00000029285HL4 082407030105  00000048520HL1 082407030329  00000058227HL0 082407030339  00000020219HL2 082407040074  00000022556HL5 082407040092  00000031024HL3 082407040216  00000032892HL2 082407040233  00000046668HL0 082407040367  00000047451HL0 082407040377  00000047885HL9 082407040378  00000029629HL3 082407070068  00000055685HL2 082407070103  00000031609HL1 082407080015  00000018302HL0 082407090029  00000028635HL1 082407100182  00000028633HL6 082407100197  00000049457HL5 082407100370  00000039732HL3 082407110020  00000030272HL9 082407120103  00000040989HL6 082407140120  00000029204HL5 082407150021  00000028176HL6 082407180023  00000040768HL4 082407180040  00000032883HL1 082407190034  00000032890HL6 082407190037  00000052991HL7 082407290040  00000057434HL3 082407290043  00000041100HL9 2608914PROTEUS  00000082525HL7 XR81884HL9  00000024733HL8 73244PROTEUS  00000037005HL6 2625645PROT  00000070588HL9 2906074PROT  00000043176HL7 CS1304RX01  00000084212HL0 CS1337RX01  00000036600HL5 600033RX04  00000043495HL1 CS1042RX01  00000016407HL9 CS1156RX01  00000018382HL2 CS1156RX02  00000018602HL3 CS1175RX01  00000018601HL5 CS1175RX02  00000043810HL1 CS1305RX01  00000052565HL9 7245PROTEUS  00000064321HL3 EG1039RX05  00000064322HL1 EG1039RX06  00000017156HL1 FI1112RX01  00000104757HL0 M4145030  00000015877HL4 M4163744  00000017154HL6 M4205408  00000035365HL6 M4207611  00000030903HL9 M4448013  00000023265HL2 M5484815  00000025375HL7 M9624229  00000031159HL7 HC3746XR02  00000095425HL5 HC7638XR01  00000041315HL3 HROS01XR04  00000019685HL7 PC8109XR01  00000101037HL0 PX5326XR02  00000016342HL8 H2014RAD02  00000019151HL0 H2022RAD08  00000048918HL7 H2097RAD06  00000058606HL5 H2165RAD02  00000044488HL5 H2188RAD04  00000058724HL6 H2199RAD02  00000058723HL8 H2212RAD01  00000058604HL0 H2213RAD01  00000058605HL7 H2214RAD01  00000058607HL3 H2216RAD01  00000085825HL8 XR85143HL6  00000081137HL2 XR84557HL8  00000081381HL6 XR84563HL6  00000061398HL4 2836332XRAY  00000072090HL4 2933763XR  00000075816HL9 2948102XR  00000016733HL8 0828070089  00000020594HL8 0828070092  00000022759HL5 0828070092  00000104752HL1 HU1008RX33  00000088857HL8 0835070198  00000099911HL0 0835070204  00000104064HL1 0835070215  00000032301HL4 IE1017RX02  00000044745HL8 IE1019RX03  00000021900HL6 IE1047RX01  00000050388HL8 IE1135RX01  00000057597HL7 IE1138RX01  00000020767HL0 IS1002RX03  00000023628HL1 A5413816  00000018107HL3 A5699201  00000018504HL1 B5356001  00000016196HL8 XU0009  00000016981HL3 XU0010  00000016740HL3 XU0011  00000018604HL9 XU0013  00000018616HL3 XU0014  00000019536HL2 XU0015  00000058130HL6 XU0017  00000060272HL2 XU0018  00000069563HL5 XU0019  00000021029HL4 KA1022RX01  00000090523HL2 KW1009RX03  00000102500HL6 LV4009RX01  00000030416HL2 LT4001RX02  00000100470HL4 0834070213  00000104277HL9 0834070227  00000021390HL0 72164PROTEUS  00000023739HL6 78797PROHM  00000020680HL5 8320106PROTEUS  00000027948HL9 21250PROTEUS  00000023630HL7 65593PRO2HM  00000032398HL0 662PROTEUS  00000031795HL8 TLB0211  00000016398HL0 0920079046  00000045284HL7 0920079084  00000045367HL0 0920079085  00000067372HL3 0920079126  00000074883HL0 0920079133  00000083057HL0 0920079145  00000045110HL4 NO1108RX03  00000055249HL7 PK1300RX05  00000079707HL6 PK1354RX03  00000084772HL3 0843070185  00000023738HL8 PL1005RX03  00000029232HL6 PL1043RX02  00000040430HL1 PL1107RX02  00000030081HL4 PL1228RX01  00000030008HL7 PL1555RX01  00000099136HL4 PL1639RX01  00000046430HL5 PL1671RX01  00000052240HL9 PL1696RX01  00000038333HL1 PL3101RX01  00000023111HL8 PF1006RX02  00000030728HL0 PER08733  00000056260HL3 PER100200  00000067236HL0 PER100300  00000055065HL7 PER100400  00000074015HL9 PER12310  00000071637HL3 PER18123  00000070886HL7 PER18127  00000028876HL1 PER18604  00000018380HL6 PER18606  00000021188HL8 PER18606  00000056261HL1 PER30002  00000041212HL2 PER39502  00000055875HL9 PER59501  00000055348HL7 PER99500  00000055345HL3 PER99600  00000019683HL2 PPR01322  00000030725HL6 PPR15708  00000021903HL0 PPR64101  00000043161JL0 PPR00172  00000041316HL1 600130RX05  00000041318HL7 600130RX07  00000029883HL6 RU1029RX01  00000055681HL1 RU1067RX02  00000017629HL7 RU1111RX07  00000029123HL7 RU1116RX01  00000040586HL0 RU1283RX05  00000018102HL4 RU2166RX01  00000027846HL5 RU2238RX01  00000110793HL7 RU2250RX02  00000020214HL3 RU2252RX01  00000054934HL5 RU2441RX02  00000034710HL4 RU2526RX01  00000032762HL7 RU2589RX01  00000032664HL5 RU2706RX01  00000040781HL7 RU2714RX01  00000039205HL0 RU2725RX01  00000064325HL4 RU3421RX01  00000070590HL5 RU3665RX02  00000069441HL4 RU3725RX02  00000069443HL0 RU3815RX01  00000075648HL6 RU7172RX01  00000090524HL0 RU7884RX01  00000021500HL4 SA1039RX06  00000033722HL0 SA1010RX03  00000033233HL8 SA1071RX02  00000023107HL6 SA1208RX01  00000024645HL4 SA1208RX03  00000022387HL5 SA1251RX01  00000022396HL6 SA1254RX01  00000024650HL4 SA1258RX01  00000024583HL7 SA1276RX01  00000029122HL9 SA1304RX01  00000029446HL2 SA1347RX01  00000032560HL5 SA1361RX01  00000036454HL7 SA1387RX02  00000046255HL6 SA2116RX01  00000024648HL8 SA1014RX09  00000033425HL0 SA1129RX02  00000019402HL7 SA1158RX02  00000019234HL4 SA1231RX04  00000018609HL8 SA1231RX05  00000060270HL6 ZA2255RX01  00000069562HL7 ZA1030RX04  00000069844HL9 ZA1030RX05  00000070884HL2 ZA1030RX06  00000070718HL2 ZA1030RX07  00000070716HL6 ZA1030RX08  00000020775HL3 ZA2122RX01  00000027094HL2 0850070692  00000018986HL0 0850070477  00000020770HL4 0850070485  00000023169HL6 0850070637  00000030005HL3 0850070707  00000042129HL7 0850070792  00000015890HL7 DE647A06  00000019954HL7 DP226703  00000058127HL2 DP226711  00000052989HL1 DPB08419  00000028981HL9 DPB17711  00000094523HL8 EPP05800  00000019162HL7 FE394804  00000043638HL6 FEG85406  00000042381HL4 FP970408  00000104066HL6 FPE60907  00000055350HL3 FPG07402  00000094655HL8 FPG10903  00000087603HL7 FPG23602  00000021327HL2 FPG30519  00000109669HL2 FPG66604  00000028032HL1 GEQ12201  00000045616HL0 GPX02109  00000037484HL3 ME203017  00000044032HL1 ME203022  00000041824HL4 ME242402  00000041759HL2 ME246210  00000042131HL3 ME246211  00000042158HL6 ME250406  00000041825HL1 ME256647  00000021697HL8 ME612427  00000022163HL0 ME667518  00000027852HL3 ME667520  00000058407HL8 ME707211  00000042077HL8 MEX00808  00000038281HL2 MP228411  00000016350HL1 MP491301  00000023631HL5 MP697607  00000046332HL3 MP926412  00000097812HL2 MPX16800  00000042457HL2 MPX28402  00000028685HL6 MPX39902  00000029021HL3 MPX39903  00000037409HL0 MPX77008  00000037482HL7 MPX77009  00000036863HL9 MPX82224  00000025566HL1 NE219630  00000040377HL4 NE435014  00000019775HL6 NE655316  00000056043HL3 NPV79304  00000056091HL2 NPV79305  00000019771HL5 UE276008  00000024366HL7 UE535A75  00000023902HL0 UE535A76  00000028781HL3 UE743A01  00000017828HL5 YP231724  00000092088HL4 YP600801  00000022385HL9 Z008RX01  00000022391HL7 Z009RX01  00000022920HL3 547110XR02  00000021326HL4 475030XR03  00000025069HL6 0853070274  00000022761HL1 0853070266  00000022755HL3 0853070267  00000023833HL7 0853070272  00000023179HL5 0853070285  00000032491HL3 0853070302  00000036324HL2 0853070305  00000037715HL0 0853070315  00000037714HL3 0853070316  00000043587HL5 0853070319  00000043743HL4 0853070326  00000059450HL7 0853070345  00000111467HL7 0853070357  00000034088HL5 0853070307  00000035238HL5 0853070309  00000057598HL5 0856070193  00000089903HL9 0856070257  00000040982HL1 0856070177  00000094361HL3 0856070248  00000034092HL7 UK1092RX02  00000019880HL4 UK5026RX01  00000037889HL3 10509RAD01  00000058727HL9 10599RAD01  00000020717HL5 77119RAD06  00000031796HL6 77119RAD07  00000019687HL3 00006RAD01  00000036611HL2 00019RAD01  00000023535HL8 00053RAD13  00000023740HL4 00053RAD14  00000034026HL5 00053RAD17  00000034430HL9 00068RAD07  00000015894HL9 00082RAD07  00000020585HL6 00082RAD09  00000050590HL9 00091RAD09  00000052079HL1 00091RAD10  00000052566HL7 00091RAD11  00000064818HL8 00094RAD04  00000024585HL2 00095RAD09  00000041649HL5 00100RAD02  00000063696HL9 00100RAD03  00000063511HL0 00100RAD04  00000018612HL2 00116RAD03  00000031789HL1 00129RAD09  00000019785HL5 00131RAD06  00000032267HL7 00131RAD07  00000031903HL8 00140RAD02  00000080059HL9 00141RAD05  00000059714HL6 00169RAD09  00000023406HL2 00233RAD10  00000023403HL9 00233RAD11  00000017452HL4 00243RAD09  00000017626HL3 00243RAD10  00000017625HL5 00243RAD11  00000017620HL6 00243RAD12  00000019409HL2 00253RAD12  00000094200HL3 00254RAD01  00000025126HL4 00263RAD15  00000025555HL4 00282RAD02  00000036188HL1 00303RAD02  00000033497HL9 00309RAD01  00000053414HL9 00313RAD07  00000053613HL6 00313RAD08  00000019684HL0 00329RAD01  00000100281HL5 00341RAD04  00000035301HL1 00373RAD04  00000037817HL4 00373RAD06  00000028251HL7 00378RAD03  00000019566HL9 00442RAD03  00000031271HL0 00488RAD07  00000019400HL1 00508RAD03  00000015886HL5 00510RAD08  00000016349HL3 00510RAD09  00000029815HL8 00521RAD03  00000031684HL4 00521RAD04  00000049935HL0 00554RAD01  00000050098HL3 00572RAD02  00000060271HL4 00572RAD04  00000099138HL0 00617RAD08  00000017148HL8 00680RAD02  00000031443HL5 00695RAD02  00000047425HL4 00701RAD05  00000047449HL4 00701RAD06  00000060158HL3 00709RAD02  00000023262HL9 00755RAD04  00000028291HL3 00755RAD05  00000063804HL9 00762RAD02  00000031608HL3 00784RAD02  00000032164HL6 00784RAD03  00000031687HL7 00785RAD08  00000017149HL6 00793RAD02  00000054813HL1 00824RAD03  00000050095HL9 00824RAD04  00000019154HL4 00833RAD02  00000046131HL9 00959RAD02  00000024466HL5 00965RAD02  00000030181HL2 01007RAD01  00000017459HL9 01203RAD02  00000031027HL6 01205RAD03  00000019686HL5 01255RAD02  00000047517HL8 05228RAD02  00000019689HL9 05452RAD03  00000039489HL0 05452RAD05  00000045943HL8 05523RAD03  00000063181HL2 05528RAD03  00000022320HL6 10134RAD02  00000094524HL6 10183RAD02  00000030912HL0 10251RAD01  00000051935HL5 10307RAD01  00000017816HL0 10362RAD01  00000018111HL5 10362RAD02  00000017143HL9 10390RAD01  00000019540HL4 10396RAD02  00000022066HL5 10428RAD01  00000021163HL1 10433RAD03  00000023743HL8 10448RAD02  00000023627HL3 10448RAD03  00000023741HL2 10448RAD04  00000030802HL3 10459RAD01  00000030803HL1 10459RAD02  00000029056HL9 10460RAD01  00000033721HL2 10469RAD01  00000033720HL4 10477RAD01  00000034160HL2 10490RAD01  00000037230HL0 10493RAD01  00000036095HL8 10494RAD01  00000035972HL9 10494RAD02  00000036793HL8 10498RAD02  00000040962HL3 10516RAD01  00000040836HL9 10516RAD02  00000041134HL8 10518RAD01  00000041262HL7 10519RAD02  00000046432HL1 10548RAD01  00000051827HL4 10563RAD01  00000054814HL9 10575RAD01  00000054933HL7 10575RAD02  00000055682HL9 10591RAD01  00000061065HL9 10605RAD01  00000060160HL9 10606RAD01  00000060065HL0 10609RAD01  00000060159HL1 10609RAD01  00000064528HL3 10625RAD01  00000081138HL0 10670RAD01  00000070370HL2 IE1502RX02  00000101403HL4 IE1564RX03  00000094653HL3 XR104248HL0  00000036979HL3 2571797PROTEUS  00000055680HL3 3882756515XRAND  00000035535HL4 AN-2561355-XR  00000042617HL1 BD-2629649-XR  00000042618HL9 BD-2629649-XR  00000056598HL6 TT-2790634-XR  00000043960HL4 GU-2647968-XR  00000049767HL7 SU-2714528-XR  00000046044HL4 3882672727XRCTR  00000049035HL9 3882706094XRAND  00000059337HL6 3882756170XRCTR  00000056258HL7 3882756601XROCC  00000028498HL4 38842510XRMET  00000022587HL0 0859070084  00000022752HL0 0859070084",
      "recall_initiation_date": "20141208",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}