  "results": [
    {
      "status": "Terminated",
      "city": "Somerville",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87088",
      "recalling_firm": "Ethicon, Inc.",
      "address_1": "Us Highway 22 West",
      "address_2": "",
      "postal_code": "08876",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-0953-2021",
      "product_description": "Ethicon Dermabond Advanced Topical Skin Adhesive-topical skin adhesive to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions   Product Code: DNX12",
      "product_quantity": "23,580 eaches",
      "reason_for_recall": "Product distributed with a red tape as part of the sterile packaging which could impact the seal and result in a sterility breach",
      "recall_initiation_date": "20210104",
      "center_classification_date": "20210204",
      "termination_date": "20231129",
      "report_date": "20210210",
      "code_info": "Lot Number: QEBBTQ  Exp. Date: April 30, 2022  GTIN: 10705031203532",
      "more_code_info": ""
    }
  ]
}