  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Reading",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81375",
      "recalling_firm": "Arrow International Inc",
      "address_1": "2400 Bernville Rd",
      "address_2": "",
      "postal_code": "19605-9607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution.",
      "recall_number": "Z-0952-2019",
      "product_description": "AF w/AS 550 x 0.5-7 ml,w/Y site,DL CPNB, Product code MVAXL2Y CPNB, Arrow AutoFuser Disposable Pain Control Pump",
      "product_quantity": "26",
      "reason_for_recall": "Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay  while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug  delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This  may result in the need for other, additional analgesic drug administration.",
      "recall_initiation_date": "20181002",
      "center_classification_date": "20190225",
      "termination_date": "20211112",
      "report_date": "20190306",
      "code_info": "74A1600493 74E1701689  74A1701278 74F1601947  74B1700147 74G1601342  74B1700922 74H1601292  74C1601403 74H1602005  74C1601458 74J1601528  74C1601839 74K1601040  74C1602928 74K1601957  74C1702397 74L1601996  74D1600766 74L1602670  74E1600739 74M1501120  74E1601599 74M1600311  74E1700481 74M1600671",
      "more_code_info": ""
    }
  ]
}