{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Murrysville",
      "address_1": "1001 Murry Ridge Ln",
      "reason_for_recall": "Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.",
      "address_2": "",
      "product_quantity": "16",
      "code_info": "Model/Material Number: ND2110X15B  Serial Numbers: H251017836F15, H2523309685A2, H252331351095, H252353329A08, H25504805F0DF, H255049394A07, H2550648781D8, H255065746A57, H25506638283F, H25506755FDD6, H255068597D7D, H25523364DD89, H255249956B93, H25524999A1FF, H2569010452BD, H2569170852B3,",
      "center_classification_date": "20200214",
      "distribution_pattern": "US: None  OUS: CH DE DK IT FR IE GB KW QA PL ES AT PH AE SI CZ",
      "state": "PA",
      "product_description": "Trilogy EVO Ventilator, Nordics - Model Number: ND2110X15B    Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories.",
      "report_date": "20200226",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Philips Respironics, Inc.",
      "recall_number": "Z-0948-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "84406",
      "termination_date": "20210331",
      "more_code_info": "",
      "recall_initiation_date": "20191119",
      "postal_code": "15668-8517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}