  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Des Plaines",
      "address_1": "1300 E Touhy Ave",
      "reason_for_recall": "The x-drive assembly may fail prematurely.  This could cause the manipulator or liquid handling arm to contact an object, shake and drip, or improperly pipet.",
      "address_2": "",
      "product_quantity": "176 devices",
      "code_info": "List Number: 09K14-02; Model: E-series; Serial Numbers: 10664 - 10833, 1004, 1005, 10142, 10154, 10171, 10416",
      "center_classification_date": "20150109",
      "distribution_pattern": "Worldwide Distribution - USA including CA, FL, GA, IL, MI, NC, NJ, NY, OH, OK, OR, and TX and Internationally to Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, China, Colombia, Costa Rica, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Indonesia, Ireland, Israel, Italy, Kazakhstan, Kenya, Malawi, Mali, Mexico, Morocco, Mozambique, New Zealand, Nigeria, Norway, Pakistan, Poland, Russia, Saudi Arabia, Senegal, South Africa, South Korea, Taiwan, Thailand, Turkey, Ukraine, and United Kingdom.",
      "state": "IL",
      "product_description": "The Abbott m2000sp is an automated system for performing sample preparation for nucleic acid testing.",
      "report_date": "20150121",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Abbott Molecular",
      "recall_number": "Z-0948-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69300",
      "termination_date": "20171204",
      "more_code_info": "",
      "recall_initiation_date": "20131123",
      "postal_code": "60018-3315",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}