  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Grand Rapids",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87009",
      "recalling_firm": "Mar-Med Co",
      "address_1": "333 Fuller Ave Ne",
      "address_2": "",
      "postal_code": "49503-3630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Australia, Canada, France, Germany, Greece, Italy, Japan, Netherlands, Singapore, Spain, Sweden, Switzerland, and United Kingdom.",
      "recall_number": "Z-0945-2021",
      "product_description": "Tourni-Cot - Universal, UDI 00855364004085, Model Number TCU,  Non-pneumatic Tourniquet - Product Usage: The device is intended to be used in surgery on fingers or toes to provide a blood-less field.",
      "product_quantity": "52,000",
      "reason_for_recall": "A failure can occur if the user attempts to remove the device by partially snipping the edge and pulling on it. The device may tear into two pieces with a portion remaining in place, which can continue to constrict the vessels. This type of failure can result in digit ischemia.",
      "recall_initiation_date": "20201228",
      "center_classification_date": "20210202",
      "termination_date": "20230914",
      "report_date": "20210210",
      "code_info": "All lots",
      "more_code_info": ""
    }
  ]
}